Five Inner Door Option; Water-Cooled Condenser - Thermo Scientific Forma 900 Series Operating And Maintenance Manual

86c ult upright freezer
Table of Contents

Advertisement

Water-cooled
Condenser
Table 5-1. Water-Cooled Condenser Specifications
Water Source
Water Pressure
Water Temperature Range
Inlet Connection
Outlet Connection
Flow Rate Required
Drain Required
Five Inner Door
Thermo Scientific
The water-cooled condenser is a factory installed option (195964 [13 cu
ft], 195965 [17, 23, 28 cu ft], 195967 [12, 17, 20 cu ft chest]) and
requires a qualified technician at freezer installation. Installation should
include proper adjustment of the regulating valve, which controls the
discharge pressure. Refer to Table 5-1 for the specifications for this option.
Tower
Not to exceed 150 psig
Not to exceed 29.4C (85F)
0.5" compression
0.5" compression
3.0 gallons (11.4 liters) per minute
No (return line is required)
The five inner door option (P/N 189405 [13 cu ft], 189406 [17 cu ft],
189407 [23 cu ft], 195652 [28 cu ft]) is factory installed. The freezer is
Option
converted to accommodate four adjustable specimen shelves with the fifth
"shelf" as the bottom of the freezer chamber.
Factory Installed Options
City
1.0 gallon (3.8 liters) per minute
Yes
Model 900 and 5900 Series
Section 5
5-9