4.6.4 Wiring Procedure
1. See
Figure 15,
2. Remove the end cap cover from the terminal block end of the electronics housing.
3. Feed loop power leads through one end of the conduit entrances on either side of the
electronics housing. The transmitter accepts up to 16AWG wire.
4. Plug the unused conduit entrance with the appropriate plug for the environment.
5. Connect the positive loop power lead to the positive (+) terminal and the negative loop power
lead to the negative (-) terminal. Note that the transmitter is not polarity-sensitive.
6. Replace the end cap, and secure it in place.
Page 32
above, for parts locations. Loosen the end cap lock using a 1.5mm Allen wrench.
ST 700 SmartLine Pressure Transmitters User's Manual
Revision 13