Disk Drive Test (if installed)
1. Insert a blank, unformatted floppy disk into the disk drive.
2. From the start-up screen, press the Mass Storage softkey.
3. Press the Next Menu softkey twice to go to the Format menu.
4. Press the Format softkey to format the floppy disk. (Note that if internal RAM is
5. A prompt displays, asking you to confirm the format operation. Press the Yes
6. After the format is complete, remove the disk from the drive and insert it into the
7. At the PC, type the following command:
8. Verify that the response on the PC screen reads:
9. Remove the disk from the PC drive and return it to the TFP2A.
10. Return to the FiberMaster start-up screen by cycling power to the OTDR.
11. Enter the Run Diagnostics menu by pressing the Change Settings softkey and the
12. Select the mass storage tests by turning the
13. Run the mass storage tests by pressing the START TEST softkey.
14. Verify successful completion of the mass storage tests by pressing the More Info
15. Press the OK softkey to return to the Run Diagnostics menu, then Exit to return to
TFP2A FiberMaster Module-Level Service Manual
installed in your instrument you must press the Format Removable softkey to
format the floppy disk.)
softkey.
floppy disk drive of an IBM PC or compatible.
dir <drive designator>:
In place of <drive designator>, type the letter designator of the floppy disk drive,
usually a or b.
Volume in drive <drive designator>: is FIBERMASTER
If the PC does not display the statement above, there is a formatting problem with
FiberMaster's disk drive.
button simultaneously.
HELP
highlighted. Then press the Select Test softkey.
softkey.
the start-up screen.
Chapter 4: Maintenance
knob until the test is
CURSOR
4-15