Fluke 187 Calibration Manual page 28

True rms multimeter
Hide thumbs Also See for 187:
Table of Contents

Advertisement

187/189
Calibration Manual
Step
Function
42
AC uA
43
DC mA
44
DC mA
45
DC mA
46
DC mA
47
DC uA
48
DC uA
49
DC uA
50
DC uA
51
AC A
52
AC A
53
DC A
54
DC A
55
DC A
56
DC A
57
AC A
1.
Do not rely on autoranging for all tests. Some test steps will require manual ranging.
2.
Must allow for source and meter settling before recording reading.
3.
Wait at least 60 seconds for source and meter to settle before recording reading.
Using Relative Mode (REL) to zero residual reading (apply 0 Ω and allow UUT to settle for 45 seconds before
4.
pushing REL
20
Table 4. Performance Test Steps (cont.)
1
Range
5000.0 uA
50.000 mA
50.000 mA
50.000 mA
400.00 mA
500.00 uA
-500.00 uA
5000 uA
500.00 uA
5.0000 A
5.0000 A
5.0000 A
5.0000 A
-5.0000 A
10.000 A
10.000 A
)
.
Frequency or
Input Level
Model
1.5 mA
1 kHz
0 A
50 mA
-50 mA
-329.00 mA
500 uA
-500 uA
-5 mA
0 A
250 mA
1 kHz
250 mA
20 kHz
0 A
5 A
-5 A
-6 A
6 A
1 kHz
2
Display
Lower
Upper
Limit
Limit
1488.2
1511.8
-0.010
0.010
49.915
50.085
-49.915
-50.085
-328.49
-329.51
498.55
501.45
-498.55
-501.45
-49187.3
-5012.7
-0.02
0.02
0.2422
0.2578
0.2310
0.2690
-0.0010
0.0010
4.9740
5.0260
-4.9740
-5.0260
-5.968
-6.032
5.905
6.095