Table of Contents Product Manual - UltraDoser 150S LN Dosing System ® Revision Log Revision Level Date Description 08/24/2015 Reformat with new layout Update pages 12 and 19 for minimum dose amount of 20mS instead of 10mS. 06/09/2016 Update CryotechFlex hose to CFlex throughout. Update trademarks, update warranty section.
• Discrete Dosing - up to 2000* containers per minute manual. If after reading this manual you are not confident in carrying out any task, please contact Chart’s service team at • Discrete or Continuous Dosing - change over defined by 1-408-371-4932.
Preface Product Manual - UltraDoser 150S LN Dosing System ® Terms Acronyms / Abbreviations Throughout this manual safety precautions will be designated The following acronyms / abbreviations are used throughout as follows: this manual: Pressure (Metric) Warning! Description of a condition that can result in personal injury or Gaseous Nitrogen death.
Handle liquid so that it will not splash or spill. Protect your eyes and cover skin where the possibility of contact with Thank you for your purchase of Chart Inc.’s (Chart) liquid, cold pipes and equipment, or cold gas exists. Safety UltraDoser 150S Liquid Nitrogen Dosing System.
Most individuals working in or around oxygen deficient atmospheres rely on the “buddy system” for protection - loosening fittings, empty the UltraDoser 150S system of liquid obviously the “buddy” is equally susceptible to asphyxiation and release any vapor pressure if he or she enters the area to assist the unconscious partner unless equipped with a portable air supply.
The application must be evaluated to determine the ideal location of the dosing head Upon arrival of the UltraDoser 150S system, it is advised to on the filling line. immediately inspect for any signs of damage. If any damage occurred in shipping, claims must be filed with the shipping •...
Chart can supply a prefabricated stand to accommodate the mounting bracket assembly. This stand can be utilized in almost all installations. If the Chart stand cannot be used in your installation, fabricating one with 1½” diameter rod or round bar will make installation of the UltraDoser 150S unit simpler.
Mount the 150S controller at a convenient location. that power is being supplied to the heater (either dosing Brackets are supplied to mount the controller on the Chart head or vent). prefabricated support stand or 1 1/2” diameter rod or round •...
150S LN Dosing System ® Installing the Nozzle Three nozzles are supplied with the UltraDoser 150S system - 0.040 inch, 0.050 inch, and 0.060 inch. Custom sizes may be ordered from Chart. Remove the dosing head heater. Select a nozzle.
Installing the Bottle Detect Sensor The bottle detect sensor must be a PNP type sensor and must be installed for the UltraDoser 150S system to operate correctly. The sensor must be mounted approximately 2 ½” from the UltraDoser dosing head.
The only adjustments are Duration (of dose) and Dose Delay. container) must be processed in the same manner. Caution! There are additional screens Chart’s UltraDoser 150S system guarantees that a precise, accessible on the 150S controller accurate dose of LN is delivered.
Operation Product Manual - UltraDoser 150S LN Dosing System ® Remove the display cover. Press and hold ‘ESC’ until the cursor appears and blinks on the Duration mS value line. Press and HOLD ‘ESC’ until cursor _ appears and blinks on the Duration mS value line (image below). Press the ►...
Daily Operating Procedures Attach the CFlex hose (½” female flare side) to the The UltraDoser 150S unit can be fed by either a portable house GN system or portable GN cylinder.
Open the liquid valve (counter-clockwise direction) on the Dura-Cyl cylinder. Note: The UltraDoser 150S system will continue to dose properly until the liquid level inside Wait until the UltraDoser unit is filled with liquid the UltraDoser unit runs low. This feature nitrogen, approximately 10 minutes.
Product Manual - UltraDoser 150S LN Dosing System ® Service and Maintenance Nozzle Change Out Procedure Caution! The dosing head heater may still be in operation. Do not expose Remove the dosing head heater. skin to prolonged contact with the dosing head heater. The Insert the nozzle tool into the nozzle area until the tool maximum temperature of the connects with the nozzle (image 1).
Note that possible causes for specific symptoms are listed in descending order of significance. That is, check out the first cause listed before proceeding to the next. If you need further assistance please contact Chart’s service team at 1-408-371- 4932...
Page 21
24 hours with a continuous purge of warm nitrogen gas. Contact Chart’s service team at 408- 371-4932 for a detailed procedure. Liquid is coming out of the dosing head The LN supply pressure is too high.
The stainless steel vacuum insulated reservoir provides a This assembly is designed to attach to the 150S controller working supply of LN for dosing operations from your and fit on Chart’s support stand or 1 1/2” stainless steel rod. liquid nitrogen supply. Bayonet Connection 150S Controller The bayonet connection allows a completely vacuum 150S controller dictates the dosing operation of the system.
Page 25
The assembly consists of the bracket attaching to the UltraDoser unit and two clamps. These clamps are designed to fit on Chart’s support stand or 1 1/2” stainless steel rod. The bracket can be mounted in 3 positions. Dosing Head Heater The UltraDoser unit has a self-regulating dosing head heater.
LN Dosing Systems, and, if said LN and hold Chart harmless from any third party claims as a Dosing Systems is found defective, shall, at Chart’s option result of the use, sale, or lease of the LN...