When the required inlet value displays, press C
6.
sel to confirm the selection and deactivate the field.
The
no longer displays to the right of the inlet
*
indicating the conflict has been resolved.
Unsuccessful Conflict Resolution: If the conflict is not
resolved by selecting a different inlet, refer to the following:
Ensure that the selected inlet is correct.
Ensure the gas cylinder that is attached to the selected
inlet is correct.
Press C OK to accept the station's default selection.

Not Found

If a value of 1 or more displays before Not Found, complete
the following:
Or
Press C cancel to quit the current bump check/calibration.
Press C more to access the gas type/inlet screen
1.
to view which gas type(s) cannot be found.
The
icon displays on the far right side on the row
*
of the gas that cannot be found. The station
automatically defaults to the Purge inlet.
Press C OK to accept the default Purge inlet.
2.
Or
Press sel to select a different inlet.
For gases not found: If sel is not selected within 30 seconds,
the station automatically defaults to the Purge inlet.
3.
If required, repeat steps #1-2 for any additional
gases not found and confirm the following:
Ensure that the selected inlet is correct.
Ensure that the gas cylinder attached to the selected
inlet is correct.
Press C OK to accept the station's default selection.
MicroDock II
Gas Conflicts
Note
51