Thermo Scientific Forma 900 Series Operating And Maintenance Manual page 43

86c ult upright freezer
Table of Contents

Advertisement

Section 6
Specifications
Model (Catalog No.)
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical- nominal
voltage ±10%
Breaker Requirements
Shipping Weight
______________________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand
safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the
category used for instruments in installations supplied from a supply comparable to public mains such as hospital and research
laboratories and most industrial laboratories, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V
supply.
2
that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity caused by condensation.
6-4
Model 900 and 5900 Series
990 (990)/5990 (990SS)
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
40.8"W x 77.9" H x 38.9" F-B (103.6cm x 197.9cm x 94.0cm)
30.6"W x 51.5"H x 25.3" F-B (77.7cm x 130.8cm x 64.3cm)
23.0 cu. ft. (651.3 liters)
Two 1 HP (2545 BTUH each)
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 4.5" (11.4 cm) door
120VAC, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
20 Amp, Dedicated Circuit,
20 Amp Time Delay Breaker
900 lbs. (408 kg)
Certifications
Declaration of Conformity is available from the factory
Safety Specifications
Indoor Use Only
Altitude - up to 2,000 meters
Temperature - 5°C to 43°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
Installation Category II
1
2
Pollution Degree 2
Class of Equipment I
Intended Use
This product is intended for use as a General Purpose Laboratory Freezer for storing samples or inventory between
-40 and -86C.
Unintended Use
This unit is not intended for use in an explosive environment, nor to be used for the storage of flammable inventory.
This unit is not intended for use in a Class II medical application as defined by Title 21 of the Federal Code of
Regulations.
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2 assumes
995 (995)/5995 (995SS)
230VAC, 50 Hz, 12.0 FLA
Operating Range: 208VAC-230VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
989 (989)/5989 (989SS)
208-230VAC, 60 Hz, 12.0 FLA
Operating Range: 208VAC-230VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker