HP Q METER 4342A Operating And Service Manual page 62

Table of Contents

Advertisement

Section
V
Paragraph
5-22
b.
Set 4342A controls
as follows:
FREQUENCY RANGE . . . . . . . . . 22k - 70k
FREQUENCY dial
. . . . . . . . . . . . . . . . 2.2
L/C dial
. . . . . . . . . . . . . . . . . . . 25(pF)
AC dial
. . . . . . . . . . . . . . . . . . . . . . . . -5
C.
Set the Test
Oscillator
frequency
to
1OOkHz and adjust
the
signal
level
until
the
Digital
Voltmeter
reads
9OO.OmV.
d.
Adjust
A6R2 (Xl GAIN adj.)
for
full
scale
Q meter
reading.
e.
Adjust
the
Test
Oscillator's
signal
level
until
the
DVM reads
450.0mV.
f.
Q meter
should
indicate
exactly
l/2
full
scale.
If
Q meter
deflection
is
insufficient,
rotate
ASR43 (LINEARITY
ADJ) CCW until
meter
reads
correctly.
If
deflection
is excessive,
rotate
AS-
R43 slightly
CW.
g-
Repeat
steps
c through
f until
meter
indicates
l/2
full
scale
within
+O.S/lOO
full
scale
(l/2
minor
divi-
sion)
in step
f.
h.
Adjust
the Test
Oscillator's
signal
level
until
the
DVM reads
300.0mV.
1.
Q meter
should
indicate
within
l/3
full
scale
?l/lOO
of full
scale
(1
minor
division).
If
not,
repeat
steps
c through
f.
Model 4342A
2) X10 GAIN adjustment.
j.
Adjust
the
Test
Oscillator's
signal
level
until
the
DVM reads
810.0mV.
k.
Depress
4342A AQ button
and set
AQ
RANGE to 10.
1.
Adjust
AQ ZERO control
for
10 (zero
scale
deflection)
on AQ scale.
m.
Adjust
the Test
Oscillator's
signal
level
until
the
DVM reads
900.0mV.
n.
Q meter
should
indicate
0 (full
scale)
on AQ scale.
If
Q meter
reading
is
not
zero,
adjust
A6R21 (X10 GAIN adj.)
for
correct
reading.
Repeat
steps
j
through
n because
both
adjustments
interact.
5-22.
Q ANALOG OUTPUT
ADJUSTMENT.
A Test
Oscillator
and a Digital
Voltmeter
are
required
for
this
adjustment.
a.
Set 4342A controls
as follows:
Q RANGE . . . . . . . . . . . . . . . . . . . . . . . . 30
other
controls
. . . . . . . any settings
b.
Connect
Digital
Voltmeter
to 4342A
rear
panel
Q ANALOG OUTPUT connector.
C.
Adjust
A6R4 (REC BAL adj.)
for
OV
?O.OlV on Digital
Voltmeter
display.
TEST OSCILLATOR
1
4342A
l-y
DIGITAL
VOLTMETER
S-10
Figure
5-6.
Voltmeter
Adjustment.