KROHNE Display and adjustment module for OPTISOUND sensors Operating Instructions Manual page 11

Table of Contents

Advertisement

Edit sensor TAG
Peak value indicator
Meas. reliability
Sensor status
Display and adjustment module for OPTISOUND sensors •
The default setting is linear.
Linearisation curve
Linear
In the menu item "Sensor-TAG" you edit a 12-digit measurement loop
name. An unambiguous designation can hence be assigned to the
sensor, e.g. the measurement loop name or the tank or product des-
ignation. In digital systems and in the documentation of larger plants,
a singular designation should be entered for exact identification of
individual measuring sites.
The available digits comprise:
Letters from A ... Z
Numbers from 0 ... 9
Special characters +, -, /, -
Factory setting is "Sensor".
Sensor-TAG
Sensor
Menu section, diagnostics
The respective min. and max. measured values are saved in the sen-
sor. The values are displayed in the menu item "Peak values".
Min. and max. distance in m(d)
Min. and max. temperature
Peak value indicator
When non-contact level sensors are used, the measurement can be
influenced by the respective process conditions. In this menu item,
the measurement reliability of the level echo is displayed as dB value.
The measurement reliability equals signal strength minus noise. The
higher the value, the more reliable the measurement. With a function-
ing measurement, the values are > 10 dB.
The instrument status is displayed in this menu item. If no failure is
detected by the sensor, "OK" will be displayed. If a failure is detected,
there will be a sensor-specific flashing fault signal, for example
"E013". The failure is also displayed in clear text, for example "No
measured value available".
Information:
The fault message as well as the clear text indication are also carried
out in the measured value display.
5 Setup
11

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Display and adjustment module for OPTISOUND sensors and is the answer not in the manual?

Table of Contents