2.Safety notes

> Before using the medical device for the first time, store it at room temperature for 24 hours.
> The operation of the medical device is permitted only on supply units which correspond to the standards
IEC 60601-1 (EN 60601-1) and IEC 60601-1-2 (EN 60601-1-2).
> Always ensure the correct operating conditions.
> Check the medical device for damage and loose parts each time before using (e.g., push-button).
> Do not operate the medical device if it is damaged.
> Only attach the medical device onto the motor when the motor is at a complete standstill.
> Do not activate the push-button of the medical device during operation. This leads to detachment of
the rotary instrument, damage to the chucking system and/or heating up of the medical device. Risk of
burning!
> Perform a test run each time before using.
> Do not touch the soft tissue with the contra-angle head (risk of burning due to the push-button heating
up)!