wtw SenTix 950 Operating Manual

Ph combination electrodes with refillable liquid reference system

Advertisement

ba75843e01
08/2009
®
SenTix
950
®
SenTix
980
pH combination electrodes with refillable
liquid reference system
Operating manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SenTix 950 and is the answer not in the manual?

Questions and answers

Summary of Contents for wtw SenTix 950

  • Page 1 ® SenTix ® SenTix pH combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009...
  • Page 2 The latest version of the present operating manual can be found on the Internet under www.WTW.com. © Copyright Weilheim 2009, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written authorization of WTW GmbH, Weilheim. Printed in Germany. ba75843e01 08/2009...
  • Page 3: General Information

    The digital transmission technique guarantees the failure-free communi- cation with the meter even with long connection cables. If the sensor firmware is enhanced by WTW, it can be updated via the meter. ba75843e01 08/2009...
  • Page 4: Technical Data

    Technical data SenTix Technical data General data WTW model Reference electrolyte Junction Special features ® SenTix KCl 3 mol/L, Ag free Ceramic Plastic shaft ® SenTix KCl 3 mol/L, Ag free Platinum Measurement and application characteristics WTW model pH measuring range...
  • Page 5 SenTix Commissioning, measuring, calibration Commissioning, measuring, calibration Commissioning The electrode is filled with reference electrolyte solution in the factory. Pre- pare the electrode for measuring as follows: Open the refilling opening for the reference electrolyte solution. Depend- ing on the model, the stopper of the refilling opening is an elastomer stop- per or a slider.
  • Page 6 Storage SenTix trolyte in the electrode must be at least 2 cm above the level of the solu- tion. Reference electrolyte Shaft Junction (Example) Glass membrane Provide approximately the same stirring conditions for measuring as for calibrating. Note Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane.
  • Page 7: Maintenance And Cleaning

    SenTix Aging Note During longer storing periods, salt sediments may develop on the watering cap. They do not affect the measuring characteristics and can easily be re- moved with deionized water when the electrode is put into operation again. Aging Every pH electrode undergoes a natural aging process.
  • Page 8 Maintenance and cleaning SenTix Cleaning Remove water-soluble contamination by rinsing with deionized water. Re- move other contamination as follows: Contamination Cleaning procedure Fat and oil Rinse with water containing household wash- ing-up liquid Lime and hydroxide deposits Rinse with citric acid (10 % by weight) Proteins Immerse in pepsin cleaning solution PEP/pH for approx.
  • Page 9: Wear Parts And Accessories

    109 648 Note Detailed information on our wide range of buffer solutions and more acces- sories is given in the price list of the WTW catalog "Laboratory and field in- strumentation". Disposal When the electrode is at the end of its operational lifetime, please dispose of it as electronic waste and not with unsorted municipal waste.
  • Page 10 Disposal SenTix ba75843e01 08/2009...