Recalling and Saving States
Saving States
Recalling States
Chapter 2
This section contains information about saving and recalling a Multiplexer
Module state. The switchbox driver can store up to 10 states.
The
<numeric_state> command saves the current instrument state.
*SAV
The state number (0-9) is specified in the state parameter. The following
settings are saved:
Channel Relay State (channels 00 through 63 open or closed)
Control Relay States (channels 90 through 94 open or closed)
•
ARM:COUNt
•
TRIGger:SOURce
•
OUTPut[:STATe]
•
INITiate:CONTinuous
•
[ROUTe:]SCAN:MODE
•
[ROUTe:]SCAN:PORT
The
<numeric_state> command recalls a previously saved state.
*RCL
Enter the number (0-9) in the state parameter of the desired saved state. If
was not previously executed using the selected number, the
*SAV
Multiplexer Module will configure to the reset values (see "Table 2-1. HP
E1476A Default Conditions for Power-on and Reset" on page 38).
Note
Scan lists are not saved when a state is saved. You must re-enter your scan
list after recalling a state.
HP E1476A Switchbox Application Examples 53