All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH. Subject to change without notice. Copyright 2022 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany) www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
4.6 Connecting the relay outputs ..................26 4.7 Connecting the external HART adapter ................. 27 4.8 Power supply ........................28 4.9 Connecting the SHD 200 for use in an active 4...20 mA current output ......29 5 Start-up 5.1 Start-up........................... 31 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 4
7.4.2 Form (for copying) to accompany a returned device............83 7.5 Disposal .......................... 83 7.5.1 Disassembly and recycling ....................84 8 Technical data 8.1 Technical data......................... 87 8.2 Dimensions ........................89 9 Notes www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
• NAMUR recommendation NE 6, NE 21, NE 43, NE 53, NE 107 • Shock resistance IEC 60721-3-4 class 4M4 • Vibration resistance IEC 60721-3-4 class 4M4 • Vibration resistance IEC 61298-3 low vibration control room 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
10/2019 ER 1.0.0_ MA SHD 200 R01 01/2020 ER 1.0.1_ MA SHD 200 R02 05/2020 ER 1.0.2 MA SHD 200 R03 03/2022 ER 2.0.X MA SHD 200 R04 www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose. ® The SHD 200 is a control unit for 4...20 mA/HART field devices. The intended use is process control by using the relays, the second current output and the NAMUR NE 107 signalisation by ®...
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Check for the correct supply voltage printed on the nameplate. Figure 2-1: Example of the standard scope of delivery 1 Ordered device 2 Documentation Optional accessories ® • External HART adapter • Weather protection cover • Wall mounting brackets 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Modem to the internal or optional ® external HART adapter. For Krohne level and pressure field devices, several functions are available to perform the commissioning directly via the SHD 200. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Electronic / Electric device waste marking, 6 Power supply data, Parameter of measuring unit 7 Order code, Manufacturing date, Internet address of manufacturer 8 Electronic revision, Serial number 9 Production order 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
• The original packing is designed to protect the equipment. It has to be used if the device is transported or sent back to the manufacturer to prevent damage of the device. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
6 HART field device 7 Power supply for devices (slaves) with passive current output 8 Load 250 (Ohm) (only required if not used in SHD 200) ® If using more than one HART master please be aware of the correct configuration because only one Primary master and one Secondary master are allowed.
HART functionality of the SHD 200 must be deactivated ® • use a HART PLC, find out its master mode and adapt the master mode of the SHD 200 appropriate 3.4 Pre-installation requirements CAUTION! Always note the following items to ensure a proper and safe installation: •...
• Put a small flat tool (e.g. screwdriver) into the right hinge lock and open the device 1. • Lift up the housing 2. Now you have access to the electronic compartment and the possibility to mount the device on a wall. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 18
Now you have access to the electronic compartment and the possibility to mount the device on a wall. • For closing the housing again, use the 2 torx screws with a torque of 1,5 Nm. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
For outdoor mounting, we recommend the optional weather protection cover XGSH010000 for wall and pipe mounting. For more flexibility and convenient mounting, we recommend wall mounting brackets XGSH010020 for plastic housing. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
For outdoor mounting, we recommend the optional weather protection cover XGSH010000 for wall and pipe mounting. For more flexibility and convenient mounting, we recommend wall mounting brackets XGSH010030 for aluminium housing. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Check for the correct supply voltage printed on the nameplate. 4.2 Important device-specific notes on electrical connection INFORMATION! The SHD 200 is a loop powered device and has no function if no field device is connected. INFORMATION! Use only shielded cables for connection with the control system (e.g. PLC).
The device is equipped with an internal HART access for the NonEx-version (A2 and B2) that ® ® offers the opportunity to access the field device via HART handheld or HART modem. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 23
3 Power supply for loop 1. Main loop for operating the device, in this case with usage of internal resistor. 4 Power supply for relay 1 5 Power supply for relay 2 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
M12 and 5...10 mm for M16. If you are using cables with a larger diameter than 3...6,5 mm for M12 and 5...10 mm for M16, you must replace the manufacturer’s cable glands with suitable ones. The operator is responsible for the correct sealing of cable glands. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
The second current output can be either used as current output for any HART® variable or as power supply for backlight and LEDs for NAMUR NE 107 signalisation if supported by field device (field device must support extended command 523: Read Condensed Status Mapping Array). 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
2 Remove the middle M16 cable gland and screw the HART adapter. ® 3 Connect the HART adapter to connections A2 and B2. Consider that the cap needs to be closed for a protection of the electronics parts. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
2-wire interface (4...20 mA). Sens+ Sens- K FE K FE A1 B1 A2 B2 A1 B1 A2 B2 Figure 4-8: Connection of the power supply www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
ELECTRICAL CONNECTIONS SHD 200 4.9 Connecting the SHD 200 for use in an active 4...20 mA current output DANGER! To avoid fatal injuries as well as destruction or damage of the device assure a correct installation before switching on the power. This includes: •...
Page 30
25 , connecting the relays refer to Connecting the relay outputs on page 26 and connecting the external HART adapter refer to Connecting the external HART adapter on page www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Polling address to 0…63 (see function table of A3.2) or set a user specific range (A.3.4). To change to 4…20 mA mode, go to menu option A2. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 32
START-UP SHD 200 When search was successful, the display shows the KROHNE product name of the field device. HART devi ce f ound . . . KROHNE SM ARTPAT PH Figure 5-2: Product name of the field device www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
• Each time you press a key, an "" flashes in the upper right corner of the display. • After 5 minutes of inactivity, there is an automatic return to the measuring mode. In this case the device does not save previously changed data! 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
C1.6 Tank type C1.7 Tank height C1.8 Current out 1 var. C1.9 0% range C1.10 100% range C1.11 Current out. range C1.12 Error function C2 Empty spectrum C2.1 Record spectrum www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 35
E3 Relay E3.1 Relay 1 E3.2 Relay 2 F Service This menu is password protected and contains functions to be used by service personnel only. Table 6-1: Menu mode structure 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Device is being operated out of specification. Check device and/or process. specification Maintenance Measurements might not be accurate anymore. Maintenance is needed. required Diagnosis OK The device is working properly. Table 6-2: NAMUR NE 107 signalisation www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 37
(8th byte of cmd 48) Electronic defect The field device connected Check field device and replace it if necessary. ® via HART reports an electronic defect (8th byte of cmd 48) 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 38
Discrete var sim The field device connected Check field device active via HART reports an active configuration. simulation mode for one or more discrete vars (9th byte of cmd 48) www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 39
(10th byte of cmd 48) Status simulation The field device connected Check field device active configuration. ® via HART reports an active status simulation (9th byte of cmd 48) 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 40
The field device connected ® Check HART configuration duplicate ID ® via HART reports that a of the field device. duplicate unique ID /long tag device has been deteced(11th byte of cmd www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
A3.4 User specific range ® ® User specific search range for HART devices (only displayed if HART mode is selected in A2 and "User specific range" is selected in A.3.2) 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 42
Settings of main value (only displayed if 4…20 mA mode is selected in A2) Options: • 4…20 mA (L1) • User A4.2 2. Measurement page ® Configuration of 2. measurement page (only displayed if HART mode is selected in A2) www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 43
Value line one ® Selection of first displayed measuring value (only displayed if HART mode is selected in A2) Options: • • • • • 4…20 mA (L1) • 4…20 mA (L2) 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 44
Options: • 4…20 mA (L1) • User A4.4.3 Select range ® Adjustment of displayed measuring range (only displayed if HART mode is selected in A2) Options: • Automatic • Manual www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 45
Setting of timespan (only activated if HART mode is selected in A2) Options: • 15 min • 30 min • 60 min • • 12 h • 24 h • • 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 46
Range: 002…006 A4.5.8 Decimals QV ® Settings of decimals for QV (only displayed if HART mode is selected in A2) Options: • Automatic • #N.XXXX • #N.XXX • #N.XX • #N.X www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 47
#N.XXX • #N.XX • #N.X A4.6 Illumination settings ® Illumination Settings (only displayed if HART mode is selected in A2 and Current output or Power supply is selected in A5.3.1). 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 48
Settings of Loop 1 (only displayed if 4…20 mA mode is selected in A2) A5.1.1 Unit Unit of displayed measuring value (only displayed if 4…20 mA mode is selected in A2) www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 51
® Time of deviation, till error message displays (only displayed if HART mode is selected in A2 and monitoring is activated in A5.2.2.1) A5.3 Loop 2 Setting for Loop 2 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 52
20 mA Trimming ® 20 mA Trimming (only displayed if HART mode is selected in A2 and Current Output is selected in A5.3.1) A5.4 Relay 1 Settings for Relay 1 www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 53
A2 and Limit Switch (Band) is selected in A5.4.1) A5.4.10 Set Low Alarm Hyst. Settings of the lower switching threshold (only displayed if 4…20 mA mode is selected in A2 and Limit Switch (Band) is selected in A5.4.2) 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 54
Settings of the Loop 1 error current, at which the relay switches (only ® displayed if HART mode is selected in A2 and Loop 1 error current is selected in A5.4.1) Options: • High • • Both www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 55
Start manual cleaning? (only displayed if Cleaning Mode is selected in A5.4.1 Options: • • A5.4.34 Stop cleaning Stop cleaning? (only displayed if Cleaning Mode is selected in A5.4.1 Options: • • A5.5 Relay 2 Settings for Relay 2 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 56
A2 and Limit Switch (Band) is selected in A5.5.1) A5.5.14 Set high alarm hyst. Settings of the upper hysteresis value (only displayed if 4…20 mA mode is selected in A2 and Limit Switch (Band) is selected in A5.5.2) www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 57
Order code of the device A6.1.1.3 CG number Number of CG A6.1.1.4 Device name Device name A6.1.1.5 Electronic revision Electronic Revision A6.1.1.6 Software version Software version A6.1.1.7 Display brd. serialnum. Display board serial number 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 58
Setting of tank type (only displayed if OPTIWAVE level transmitter is connected; ) Options (displayed selection depends on OPTIWAVE level transmitter type): • Storage • Process • Agitator • Stillwell • Flat surface • Medium slope • Steep slope www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 59
Partial, Average (if tank can not be discharged and/or device is close to a product inlet) • Partial, Max. (if tank can not be discharged and/or device is close to a product inlet) 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 60
Setting of damping value (only displayed if OPTIBAR transmitter is connected) Range: 000.0…999.0 Set unit Setting of units (only displayed if OPTIBAR transmitter is connected) D5.1 Pressure unit Setting of pressure units (display depends on connected OPTIBAR transmitter) www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 62
• Linear • Square root • Free prog. • Horiz. round tank • Spherical tank • Bidirect. linear • Bidirect. square root Test Simulation menu System Simulation of process variables www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Figure 6-1: 1. Measurement page 1 TAG 2 Measuring unit 3 Relay 2 status 4 Relay 1 status 5 Measuring value The 1. Measurement page shows one large value with the corresponding unit. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 64
This 2. measurement page shows two values with the corresponding unit. For the displayed values you can chose between PV, SV, TV, QV, 4…20 mA (L1) in menu option A4.2. Those options are for both variables. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 65
The number of values is freely configurable. The status of the relays is displayed on the right side in this case. Go to menu option A4.3 to choose between the parameters PV, SV, TV, QV, 4…20 mA (L1) and 4…20 mA (L2) for each value. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 66
In manual mode, you can define the lower range and the upper range value. Decide, wether the limit lines of relay 1 and relay 2 should be displayed. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 67
Press > to get detailed information about the displayed error and leave that page by pressing ^. Press or to leave the status page to the measurement pages for entering the menu. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Press or until the submenu Trimming 20mA is highlighted and press > to enter the submenu. Define the 20mA Trimming and press ^. Press ^ several times to save the configuration and to return to the measuring mode. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Cleaning mode, System alarm, Loop 1 error current, USP<645> and Field device status. Limit switch (threshold) In Limit Switch (Threshold) mode, you define a threshold. When the measuring value exceeds or falls below this threshold, the relay switches. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 72
Press or until the submenu Set high alarm hysteresis is highlighted and press > to enter the submenu. Define the value and press ^. Press ^ several times to save the configuration and to return to the measuring mode. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 73
Press or until the submenu Set factor for USP<645> is highlighted and press > to enter the submenu. Define the value and press ^. Press ^ several times to save the configuration and to return to the measuring mode. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 74
Choose between Maintenance requir., Failure, Check function, Out of specification. The relay switches when the field device transmits the selected error category and press ^. Press ^ several times to save the configuration and to return to the measuring mode. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 75
Press or until the submenu Stop cleaning is highlighted and press > to enter the submenu. Choose between Yes and No to stop cleaning manually or not and press ^. Press ^ several times to save the configuration and to return to the measuring mode. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 76
Press or until the submenu Set characteristic is highlighted and press > to enter the submenu. Choose between Overrun and Underrun to define when the relay should switch and press ^. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Choose between Yes and No when the LEDs are required or not and press ^. Press ^ several times to save the configuration and to return to the measuring mode. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Figure 6-10: Overflow sign Exponential format If the format of the related display line is set to "automatic" in menu A4.5 long measuring results appear in an exponential format: Figure 6-11: Exponential format www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
The device can then use this data to put the measurement signal through a filter (empty spectrum). For more detailed information about the different setting options see the transmitter manual. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
PM 5060, OPTIBAR PC 5060, OPTIBAR DP 7060. Depending on the transmitter type, the setting options in the table below may vary. For more detailed information about the different setting options see the transmitter manual. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 81
Press ^ twice after density unit settings and choose Set transfer functionwith or . Press > to enter Set transfer function. Choose type of transfer function with or and confirm adjusted type with ^. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
• to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from such dangerous substances, • to enclose a certificate with the device confirming that it is safe to handle and stating the product used. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative. 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
(EOL) or is disposed of after usage. The information given is sufficient to gather the most important parts of the device (by the end-user) which can be used for recycling. Figure 7-1: Overview disassembly - aluminium housing www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Page 85
52 / 0.12 52 / 0.12 size<82 cm PCB basic 20 / 0.04 20 / 0.04 cover, size<81 Table 7-1: Disassembly and recycling No battery and no capacitors >25 mm included 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 86
• Open the housing with a flat tool(plastic version) or with a torx screwdriver T20 (aluminium version) • Disconnect all cables • Remove the circuit boards All main parts are now disassembled and can be shipped separately for reuse and/or recycling. www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
Storage -30...+70°C / -22...+158°F Altitude 2000 m / 6561.68 f Other conditions Humidity RH4...100% rH (IEC 60721-3-4:1995 Class 4K4) Ingress protection IP66/ IP67 acc. to IEC 60529: 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 88
NAMUR recommendation NE 6, NE 21, NE 43, NE 53, NE 107 Shock resistance IEC 60721-3-4 class 4M4 Vibration resistance IEC 60721-3-4 class 4M4 IEC 61298-3 low vibration control room www.krohne.com 07/2022 - 4007792904 - MA SHD 200 R04 en...
NOTES SHD 200 07/2022 - 4007792904 - MA SHD 200 R04 en www.krohne.com...
Page 92
Engineering, commissioning, calibration, maintenance and training services Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.: +49 203 301 0 Fax: +49 203 301 10389 info@krohne.de The current list of all KROHNE contacts and addresses can be found at: www.krohne.com...