Siemens MAMMOMAT 300 Parts Replacement page 12

Compression mechanism and potentiometer r863
Table of Contents

Advertisement

2 - 6
.
Fig. 5
Instead of exchanging the whole Compression mechanism (gear) delivered, the CSE can
decide to insert the spring pin 2.5 x 14 only - if that one was found shorn off. The new
Compression mechanism has nowadays a larger pin mounted, (sized 4 x 14 mm), the
danger of pin getting broken decreased.
M300/1000/3000/3000 Nova
Service
Procedure - Compression mechanism
SPB7-230.841.03
Rev. 02
04.04
Page 6 of 6
CS PS 24
Siemens AG
Medical Solutions