Removing The Battery(S) - Invacare Platinum Mobile User Manual

Oxygen concentrator
Hide thumbs Also See for Platinum Mobile:
Table of Contents

Advertisement

flashes to show the battery is charging. When the battery is
fully charged, the fourth segment B of the battery gauge A
stops flashing and remains lit.
Battery Gauge for Uninstalled Battery(s)
When the battery D is removed from the concentrator,
press the battery status button C to show the status of the
charge level on the battery gauge A. Refer to the charge
level B table in Reading the Battery Gauge for Installed and
Uninstalled Battery(s) in Setup.
1194969-C

3.3.4 Removing the Battery(s)

1. Open the battery compartment door F of the
concentrator B.
2. Pull straight up on the lift ring E to remove the battery
C from the battery compartment D.
3. If a second battery is being used (sold separately),
remove it in the same manner.
4. Close the battery compartment door.
A steady audible beep will occur when all power
sources are removed from the concentrator. Refer to
Shut Down Alarms in Troubleshooting.
A battery can be removed without affecting the operation
of the concentrator as long as an external power source or
a second charged battery is installed.
Setup
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Poc1-100bPoc1-100c

Table of Contents