3000 Series Indicators

3.4.3 Graduation

Set the scale readability.
0.0005, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20.
NOTE: Not all settings are available for each capacity. Refer to the Setup Table 3-3 for available settings.

3.4.4 Power On Unit

Set the unit that will be active at power on.
Auto (last unit in use when power was turned off.), kg, g, lb

3.4.5 Zero Range

Set the percentage of scale capacity that may be zeroed.
2%
= zero up to 2 percent of capacity
100% = zero up to full capacity

3.4.6 End Setup

Advance to the next menu.
3.5

Readout Menu

Enter this menu to customize display functionality.

3.5.1 Reset

Set the Readout menu to factory default settings.
No
= not reset
Yes
= reset
If the Legal for Trade menu item is set to ON, the Stable Range, Averaging Level, Auto Zero Tracking and Auto
Off settings are not reset.
Reset: No, Yes
Stable Range: 0.5d, 1d, 2d, 5d
Filter: Lo, Med, Hi
Auto Zero Tracking
Off, 0.5d, 1d, 3d
Sleep: Off, On
Light: HI, Med, Low
Auto Off: Off, 1, 5, 10 (min)
Expand: Off, On
End Readout
Exit READOUT menu
EN-19
0.0005
pr.UNit
UNIt
kg