User guide Ultrasonic heat meter SonoMeter 30 Test and calibration instruction Table of Contents General information ....................... 2 Test instruction ........................2 Calculator energy measurement error (δ ) test ..................2 Volume measurement error (δ ) test ......................2 Calibration instruction ......................3 Calibration mode .............................
Page 2
Ultrasonic heat meter SonoMeter 30 1. General information This instruction designed for ultrasonic heat measurement error (δ ) and temperature sensor meter SonoMeter 30 flow and energy verification temperature difference measurement error (δ and calibration. Heat meter energy measurement error (δ ) are all components of the arithmetic sum of the Heat meter energy measurement error (δ...
Page 3
Fig. 2 3.2 Software installation The software “SonoMeter30Config.exe” is used also via M-Bus interface module then converter to read and configure SonoMeter 30 device via COM/M-Bus (MB-1) can be used. OG-1-USB optical reader. Open “SonoMeter30Config.exe”. Press “Next”. Meter configuration / calibration cab be made Then, you can choose installation directory.
Page 4
User guide Ultrasonic heat meter SonoMeter 30 3.5 Volume calibration Measurement units, flow pipe tagging, and press “set” in order to configure what energy calibration can be configured in “Manufacturer” units are used for measurements. If device is window. successfully configured message “Operation Select energy units option in “energy unit”...