SECTION 5—CABINET
SECTION 5—CABINET

Removing Cabinet

DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
NOTE: For this procedure, refer to FIGURE 5.1.
1. Unplug unit.
2. Remove the four mounting screws that secure cabinet assembly to the base assembly.
3. Lift the cabinet straight up.
NOTE: When required, vacuum inside of the cabinet and exposed foam insulation.
4. To re-install cabinet, reverse STEPS 2-3.
L
Cabinet Assembly
Mounting
Screws
Mounting
Screws
Base Assembly
FIGURE 5.1 Removing Cabinet