Mounting And Connecting Connectors (When Using T Type); Mounting Nq To Target System - Renesas QB-RL78F14 User Manual

In-circuit emulator
Table of Contents

Advertisement

QB-RL78F14 In-Circuit Emulator

2.4 Mounting and Connecting Connectors (When Using T Type)

This section describes the methods of connecting the QB-RL78F14 and target system.
Make connections with both the QB-RL78F14 and target system powered OFF.
The following abbreviations are used in this section:
- NQ: Target connector
- YQ: YQ connector
- EA: Exchange adapter
- MA: Mount adapter
- CA: Check pin adapter
- SA: Space adapter

2.4.1 Mounting NQ to target system

(1) Thinly apply a two-component epoxy adhesive (hardening time at least 30 minutes) to the ends of the four
projections on the base of the NQ and adhere the NQ to the user board (clean the surface of the target system
board using alcohol or the like). If alignment of target system pads to NQ leads is difficult, align them as in (2).
(2) Align by inserting the guide pins for alignment for the NQ (NQGUIDE) through the pin holes on the top of the
NQ. Accessory holes are 1.0 mm non-through holes in two or three places.
(For hole positions, see the particular NQ drawing.)
(3) Solder after fitting the MA to the NQ. This is to prevent troubles such as flux or solder splashing and adhering
to the NQ contact pins when soldering.
- Soldering conditions
Caution Do not perform washing by flux immersion or vapor.
(4) Take away the guide pins.
R20UT2058EJ0300 Rev.3.00
Oct 31, 2013
260°C  10 seconds or less
Solder reflow
350°C  5 seconds or less (1 pin)
Manual soldering