Faulty Q Limit Operation - HP Q METER 4342A Operating And Service Manual

Table of Contents

Advertisement

Model 4342A
Section
V
Paragraphs
5-45 and 5-46
5-45.
Low Q indication
in
high
frequency
measurements.
If
the Q meter
shows lower
Q indication
at
higher
frequencies
(above
approx.
lOMltz),
it
is conceivable
that
the symptom is being
caused by a drop in Q of the tuning
capacitor.
The tuning
capacitor
has a spring
contact
brush
for
grounding
the capacitor
rotor
plates
with
minimal
residual
impedance
to
maintain
the inherent
loss of the
capacitor
at minimum in the high
frequency
region.
A
contact
brush
in service
for
a long period
may possibly
cause an increase
in contact
resistance
and resultant
increase
in capa-
citor
loss.
The remedy for
this
trouble
is
to clean
the contact
brush.
Clean with
a
cloth
moistened
with
alcohol.
To take
out
the contact
brush,
proceed
as follows:
a.
Remove top cover.
b.
Remove
white
plastic
top plate
on
measurement
terminal
deck.
c.
Unsolder
center
conductor
1 of
0
coaxial
module
connected
to A4 Imped-
ance Converter
(see Figure
S-11).
d.
Remove nut 0
retaining
the coaxial
assembly
module.
e.
Remove the
six terminal
deck retaining
screws
0.
f.
Lift
terminal
deck up and out.
The
contact
brush
is located
on bottom
side
of terminal
deck.
5-46.
Faulty
Q Limit
Operation.
If
4342A operates
normally
in Q measurements
but Q OVER LIMIT
indication
malfunctions,
A7
Q Limit
Selector
assembly
is probably
faulty.
If
OVER LIMIT
lamp does not
light,
first
check lamp AlODSS.
A2
TUNING
CAPACITOR
ASS?
Figure
S-11.
Tuning
Capacitor
Disassembly
(top
view).
5-19