‡ SIGNET 5500 Flow Monitor ENGLISH CAUTION! • Remove power to unit before wiring input and output connections. 3-5500.090-1 H (6/01) English • Follow instructions carefully to avoid personal injury. Contents 8. Output Functions 1. Power Connections 9. Menu Functions 2.
• An external heavy-duty relay must be used for devices with surge de-energized. Power is applied after the relay 2 setpoint is reached. currents or operating currents that exceed the above specifications. ‡ SIGNET 5500 Flow Monitor Pg 3 of 12...
8.1 Sensor pulse output (sec. 3): This output emits an open-collector frequency equal to the sensor input. The output connects directly to a second + GF + SIGNET instrument for expanded process control with a single sensor input (sec. 3.1). The output is an open-collector type that requires an external pull up resistor and power supply for other instrument inputs (sec.
(sec. 9.2 J-M). Relay 2 H. Last calibration: this display shows a user defined setup date for maintenance records. This feature is not an internal timer or calender. Aux output Last calibration A↑ ‡ SIGNET 5500 Flow Monitor Pg 5 of 12...
Page 6
The flow and total K-Factors are independent of each other. The K-Factor settings represent the number of pulses generated by the + GF + SIGNET flow sensor for each engineering unit of fluid measured (published in flow sensor Last calibration Setup date manual).
Page 7
Active high logic: Output changes from low to high VIEW menu: when active, see diagram below. quick quick press press Active Low Logic (factory default) Inactive (Off) Active (On) Active High Logic Active (On) Inactive (Off) ‡ SIGNET 5500 Flow Monitor Pg 7 of 12...
Note: totalizer displays remain undistrubed and will resume after instrument reset. 14. Maintenance Clean the instrument case and front panel with a soft cloth and mild liquid soap solution. ‡ SIGNET 5500 Flow Monitor Pg 11 of 12...