Fluke 187 Calibration Manual page 15

True rms multimeter
Hide thumbs Also See for 187:
Table of Contents

Advertisement

Function
Range
DC mV
50.000 mV
500.00 mV
3000.0 mV
DC V
5.0000 V
50.000 V
500.00 V
1000.0 V
DC µA
500.00 µA
5,000 µA
DC mA
50.000 mA
400.00 mA
DC A
5.0000 A
1
10.000 A
1. 10 A continuous up to 35 °C, less than 10 minutes 35 °C to 55 °C. 20 A overload for 30 seconds maximum.
2. 20 counts in dual display DC or AC+DC.
3. See AC conversions notes for AC mV and V
Function
1
Resistance
Conductance
1. For the 5,000 count mode, divide the number of least significant digits (counts) by 10.
2. For relative humidity greater than 70 %, resistance accuracy is 0.5 % over 1 MΩ and 2.5 % over 10 MΩ.
3. Using relative mode (REL ∆) to zero residual reading.
4. To ensure stated accuracy, switch to conductance mode and verify that the open circuit reading is less than
0.10 nS.
Accuracy
Resolution
DC
0.001 mV
0.1% + 20
0.01 mV
0.03 % + 2
0.1 mV
0.025 % + 5
0.0001 V
0.025 % + 10
0.001 V
0.03 % + 3
2
0.01 V
0.1 % + 2
2
0.1 V
0.1 % + 2
0.01 µA
0.25 % + 20
0.1 µA
0.25 % + 2
0.001 mA
0.15 % + 10
0.01 mA
0.15 % + 2
0.0001 A
0.5 %+ 10
0.001 A
0.5 %+ 2
Range
Resolution
500.00
0.01
5.0000 k
0.0001 k
50.000 k
0.001 k
500.00 k
0.01 k
5.0000 M
0.0001 M
30.000 M
0.001 M
100.0 M
0.1 M
500.0 M
0.1 M
50.00 nS
0.01 nS
Accuracy Dual Display AC or AC+DC
20 Hz-
45 Hz -
45 Hz
2 % + 80
0.5 % + 40
2
2
1 % + 20
1.0 % + 20
1 % + 10
0.75 % + 10
1 % + 20
0.75 % + 20
1 % + 10
1 % + 10
2 % + 20
2 % + 20
1.5 % + 10
1.5 % + 10
True RMS Multimeter
Detailed Accuracy Specifications
3
1 kHz-
1 kHz
20 kHz
6 % + 40
2 % + 40
Not specified
Not specified
2 % + 40
2 % + 40
2 % + 40
3 % + 40
6 % + 40
5 % + 10
Accuracy
3
0.05 % + 10
0.05 % + 2
0.05 % + 2
0.05 % + 2
2
0.15 % + 4
2
1 % + 4
4
3 % + 2
4
10 % + 2
1 % + 10
7