Knick Stratos Pro A201MSCONDI User Manual

Knick Stratos Pro A201MSCONDI User Manual

Conductivity measurement with toroidal sensors via memosens protocol
Table of Contents

Advertisement

Quick Links

®
User Manual
Read before installation.
Keep for future use.
Stratos Pro A201MSCONDI
Conductivity Measurement
with Toroidal Sensors via Memosens Protocol
www.knick.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Stratos Pro A201MSCONDI and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Knick Stratos Pro A201MSCONDI

  • Page 1 ® Stratos Pro A201MSCONDI Conductivity Measurement User Manual with Toroidal Sensors via Memosens Protocol Read before installation. Keep for future use. www.knick.de...
  • Page 2: Supplemental Directives

    Supplemental Directives Read this document and retain it for future reference. Before assem- bling, installing, operating, or maintaining the product, ensure that you fully understand the instructions and risks. Observe all safety instructions. Failure to follow the instructions in this document may result in serious injury and/or property damage.
  • Page 3: Documents Supplied

    Quickstart Guides Installation and first steps: • Operation • Menu structure • Calibration • Error messages and recommended actions Test Report 2.2 According to EN 10204 Electronic Documentation www.knick-international.com: Manuals + software Ex devices: Control Drawings EU Declaration of Conformity...
  • Page 4: Table Of Contents

    Supplemental Directives ............... 2 Documents Supplied ..............3 Safety ....................7 Intended Use ..................7 Introduction ..................9 Overview of Stratos Pro A201MSCONDI ........12 Assembly ..................13 Package Contents ................13 Mounting Plan, Dimensions ............14 Pipe Mounting, Protective Hood ..........15 Panel Mounting ..................16 Electrical Installation ..............17...
  • Page 5 Table of Contents Configuration ................32 Menu Structure of Configuration..........32 Manual Switchover of Parameter Sets A/B .......34 Configuration (Template for Copy) ..........39 Sensor .....................42 Sensor Verification (TAG, GROUP) ..........50 Current Output 1 ................52 Current Output 2 ................62 Temperature Compensation ............64 CONTROL Input (TAN SW-A005) ...........68 Alarm Settings ..................70 Time and Date ..................74 Measuring Points (TAG/GROUP) ..........76...
  • Page 6 Table of Contents Product Line and Accessories ..........102 Specifications ................103 Calibration Solutions ..............109 Concentration Measurement ...........111 Concentration Curves ..............112 Error Handling ................117 Error Messages ................118 Decommissioning ...............121 Disposal ....................121 Returns ....................121 Sensoface ..................122 HART: Typical Applications ............124 FDA 21 CFR Part 11 ..............125 Electronic Signature –...
  • Page 7: Safety

    Safety Always Read and Observe the Safety Instructions! The device is constructed in accordance with the latest technology and generally accepted safety rules and regulations. Under certain circumstances, however, usage may pose risks to users or cause damage to the device. Commissioning must be carried out by specialist personnel authorized by the operating company.
  • Page 8 Safety Function Check Mode (HOLD Function) After activating configuration, calibration, or service, Stratos Pro enters function check mode (HOLD). The current outputs respond in accordance with the configuration. Operations must not be carried out while Stratos Pro is in function check (HOLD) mode, as the system may behave unexpectedly and put users at risk.
  • Page 9: Introduction

    M12 socket accessory (ZU0822) instead of a cable gland – which allows simple replacement of the sensor cable without opening the device. Sensors The device has been designed for application with the Model SE670 and SE680 toroidal sensors (Knick).
  • Page 10 Introduction Display Plain-text messages on a large, backlit LC display enable intuitive op- eration. You can specify which values are to be displayed in standard measuring mode (“Main Display, ” see page 25). Color-Coded User Interface The colored display backlighting indicates different operating states (e.g., alarm: red, HOLD mode: orange;...
  • Page 11 Introduction Control Inputs (TAN SW-A005) I input Current The analog (0) 4 ... 20 mA current input can be used for external temperature compen- input sation; see page 64. HOLD HOLD (Floating digital control input) The HOLD input can be used for external input activation of HOLD mode;...
  • Page 12: Overview Of Stratos Pro A201Mscondi

    Overview Overview of Stratos Pro A201MSCONDI Power supply + Output 1, 2 / HART Output 1 RS-485 + 3 V RS 485 A – Output 1 / HART RS 485 B GND/Shield Output 2 – Output 2 Current Input + Input –...
  • Page 13: Assembly

    Assembly Package Contents Note: Check all components for damage upon receipt. Do not use damaged parts. The package should contain: • Front unit, rear unit, bag containing small parts • Specific test report • Documentation (see page 3) Fig.: Assembling the enclosure 1) Insertable jumper (3x) 6) Blanking plug (2x, non-Ex 2) Plate (1x), for conduit mount-...
  • Page 14: Mounting Plan, Dimensions

    Assembly Mounting Plan, Dimensions 1) Cable gland (3 x) 2) Knockouts for cable gland or ½” conduit, 21.5 mm dia. (2 knockouts) Conduit couplings not included! 3) Knockout for pipe mounting (4 x) 4) Knockout for wall mounting (2 x) Fig.: Mounting plan (All dimensions in mm!)
  • Page 15: Pipe Mounting, Protective Hood

    Assembly Pipe Mounting, Protective Hood ø40...ø60 1) Hose clamp with worm gear drive to DIN 3017 (2 x) 2) Pipe-mount plate (1 x) 3) For vertical or horizontal posts or pipes 4) Self-tapping screw (4 x) Fig.: Pipe-mount kit, accessory ZU 0274 (All dimensions in mm!) Fig.: Protective hood for wall and pipe mounting, accessory ZU 0737 (All dimensions in mm!)
  • Page 16: Panel Mounting

    Assembly Panel Mounting < 30 1) Circumferential sealing (1 x) 2) Screws (4 x) 3) Position of control panel 4) Span piece (4 x) 5) Threaded sleeve (4 x) Cutout 138 x 138 mm (DIN 43700) 1...22 Fig.: Panel-mount kit, accessory ZU 0738 (All dimensions in mm!)
  • Page 17: Electrical Installation

    Electrical Installation Before commencing with the installation, make sure that all lines to be connected are de-energized. Observe the safety instructions; see page 7. Cable Glands In a hazardous location, only cable glands with suitable approvals may be used. The installation instructions of the manufacturer must be observed.
  • Page 18: Rating Plates / Terminal Assignments

    Electrical Installation Rating Plates / Terminal Assignments Fig.: Terminal assignments of Stratos Pro A201 Fig.: Stratos Pro A201N rating plate at outside bottom of front (illustrative example)
  • Page 19: Wiring Of Stratos Pro A201Mscondi

    Electrical Installation Wiring of Stratos Pro A201MSCONDI Areas for placing the screwdriver to pull out the terminals Terminal row 1 Terminal row 2 In addition: 2 HART pins (between terminal row 1 and 2) Sensor connection RS-485 Fig.: Terminals, device opened, back of front unit...
  • Page 20: Wiring Example: Se670, Se680

    Wiring Example: SE670, SE680 Measuring task: Conductivity, temperature Sensor: SE670, SE680 NOTICE! Connection to RS-485 interface! The SE670 / SE680 sensors are connected to the RS-485 interface of the device – for an A2... Series (2-wire) device, the measuring module slot must be empty.
  • Page 21: User Interface, Keypad

    User Interface, Keypad MEMO SENS 1 Display 2 Keypad 3 Nameplate (enclosure bottom) Function • Return to last menu level • Directly to measuring mode (press > 2 s) • Measuring mode: other display • Retrieve information • Show error messages •...
  • Page 22: Display

    Display MEMO SENS 1 Temperature 13 Info available 2 Sensocheck 14 Hold mode active 3 Interval/response time 15 Main display 4 Sensor data 16 Secondary display 5 Not used 17 Proceed using enter 6 Limit message: 18 Not used Limit 1 or Limit 2 19 Diagnostics 7 Alarm...
  • Page 23: Measuring Mode

    Measuring Mode After the operating voltage has been connected, the analyzer auto- matically goes to “Measuring“ mode. To call the measuring mode from another operating mode (e.g. Diagnostics, Service): Hold meas key depressed (> 2 s). Sensoface indicator Active (sensor status) parameter set (configuration) Time...
  • Page 24: Selecting The Mode / Entering Values

    Selecting the Mode / Entering Values To select the operating mode: 1) Hold meas key depressed (> 2 s) (directly to measuring mode) 2) Press menu key: the selection menu appears 3) Select operating mode using left / right arrow key 4) Press enter to confirm the selected mode Selection menu Selected mode...
  • Page 25: Display In Measuring Mode

    Display in Measuring Mode The MAIN DISPLAY is the display which is shown in measuring mode. To call the measuring mode from any other mode, hold the meas key depressed for at least 2 sec. meas key enter key meas By pressing meas briefly you can step through further displays such as tag number (TAG) or flow (L/h).
  • Page 26: Color-Coded User Interface

    Color-Coded User Interface The color-coded user interface* guarantees increased operating safety. Operating modes are clearly signaled. The normal measuring mode is white. Information text appears on a green screen and the diagnostic menu appears on turquoise. The orange HOLD mode (e.g. during calibration) is quickly visible as is the magenta screen which indicates asset management messages for predictive diagnostics –...
  • Page 27: Operating Modes

    Operating Modes Diagnostics Display of calibration data, display of sensor data, performing a device self-test, viewing the logbook entries, display of hardware/software versions of the individual components. The logbook (TAN SW-A002) can store 100 events (00...99). They can be displayed directly on the device.
  • Page 28: Menu Structure Of Modes And Functions

    Menu Structure of Modes and Functions Meas. mode TAG display CLK display (main display selectable) after 60 s after 60 s Pressing the menu key (down arrow) opens the selection menu. Select the menu group using the left/right arrow keys. ...
  • Page 29: Hold Mode

    HOLD Mode The HOLD mode is a safety state during configuration and calibration. Output current is frozen (LAST) or set to a fixed value (FIX). The HOLD mode is indicated by orange display backlighting. HOLD mode, display icon: Output signal response •...
  • Page 30: Alarm

    Alarm External activation of HOLD (SW-A005) The HOLD mode can be activated from outside by sending a signal to the HOLD input (e.g. from the process control system). Power supply HOLD 12...24 V AC/DC input Stratos Pro A201 Process control system HOLD inactive 0...2 V AC/DC HOLD active...
  • Page 31: Alarm And Hold Messages

    Alarm and HOLD Messages Message Released by Cause Alarm Sensocheck Polarization / Cable (22 mA) Error Messages Flow (CONTROL input) ERR 10: Conductance > 3500 mS HOLD HOLD via menu or input (Last/Fix) Configuration Calibration Service Generating a message via the CONTROL input (TAN SW-A005) (min.
  • Page 32: Configuration

    Configuration CAUTION! Incorrect parameter settings or adjustments can result in incorrect outputs. Stratos Pro must therefore be commissioned by a system specialist, all its parameters must be set, and it must be fully adjusted. For detailed information on parameter setting and adjustment, see the user manual Menu Structure of Configuration The device provides 2 parameter sets “A“...
  • Page 33 Configuration Parameter Set A/B: Configurable Menu Groups Menu group Parameter set A Parameter set B Sensor selection Current output 1 Current output 1 Current output 2 Current output 2 Compensation Compensation Control input Alarm mode Alarm mode Parameter set selection Setting the clock TAG of measuring point TAG of measuring point...
  • Page 34: Manual Switchover Of Parameter Sets A/B

    Configuration Manual Switchover of Parameter Sets A/B Display Action Remark To switch between Manual selection of parameter sets must have parameter sets: been preset in CONFIG Press meas. mode. Default setting is a fixed parameter set A. Wrong settings change the measurement properties! PARSET blinks in the...
  • Page 35 Configuration Configuration Choices Default Sensor (SENSOR) SNS: MEMOSENS MEMOSENS SE 680-M SE 670-K SE 670 MEAS MODE Cond Cond Conc % Sal ‰ Cond MEAS RANGE xxx.x μS/cm x.xxx mS/cm x.xxx mS/cm xx.xx mS/cm xxx.x mS/cm x.xxx S/m xx.xx S/m Conc Solution -01- (NaCl)
  • Page 36 Configuration Configuration Choices Default Output 1 (OUT1) CHANNEL Cond/TMP Cond OUTPUT (with Cond only) LIN / BiLIN / LOG BEGIN 4mA xxxx 000.0 mS/cm END 20 mA xxxx 100.0 mS/cm BiLIN BEGIN 4 mA END 20 mA CORNER X Input range: selected CHANNEL Vertex X : BEGIN ≤...
  • Page 37 Configuration Configuration Choices Default Output 2 (OUT2) CHANNEL Cond/TMP ... other steps like output 1 Temperature compensation (CORRECTION) Compensation for ultrapure water: NaCl, HCL, NH3 TC LIQUID 00.00 ...19.99%/K 00.00%/K REF TEMP 000.0 ... 199.9 °C 025.0 °C TEMP EXT ON/OFF I-INPUT 0–20 mA / 4–20 mA...
  • Page 38 Configuration Configuration Choices Default Parameter set (PARSET) Select fixed parameter set (A) PARSET FIX / or switch between A/B via CNTR INPUT / (fixed parameter control input or manually in MANUAL set A) measuring mode Real-time clock (CLOCK) CLK: FORMAT 24 h / 12 h 24 h 24 h...
  • Page 39: Configuration (Template For Copy)

    Configuration (Template for Copy) Two complete parameter sets are stored in the EEPROM. As delivered, the two sets are identical but can be edited. Note: Fill in your configuration data on the following pages or use them as original for copy.
  • Page 40 Configuration (Template for Copy) Parameter Parameter set A Parameter set B SNS: Sensor type SNS: Measuring mode SNS: Measuring range SNS: Concentration determination SNS: Temperature unit SNS: Temp detection SNS: Manual temp SNS: CIP counter SNS: SIP counter SNS: CHECK TAG SNS: CHECK GROUP OT1: Process variable OT1: Lin/bilin/log output...
  • Page 41 Configuration (Template for Copy) Parameter Set A Set B OT2: Process variable OT2: Lin/bilin/log output OT2: Current start OT2: Current end OT2: Vertex X (bilinear curve only) OT2: Vertex Y (bilinear curve only) OT2: Filter time OT2: FAIL 22 mA (error messages) OT2: FACE 22 mA (Sensoface messages) OT2: HOLD mode OT2: HOLD FIX current...
  • Page 42: Sensor

    Configuration Sensor Select: Sensor type, measuring mode, range 1) Press menu key. 2) Select CONF using , press enter. 3) Select parameter set using keys, press enter. 4) Select SENSOR menu using keys, press enter. 5) All items of this menu group are indicated by the “SNS:”...
  • Page 43 Configuration Menu item Action Choices Sensor type MEMOSENS Select sensor type using  keys. SE 680-M SE 670-K Press enter to confirm. SE 670 Measuring mode Cond Select desired measuring mode using  keys. Conc % Sal ‰ Press enter to confirm. Measuring range For cond measurement xxx.x μS/cm...
  • Page 44 Configuration Sensor Select: Concentration determination 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select SENSOR menu using  keys, press enter. 5) All items of this menu group are indicated by the “SNS:”...
  • Page 45 Configuration Menu item Action Choices Concentration For conc measurement -01- (NaCl), -02- (HCl), only -03- (NaOH), -04- (H determination Select desired concentra- -05- (HNO ), -06- (H tion solution using . -07- (HCl), -08- (HNO -09- (H ), -10- (NaOH), Press enter to confirm.
  • Page 46 Configuration Sensor Select: Temperature unit, temperature detection 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select SENSOR menu using  keys, press enter. 5) All items of this menu group are indicated by the “SNS:”...
  • Page 47 Configuration Menu item Action Choices Temperature unit Select °C or °F using  keys. Press enter to confirm. Temperature Select mode using : AUTO: Measured by detection sensor MAN: Direct input of temperature, no measure- ment (see next step) EXT: Temperature speci- fied via current input (only if TAN E enabled) Press enter to confirm.
  • Page 48 Configuration Sensor Adjust: Cleaning cycles, sterilization cycles 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select SENSOR menu using  keys, press enter. 5) All items of this menu group are indicated by the “SNS:”...
  • Page 49 Configuration Menu item Action Choices CIP / SIP Cleaning cycles Select ON or OFF using ON/OFF  keys. Activates/deactivates log‑ ging in extended logbook (TAN SW‑A003). Press enter to confirm. Sterilization cycles Select ON or OFF using ON/OFF  keys. Activates/deactivates log‑...
  • Page 50: Sensor Verification (Tag, Group)

    Configuration Memosens Sensor Sensor Verification (TAG, GROUP) 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select SENSOR menu using  keys, press enter. 5) All items of this menu group are indicated by the “SNS:”...
  • Page 51 Configuration Menu item Action Choices Select ON or OFF using ON/OFF  keys. Press enter to confirm. When switched on, the entry for “TAG” in the Memosens sensor is com- pared to the entry in the analyzer. If the entries differ, a mes- sage will be generated.
  • Page 52: Current Output 1

    Configuration Current Output 1 Output current range. Linear/Logarithmic. Current start. 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select OUT1 menu using  keys, press enter. 5) All items of this menu group are indicated by the “OT1:”...
  • Page 53 Configuration Menu item Action Remark Process variable Selectable decades with Select using  keys: logarithmic setting (LOG): Cond: Conductivity S/cm: 0.001 mS/cm, 0.01 mS/cm, TMP: Temperature 0.1 mS/cm, 1.0 mS/cm, 10.0 mS/ Press enter to confirm. cm, 100.0 mS/cm, 1000 mS/cm Then select characteristic S/M: 0.001 S/m, 0.01 S/m, 0.1 S/m, 1.0 S/m, 10.0 S/m, 100 S/m...
  • Page 54 Configuration Current Output 1 Output current curve, bilinear 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select OUT1 menu using  keys, press enter. 5) All items of this menu group are indicated by the “OT1:”...
  • Page 55 Configuration Menu item Action Choices Output current Select using  keys. Linear characteristic Press enter to confirm. curve biLIN Bilinear curve Logarithmic curve Current start Entered value applies to Enter value using  selected process variable/  keys. and current end range If the adjusted range is exceeded, the device...
  • Page 56 Logarithmic Curve Nonlinear output current characteristic: allows measurements over several decades, e.g. measuring very low values with a high resolution and high values with a low resolution. Parameters required: Start and end value Possible start and end values The start value must be at least one decade lower than the end value. Start value and end value must be specified in the same units (either in µS/cm or in S/m, see listing): 0.001 mS/cm...
  • Page 57 Configuration Menu item Action Choices Logarithmic curve Select using  keys. Logarithmic curve Press enter to confirm. of output current biLIN Bilinear curve Linear characteristic Start value Start value of logarithmic Enter value using  output curve  keys. Press enter to confirm. End value End value of logarithmic Enter value using ...
  • Page 58 Configuration Current Output 1 Adjusting the time interval of the output filter 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select OUT1 menu using  keys, press enter.
  • Page 59 Configuration Menu item Action Choices Time averaging filter 0...120 SEC Enter value using  (0000 SEC)  keys. Press enter to confirm. Time averaging filter To smoothen the current output, a low-pass filter with adjustable filter time constant can be switched on. When there is a jump at the input (100 %), the output level is at 63 % after the time interval has been reached.
  • Page 60 Configuration Current Output 1 Output current during Error and HOLD 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select OUT1 menu using  keys, press enter. 5) All items of this menu group are indicated by the “OT1:”...
  • Page 61 Configuration Menu item Action Choices Output current dur- Select ON (22 mA for ON/OFF error message) or OFF ing error message using  keys. Press enter to confirm. Output current Select ON or OFF using ON/OFF  keys. during Sensoface Confirm by pressing messages enter...
  • Page 62: Current Output 2

    Configuration Current Output 2 Output current range. Process variable . . . 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select OUT2 menu using  keys, press enter.
  • Page 63 Configuration Menu item Action Choices Process variable Cond/TMP Select using  keys: Begin: 0 °C Cond: Conductivity End: 100°C TMP: Temperature Press enter to confirm. All the following adjustments are made as for current output 1 (see there)!
  • Page 64: Temperature Compensation

    Configuration Temperature Compensation Selecting the compensation method. TC process medium. 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select CORRECTION menu using  keys, press enter. 5) All items of this menu group are indicated by the “COR:”...
  • Page 65 Configuration Menu item Action Choices Temperature Select desired compensation using keys: compensation OFF: Temp compensation LIN: Linear temperature compensation with entry of temperature coefficient nLF: Temperature compensa- tion for natural waters to EN 27888 NaCl: Ultrapure water with NaCl traces (0 ... +120 °C / +32 ... +248 °F) HCL: Ultrapure water with HCl traces...
  • Page 66 Configuration Temperature Compensation Current input for temp measurement. 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select CORRECTION menu using  keys, press enter. 5) All items of this menu group are indicated by the “COR:”...
  • Page 67 Configuration Menu item Action Choices With external temp measurement (current input enabled / TAN): Current range 4-20 mA / 0-20 mA Select desired range using  keys. Press enter to confirm. Current start Input range: Modify digit using  keys, select next digit using ...
  • Page 68: Control Input (Tan Sw-A005)

    Configuration CONTROL Input (TAN SW-A005) Parameter set selection via external signal or flow measurement 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select CNTR_IN menu using  keys, press enter.
  • Page 69 Configuration Menu item Action Choices Select function of PARSET Select using keys. (selecting parameter Press enter to confirm. CONTROL input set A/B via signal at CONTROL input) Flow (for connecting a pulse- output flow meter) Adjust to flow meter: With “Flow” selected, 12000 pulses/liter you must adjust the device to the flow meter...
  • Page 70: Alarm Settings

    Configuration Alarm Settings Delay. Sensocheck. Tempcheck. 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select ALARM menu using  keys, press enter. 5) All items of this menu group are indicated by the “ALA:”...
  • Page 71 Configuration Menu item Action Choices Delay 0...600 SEC Enter value using  (010 SEC)  keys. Press enter to confirm. Sensocheck Select Sensocheck ON/OFF (continuous monitoring of sensor). Select ON or OFF using  keys. Press enter to confirm (At the same time, Sensoface is activated.
  • Page 72 Configuration Alarm Settings CONTROL input (TAN SW-A005) 1) Press menu key. 2) Select CONF using  keys, press enter. 3) Select parameter set using , press enter. 4) Select ALARM menu using  keys, press enter. 5) All items of this menu group are indicated by the “ALA:”...
  • Page 73 Configuration Menu item Action Choices CONTROL input The CONTROL input can generate an alarm when assigned to FLOW (flow monitoring) in the CONF menu: FLOW CNTR Flow measurement: allows monitoring the minimum and maximum flow (pulse counter) Alarm Specify value Default: 05.00 liters/h Minimum flow FLOW MIN...
  • Page 74: Time And Date

    Configuration Time and Date 1) Press menu key. 2) Select CONF using , press enter. 3) Select parameter set A using  keys, press enter. 4) Select CLOCK using  keys, press enter. 5) All items of this menu group are indicated by the “CLK:”...
  • Page 75 Configuration Time and Date Control of the calibration and cleaning cycles is based on the time and date of the integrated real-time clock. In measuring mode the time is shown in the lower display. When using digital sensors, the calibration data is written in the sensor head. In addition, the logbook entries (cf Diagnostics) are provided with a time stamp.
  • Page 76: Measuring Points (Tag/Group)

    Configuration Measuring Points (TAG/GROUP) Display Backlighting 1) Press menu key. 2) Select CONF using , press enter. 3) Select parameter set A using  keys, press enter. 4) Select TAG or DISPLAY using  keys, press enter. 5) All items of this menu group are indicated by the “TAG:”...
  • Page 77 Configuration Sensor Verification (TAG, GROUP) When Memosens sensors are calibrated in the lab, it is often useful and sometimes even mandatory that these sensors will be operated again at the same measuring points or at a defined group of measuring points. To ensure this, you can save the respective measuring point (TAG) or group of measuring points (GROUP) in the sensor.
  • Page 79: Calibration

    Calibration Note: • All calibration procedures must be performed by trained person- nel. Incorrectly set parameters may go unnoticed, but change the measuring properties. Calibration can be performed by: • Determining the cell factor with a known calibration solution taking account of the temperature •...
  • Page 80: Calibration With Calibration Solution

    Calibration Calibration with Calibration Solution Input of temperature-corrected value of calibration solution with simultaneous display of cell factor. Be sure to use known calibration solutions and the respective temper- ature-corrected conductivity values (see calibration solution tables in the appendix). During the calibration procedure the temperature must be kept constant.
  • Page 81 Calibration Display Action Remark The cell factor and zero point are displayed. The “hourglass” icon is blinking. Use the arrow keys to select: • Repeat (repeat calibration) or • Measuring. Press enter to confirm. With MEAS selected: Display of measured End calibration by variable, Sensoface pressing enter.
  • Page 82: Calibration By Input Of Cell Factor

    Calibration Calibration by Input of Cell Factor You can directly enter the value for the cell factor of a sensor. This value must be known, e.g. determined beforehand in the laboratory. The selected process variable and the temperature are displayed. This method is suitable for all process variables.
  • Page 83: Calibration By Input Of An Installation Factor

    Calibration Calibration by Input of an Installation Factor When using a Memosens sensor in a tight space, the installation factor is entered. Display Action Remark Select Calibration. Press enter to proceed. Select CAL_INSTALL calibration method. Press enter to proceed. Ready for calibration. Display (3 sec) Hourglass blinks.
  • Page 84: Product Calibration

    Calibration Product Calibration (Calibration by sampling) For product calibration, the uncompensated conductivity (mS/cm, S/m) is used. During product calibration the sensor remains in the process. The measurement process is only interrupted briefly. Procedure: 1) The sample is measured in the lab or directly on the site using a portable meter.
  • Page 85 Calibration Display Action Remark The device returns to From the blinking measuring mode. CAL mode indicator you see that product calibration has not been terminated. Product calibration Display (3 sec) step 2: Now the device is in When the sample value HOLD mode.
  • Page 86: Zero Calibration In Air

    Calibration Zero Calibration in Air Display Action Remark Select Calibration. Press enter to proceed. Select CAL_ZERO calibration method. Press enter to proceed. Ready for calibration. Display (3 sec) Hourglass blinks. Now the device is in HOLD mode. Calibration in air SE 670, SE 680-K only Memosens: Edit digits until the lower...
  • Page 87: Temp Probe Adjustment

    Calibration Temp Probe Adjustment Display Action Remark Select Calibration. Wrong settings Press enter to proceed. change the Select CAL_RTD measurement calibration method. properties! Press enter to proceed. Measure the tempera- Display (3 sec) ture of the process Now the device is in medium using an HOLD mode.
  • Page 88: Measurement

    Measurement Display Remark From the configuration or calibration menus, you can switch the device to measuring mode by pressing the meas key. In the measuring mode the upper display line shows the configured process variable or AM/PM and °F: (Cond or temperature), the lower display line shows the time and the second configured process variable (Cond or temperature).
  • Page 89: Diagnostics

    Diagnostics In the Diagnostics mode you can access the following menus without interrupting the measurement: CALDATA viewing the calibration data SENSOR viewing the sensor data SELFTEST starting a device self-test LOGBOOK viewing the logbook entries MONITOR displaying currently measured values VERSION displaying device type, software version, serial number Access to diagnostics can be protected with a passcode...
  • Page 90 Diagnostics Display Menu item Displaying the calibration data Select CALDATA using, press enter to confirm. Use thekeys to select the desired parameter from the bottom line of the display (LAST_CAL CELLFACTOR ZERO). The selected parameter is shown in the main display. Press meas to return to measurement.
  • Page 91 Diagnostics Display Menu item Device self-test (To abort, you can press meas.) 1) Display test: Display of all segments with changing background colors white/green/red. Press enter to proceed. 2) RAM test: Hourglass blinks, then display of --PASS- - or --FAIL-- Press enter to proceed.
  • Page 92 Diagnostics Display Menu item Displaying the logbook entries (TAN SW-A002) Select LOGBOOK using, press enter to confirm. By using the  keys, you can scroll backwards and forwards through the logbook (entries -00-...-99-), -00- being the last entry. If the display is set to date/time, you can search for a particular date using the ...
  • Page 93 Diagnostics Display Menu item Version Here, you find the data you require for requesting a device-specific Option. Display of device type, software/hardware version, and serial number for all device components. Use the  keys to switch between software and hardware version. Press enter to proceed to next device component.
  • Page 95: Service

    Service In the Service mode you can access the following menus: MONITOR Displaying currently measured values. OUT1 Testing current output 1. OUT2 Testing current output 2. (Only if equipped with 2nd current output.) CODES Assigning and editing passcodes. DEFAULT Resetting the device to factory settings. OPTION Enabling options via TAN.
  • Page 96 Service Menu item Remark Displaying currently measured values (sensor monitor) with HOLD mode activated: Select MONITOR using, press enter to confirm. Select variable in the bottom text line using. The selected parameter is shown in the upper display line. As the device is in HOLD mode, you can perform validations using simulators without influencing the signal outputs.
  • Page 97 Service Menu item Remark Assigning passcodes: In the “SERVICE - CODES“ menu you can assign pass- codes to DIAG, HOLD, CAL, CONF and SERVICE modes (Service preset to 5555). When you have lost the Service passcode, you have to request an “Ambulance TAN“ from the manufac- turer specifying the serial number of your device.
  • Page 98: Operating States

    Operating States Operating status Measuring Diag 60 s CAL_SOL Calibration solution CAL_CELL Cell factor P_CAL Product cal S1 P_CAL Product cal S2 CAL_ZERO Zero cal CAL_RTD Temp adjustment CONF ParSet A CONF ParSet B...
  • Page 99 Operating States Operating status HOLD input Explanation: as configured (Last/Fix or Last/Off ) active manual...
  • Page 100: Maintenance And Repair

    • Opening the Service menu • Opening the Confiuration menu Repair The Stratos Pro and the measuring modules cannot be repaired by the user. To request a repair, please contact Knick Elektronische Messgeräte GmbH & Co. KG by visiting www.knick.de.
  • Page 101: A201B/X: Supply Units And Connection

    A201B/X: Supply Units and Connection Recommended Power Supply Units Order No. Stratos Pro A201X, Zone 1: Repeater power supply, Ex, 90...253 V AC, WG 21 A7 output 4...20 mA Repeater power supply, Ex, 90...253 V AC, WG 21 A7 Opt. 470 HART, output 4...20 mA Repeater power supply, Ex, 24 V AC/DC, WG 21 A7 Opt.
  • Page 102: Product Line And Accessories

    Product Line and Accessories Order Code Stratos Pro A201 Example MSCONDI 2-wire / 4-20 mA B,C,E Communication Without (HART retrofittable via TAN) 0 Version number Version Approvals General Safety ATEX / IECEx Zone 2 ATEX / IECEx / FM Zone 1 / Cl 1 Div 1 Measuring channel Memosens pH / Redox digital...
  • Page 103: Specifications

    Specifications CONDI input Input for toroidal conductivity sensors with Memosens protocol (e.g., SE680-M) or SE680-K or SE670 Effective ranges Conductivity 0.0 µS/cm ... 2000 mS/cm Display ranges Conductivity 000.0 ... 999.9 µS/cm 0.000 ... 9.999 mS/cm 00.00 ... 99.99 mS/cm 000.0 ...
  • Page 104 Specifications Concentration -06- H 94 – 99 wt% (-17 °C / 1.4 °F) ... 89 – 99 wt% (115 °C / 239 °F) determination -07- HCl 22 – 39 wt% (-20 °C / -4°F) ... 22 – 39 wt% (50 °C / 122 °F) (continued) -08- HNO 35 –...
  • Page 105 Specifications I input (TAN) Current input 0/4 ... 20 mA / 50 Ω for external temperature signal Start/end of scale Configurable –50 ... 250 °C / –58 ... 482 °F Characteristic Linear Measurement error 1.3) < 1% current value + 0.1 mA HOLD input Galvanically separated (optocoupler) Function...
  • Page 106 Specifications Output 2 Current loop, 4 ... 20 mA, floating, reverse polarity protected For version with 2nd current output only Supply voltage 14 ... 30 V Process variable * Conductivity, resistivity, concentration, salinity, or temperature Characteristic Linear, bilinear or logarithmic Overrange * 22 mA in the case of error messages Output filter *...
  • Page 107 Specifications Diagnostic functions Calibration data Calibration date, sensor parameters Device self-test Display test, automatic memory test (RAM, FLASH, EEPROM), module test Logbook (TAN) 100 events with date and time Extended logbook (TAN) Audit Trail: 200 events with date and time Service functions Sensor monitor Display of direct sensor signals...
  • Page 108 Specifications Wiring Stripping length Max. 7 mm Temperature resistance > 75 °C / 167 °F Rated operating conditions Climatic class 3K5 according to EN 60721-3-3 Location class C1 according to EN 60654-1 Ambient temperature –20 ... 65 °C / –4 ... 149 °F Relative humidity 5 ...
  • Page 109: Calibration Solutions

    Calibration Solutions Potassium Chloride Solutions (Conductivity in mS/cm) Temperature Concentration [˚C] 0.01 mol/l 0.1 mol/l 1 mol/l 0.776 7.15 65.41 0.896 8.22 74.14 1.020 9.33 83.19 1.147 10.48 92.52 1.173 10.72 94.41 1.199 10.95 96.31 1.225 11.19 98.22 1.251 11.43 100.14 1.278 11.67...
  • Page 110 Calibration Solutions Sodium Chloride Solutions (Conductivity in mS/cm) Temperature Concentration [˚C] 0.01 mol/l 0,1 mol/l Saturated 0.631 5.786 134.5 0.651 5.965 138.6 0.671 6.145 142.7 0.692 6.327 146.9 0.712 6.510 151.2 0.733 6.695 155.5 0.754 6.881 159.9 0.775 7.068 164.3 0.796 7.257 168.8...
  • Page 111: Concentration Measurement

    Concentration Measurement Ranges Substance Concentration ranges NaCl 0-26 wt% (0 °C / 32 °F) 0-26 wt% (100 °C / 212 °F) -01- Configuration 0-18 wt% (–20 °C / –4 °F) 22-39 wt% (–20 °C / –4 °F) 0-18 wt% (50 °C / 122 °F) 22-39 wt% (50 °C / 122 °F) Configuration -02-...
  • Page 112: Concentration Curves

    Concentration Curves -01- Sodium chloride solution NaCl -01- c [% by wt]  Concentration measurement not possible in this range. Conductivity versus substance concentration and process temperature for sodium chloride solution (NaCl)
  • Page 113 Concentration Curves -02- Hydrochloric acid HCl -07- -07- -02- c [% by wt]  Concentration measurement not possible in this range. Conductivity versus substance concentration and process temperature for hydrochloric acid (HCl) Source: Haase/Sauermann/Dücker; Z. phys. Chem. New Edition, Vol. 47 (1965)
  • Page 114 Concentration Curves -03- Sodium hydroxide solution NaOH -10- -10- -03- c [% by wt]  Concentration measurement not possible in this range. Conductivity versus substance concentration and process temperature for sodium hydroxide solution (NaOH)
  • Page 115 Concentration Curves -04- Sulfuric acid H -06- -09- -06- -04- -09- c [% by wt]  Concentration measurement not possible in this range. Conductivity versus substance concentration and process temperature for sulfuric acid (H Source: Darling; Journal of Chemical and Engineering Data; Vol.9 No.3, July 1964...
  • Page 116 Concentration Curves -05- Nitric acid HNO -08- -05- -08- c [% by wt]  Concentration measurement not possible in this range. Conductivity versus substance concentration and process temperature for nitric acid (HN0 Source: Haase/Sauermann/Dücker; Z. phys. Chem. New Edition, Vol. 47 (1965)
  • Page 117: Error Handling

    Error Handling Alarm condition: • The display backlighting turns red • The alarm icon is displayed • The complete measured-value display blinks • “ERR xxx“ is displayed in the lower menu line Press the [info] key to view a short error text: •...
  • Page 118: Error Messages

    Error Messages Info text Problem (is displayed in case of Error Possible causes fault when the Info key is pressed) ERR 99 Error in factory settings DEVICE FAILURE EEPROM or RAM defective This error message only occurs in the case of a total defect. The device must be repaired and recalibrated at the factory.
  • Page 119 Error Messages Info text Problem (is displayed in case of Error Possible causes fault when the Info key is pressed) ERR 10 CONDUCTANCE TOO HIGH Measuring range of conductance exceeded > 3500 mS ERR 11 Display range limits exceeded CONDUCTIVITY RANGE Cond >...
  • Page 120 Error Messages Info text Problem (is displayed in case of Error Possible causes fault when the Info key is pressed) ERR 100 INVALID SPAN OUT1 Span Out1 configuration error Selected span too small ERR 101 INVALID SPAN OUT2 Span Out2 configuration error Selected span too small ERR 105 INVALID SPAN I-INPUT...
  • Page 121: Decommissioning

    Decommissioning Disposal Local codes and regulations must be observed when disposing of the product. Returns If required, send the product in a clean condition and securely packed to your local contact. See www.knick.de.
  • Page 122: Sensoface

    Sensoface (Sensocheck must have been activated during configuration.) The smiley in the display (Sensoface) alerts to sensor problems (defective sensor, defective cable, maintenance required). The permitted calibration ranges and the conditions for a friendly, neutral, or sad Sensoface are summarized in the following table.
  • Page 123 Sensoface Display Problem Status Sensor defect Wrong or defective sensor or excessive cable capacitance (see also error message Err 15). Temperature Temperature outside range for TC, conc, sal...
  • Page 124: Hart: Typical Applications

    HART: Typical Applications (SW-A001) HART Terminal + out Power supply 250 Ω 4 ... 20mA min 19 V - out 1 - out 2 A201 HART Terminal HART Terminal Repeater + out power supply 4 ... 20mA - out 1 - out 2 A201...
  • Page 125: Fda 21 Cfr Part 11

    FDA 21 CFR Part 11 Conformity with FDA 21 CFR Part 11 In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11, Electronic Records; Electronic Signatures“ the American health agency FDA (Food and Drug Administration) regulates the production and processing of electronic documents for pharmaceutical develop- ment and production.
  • Page 126: Index

    Index Access code assignment 97 Accessories 102 Activating an option 97 AIR-SET calibration 86 Alarm 30 Alarm and HOLD messages 31 Alarm settings 70 Ambulance TAN 97 Assembly 13 Audit Trail 125 Audit Trail, diagnostics 92 Autorange 53 Backlighting 22 Block diagram 12 Cable glands 17 Calibration 79...
  • Page 127 Index Concentration solution, selection 44 Configuration, alarm 70 Configuration, CONTROL input 68 Configuration, current output 1 52 Configuration, current output 2 62 Configuration, display backlighting 76 Configuration, general 27 Configuration, individual settings 39 Configuration, measuring point 76 Configuration, menu groups 33 Configuration, menu structure 32 Configuration, output current during Error and HOLD 60 Configuration, overview 35...
  • Page 128 Index Diagnostics, general 27 Diagnostics, logbook 92 Diagnostics mode 89 Diagnostics, sensor monitor 92 Diagnostics, version 93 Dimensions 14 Display 22 Display backlighting 22 Display data in Diagnostics mode 89 Display test 91 Disposal 121 Documentation: package contents 3 EEPROM test 91 Electrical installation 17 Electronic Signature 125 Enabling an option 97...
  • Page 129 Index HART, typical applications 124 HOLD, configuration 61 HOLD mode, description 29 HOLD mode, exit 29 HOLD mode, external activation 30 HOLD mode, manual activation 30 HOLD mode, output signal 29 Info text 118 Installation, notices 17 Intended use 7 Keypad 21 Logarithmic curve 56 Logbook, diagnostics 92...
  • Page 130 Index Operating modes 27 Operating mode, selection 24 Operating states 98 Option request 97 Option request: Conditions 93 Options, overview 102 Order code 102 Output current, display 88 Output current range 62 Output current, specify value 96 Output filter 58 Output signal during HOLD 29 Output signal during HOLD, configuration 61 Overview, device properties 9...
  • Page 131 Index RAM test 91 Rated operating conditions 108 Rating plates 18 Repair 100 Reset to factory settings 97 Returns 121 Safety 7 Safety guide 3 SE 670/680, wiring (via RS-485) 20 Selection menu 24 Sensocheck, configuration 70 Sensocheck, description 122 Sensoface, description 122 Sensoface, troubleshooting 117 Sensor connection, assignment 19...
  • Page 132 Index Tag number, display 88 TAG (point of measurement) 77 TAN input 97 TAN options, activation 97 TAN options, overview 102 Technical data 103 Tempcheck 70 Temperature compensation 65 Temperature probe adjustment 87 Temperature probe, monitoring 71 Temperature unit 46 Terminal assignments 18 Test report 2.2 3 Time and date (configuration) 74...
  • Page 133 Index...
  • Page 134 Knick Elektronische Messgeräte GmbH & Co. KG Headquarters Beuckestraße 22 • 14163 Berlin Germany Phone: +49 30 80191-0 Fax: +49 30 80191-200 info@knick.de www.knick.de Local Contacts www.knick-international.com Translation of the original instructions Copyright 2022 • Subject to change Version: 4 • This document was published on February 18, 2022.