MicroDock II
Users Manual

Troubleshooting

If a problem is encountered, refer to the solutions provided in the following table. If the problem cannot be resolved, contact
BW
Technologies. [
Select GasAlertMicro, GasAlertMicro 5, and GasAlertMicro 5 PID docking modules only].
*
Problem
The station does not activate.
The detector fails a bump test.
70
Table 20. Troubleshooting Tips
Possible Cause
No power connection
No batteries
Depleted batteries
Damaged or defective station
Barbed fittings not connected correctly
with the gasket
Detector alarm setpoints are set higher
than the gas concentration levels of the
gas cylinder
Gas connections are not attached
correctly or are blocked
Gas inlets are not setup correctly
Gas cylinder is empty
Detector and/or station are damaged or
defective
Solution
Connect the power adapter.
Insert the batteries.
Replace the batteries.
Contact
BW
Technologies.
Confirm connection - refer to
Adding Another Docking
Module.
Connect a gas cylinder that has a
higher gas concentration level than the
alarm setpoints.
Confirm that all gas connections are
attached correctly.
Refer to
Inlet Setup
and Installation.
Use a new gas cylinder.
Contact
BW
Technologies.