Balluff BTL5-S1 M B-DEXA/B K Series User Manual page 29

Micropulse transducer - rod style
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 24
BTL5-S1_ _-M_ _ _ _-B/Z-DEX_-K_ _/KA_ _
Micropulse Transducer - Rod Style
3
Installation (cont.)
The installation surface of the
BTL housing must lie fully
against the mating surface. The
appropriate O-ring must seal the
bore hole perfectly, i.e. the
chamfer for the O-ring must be
finished according to ➥ ➥ ➥ ➥ ➥ Fig. 3-4.
Suitability of the installed O-ring
(Viton GFLT90) must be verified
for the specific application.
The depth of the threaded hole
must be at least 25 mm.
At least 5 threads must be en-
gaged and there must be a
thread overlap of at least 8 mm.
aaaaaaaaaaaa
➥ ➥ ➥ ➥ ➥
➥ ➥ ➥ ➥ ➥
Observe the prescribed
installation orientation
exactly!
➥ ➥ ➥ ➥ ➥
BTL2-S-6216-8P-Ex
➥ ➥ ➥ ➥ ➥
Float Type
Flameproof enclosure
BTL2-S-5113-4K-Ex
BTL2-S-4414-4Z-Ex
BTL2-S-4414-4Z01-Ex
3
min. density
1 g/cm
Ignition Class "d"
3
(H
O)
0.7 g/cm
2
7

Advertisement

Table of Contents
loading

Table of Contents