PS‐2600; Wireless Spectrometer Battery Replacement Instructions  
Reference:  Battery Model Number: PS‐2569 (Part Number 540‐079); Li‐Poly 2500 mAH 
 
 
 
Recommended Equipment: (PS‐3299; Battery Replacement Tool Set)
 
 
Step 1:
 Position the Spectrometer looking at the back surface as shown below.  
                  Note the location of the (4) T‐8 Screws 
 
 
 
PASCO Form Number
Description
Date
Revision
013-17354
PS-2600 Battery Replacement Instructions
05-05-22
Page 1 of 4
A
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PS-2600 and is the answer not in the manual?

Questions and answers

Summary of Contents for PASCO PS-2600

  • Page 1     PS‐2600; Wireless Spectrometer Battery Replacement Instructions   Reference:  Battery Model Number: PS‐2569 (Part Number 540‐079); Li‐Poly 2500 mAH        Recommended Equipment: (PS‐3299; Battery Replacement Tool Set)     Step 1:  Position the Spectrometer looking at the back surface as shown below.                     Note the location of the (4) T‐8 Screws        PASCO Form Number Description Date Revision 013-17354 PS-2600 Battery Replacement Instructions 05-05-22 Page 1 of 4  ...
  • Page 2     Step 2:  Unscrew the (4) T‐8 Screws with a T‐8 Screwdriver. While holding the Spectrometer                     plastic halves together, turn the Spectrometer over and let the (4) screws fall out. Set                     the Spectrometer down on a flat surface. Put the screws off the side and lift off the                     top cover.           Step 3:  Rotate the bottom half as shown.        PASCO Form Number Description Date Revision 013-17354 PS-2600 Battery Replacement Instructions 05-05-22 Page 2 of 4  ...
  • Page 3     Step 4a:    Unplug the battery connector and remove the Phillips Screw.   Step 4b:    Lift the Printed Circuit Board Assembly by the plastic center and set apart.            PASCO Form Number Description Date Revision 013-17354 PS-2600 Battery Replacement Instructions 05-05-22 Page 3 of 4  ...
  • Page 4     Step 5:  Remove the old battery and add the new replacement battery at the location shown  in the picture above. Replace the printed circuit board above the battery and plug the battery  into its mating connector. Secure the Phillips Screw in place.      Step 6:  Locate the top cover and while holding the Spectrometer plastic halves together, turn  the Spectrometer over and secure the (4) mounting screws in place with the T‐8 driver.        If you need further assistance, please contact Technical Support at support@pasco.com or  through www.pasco.com on chat.  To access chat through your browser, please enable all  cookies on our website.    PASCO Form Number Description Date Revision 013-17354 PS-2600 Battery Replacement Instructions 05-05-22 Page 4 of 4  ...