5 Setup
Menu
Display
Service
Unit of measurement
Language
Copy sensor data
14
Menu item
Sensor-TAG
Displayed value 1
Scaling
Current output
Reset
Select reset?
In this menu item you select the internal unit of calculation of the sen-
sor: m(d) or ft(d).
Unit of measurement
m(d)
The sensor is already set to the ordered national language. In this
menu item you can change the language. The following languages
are available:
•
Deutsch
•
English
•
Français
•
Espanõl
•
Pycckuu
Language
German
With this function
•
data are read from the sensor
•
Data written into the sensor.
The data are permanently saved in an EEPROM memory in the
display/adjustment module and remain there even in case of voltage
failure. From there, they can be written into one or several sensors
and kept as a backup for a possible sensor exchange. When data is
written into the sensor, the instrument type from which the data origi-
nate as well as the TAG no. of that sensor are displayed.
Copy sensor data
Display and adjustment module for OPTISOUND sensors •
Ultrasonic
Sensor
Distance
0 % = 0.0, 100 % = 100.0
Output mode: 4-20 mA
Failure mode: < 3.6 mA ▼
Min. current: 3.8 mA