Download Print this page
Fluke 187 Extended Specifications

Fluke 187 Extended Specifications

True-rms digital multimeter
Hide thumbs Also See for 187:

Advertisement

Page 1 of 3, Document #2153
187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
True-rms Digital Multimeter
Nominal Specifications
Function
Ranges/Description
DC Voltage
0 to 1000V
DC Current
0 to 10A (20A for 30 seconds)
AC Voltage, True-rms
2.5 mV to 1000V - 100 kHz bandwidth
AC Current, True-rms
25 A to 10A (20A for 30 seconds)
Resistance
0 to 500 Mohms
Conductance
0 to 500 Nanosiemens
Capacitance
0.001 nF to 50 mF
Diode Test
3.1V
Temperature
-200 C to 1350 C (-328 F to 2462 F)
Frequency
0.5 Hz to 1000 kHz
Accuracy
(Basic DC V)
0.025%
(Basic AC V)
0.4%
Features
Feature
Description
Dual Displays
50,000 count primary display
5,000 count secondary display
Backlight with 2
Bright white backlight for clear readings
brightness selections
in poorly lighted areas
Fast Autorange
Meter automatically selects best range -
instantly
AC+DC True RMS, ac
Choices for AC only, AC and DC dual
rms specified to 100 kHz
display, or AC+DC readings
dBm, dBV
User selectable impedance references
for dBm
AutoHOLD
Holds readings on display
Continuity/Open Test
Beeper sounds for Ohms readings below
threshold or for momentary
open circuit indication
Fast Bar Graph
51 Segments for peaking and nulling
Duty Cycle/Pulse Width
Measure time a signal is on or off in %
or milliseconds
MIN MAX/Fast MIN MAX
Record Maximum, Minimum, and Average
with elapsed and
values. Real Time for MAX or MIN, elapsed
Real Time Stamp
time for AVG.
Fast MIN MAX captures peaks to 250 sec.
Closed Case Calibration
No internal adjustments needed
Battery/Fuse
Battery or fuse replacement without
Access Door
voiding calibration
Hi-Impact Overmold Case Integrated Protective Holster provides
superior impact protection for your meter
1
For the 5,000 count mode, divide the number of least significant digits (counts) by 10.
2
A residual reading of 8 to 180 digits with leads shorted, will not affect stated
accuracy above 5% of range.
3
20 counts in dual display DC or AC + DC
4
C, less than 10 minutes 35 to 55 . 20A overload for
10A continuous up to 35
30 seconds maximum
5
See AC conversion notes for AC mV and V.
Above specifications are subject to change without notice.
Fluke 187/189
Extended Specifications
Detailed Specifications
Accuracy is specified for a period of one year after calibration, at 18 C to 28 C (64 F to 82 F) with
relative humidity to 90%. Accuracy specifications are given as ([% of reading] + [number of least
significant digits])
Function
AC mV
AC V
dBV
Function
AC A
AC mA
AC A
Function
DC mV
DC V
DC A
DC mA
DC A
45 Hz-
Range
Resolution 1 kHz
1,2
50,000mV
0.001mV
0.4% + 40 2% + 80
500.00 mV
0.01 mV
0.4% + 40 2% + 80
3000.0 mV
0.1 mV
0.4% + 40 2% + 80
1,2
5.0000V
0.0001V
0.4% + 40 2% + 80
50.000V
0.001V
0.4% + 40 2% + 80
500.00V
0.01V
0.4% + 40 2% + 80
1000.0V
0.1V
0.4% + 40 2% + 80
-56 to -6
0.01 dB
0.1 dB
-6 to +34
0.01 dB
0.1 dB
+34 to +60 0.01 dB
0.1 dB
45 Hz-
Range
Resolution 1 kHz
500.00 A
0.01 A
0.75% + 20
5000.0 A
0.1 A
0.75% + 5
50.000 mA
0.001 mA
0.75% + 20
400.00 mA
0.01 mA
0.75% + 5
5.0000A
0.0001A
1.5% + 20
10.000A
4
0.001A
1.5% + 5
Accuracy
Range
Resolution DC
50.000 mV
0.001mV
0.1% +20
500.00 mV
0.01 mV
0.03% + 2
3000.0 mV
0.1 mV
0.025% + 5
5.0000V
0.0001V
0.025% + 10
50.000V
0.001V
0.03% + 3
500.00V
0.01V
0.1% + 2
1000.0V
0.1V
0.1% + 2
500.00 A
0.01 A
0.25% + 20
5.000 A
0.1 A
0.25% + 2
50.000 mA
0.001 mA
0.15% + 10
400.00 mA
0.01 mA
0.15% + 2
5.0000A
0.0001A
0.5% + 10
4
10.000A
0.001A
0.5% + 2
Accuracy
20 Hz-
1 kHz-
10 kHz-
45 Hz
10 kHz
20 kHz
5% + 40
5.5% + 40
5% + 40
5.5% + 40
0.4% + 40
1.5% + 40
0.4% + 40
1.5% + 40
0.4% + 40
1.5% + 40
0.4% + 40
Not specified Not specified
0.4% + 40
Not specified Not specified
0.2 dB
0.5 dB
0.5 dB
0.2 dB
0.1 dB
0.2 dB
0.2 dB
0.1 dB
Not specified Not specified
Accuracy
20 Hz-
1 kHz-
45 Hz
20 kHz
1% + 20
0.75% + 20
1% + 5
0.75% + 10
1% + 20
0.75% + 20
1% + 5
1.5% + 10
1.5% + 20
6% + 40
1.5% + 5
5% + 10
Accuracy Dual Display AC or AC+DC
20 Hz-
45 Hz-
45 Hz
1 kHz
2% + 80
0.5% + 40
2% + 80
0.5% + 40
2% + 80
0.5% + 40
3
2% + 80
0.5% + 40
3
2% + 80
0.5% + 40
3
2% + 80
0.5% + 40
3
2% + 80
0.5% + 40
1% + 20
1% + 20
1% + 10
0.75% + 10 2% + 40
1% + 20
0.75% + 20 2% + 40
1% + 10
1% + 10
2% + 20
2% + 20
1.5% + 10
1.5% + 10
20 kHz-
100 kHz
15% + 40
8% + 40
8% + 40
8% + 40
8% + 40
1.4 dB
0.8 dB
20 kHz-
100 kHz
6% + 40
2% + 40
9% + 40
4% + 40
Not specified
Not specified
5
1kHz-
20 kHz
6% + 40
2% + 40
Not specified
Not specified
2% + 40
3% + 40
6% + 40
5% + 40

Advertisement

loading

Summary of Contents for Fluke 187

  • Page 1 Page 1 of 3, Document #2153 187/189 Extended Specifications ©2000 Fluke Corporation Rev. A-8/00 Fluke 187/189 True-rms Digital Multimeter Extended Specifications Nominal Specifications Detailed Specifications Accuracy is specified for a period of one year after calibration, at 18 C to 28 C (64 F to 82 F) with...
  • Page 2 Accuracy is specified for a period of one year after calibration, at 18 C to 28 C (64 F to 82 F) with relative humidity to 90%. The Fluke 189 adds the following capabilities: Accuracy specifications are given as ([% of reading] + [number of least...
  • Page 3: Safety Information

    10.0 cm X 20.3 cm X 5.0 cm (3.94" X 8.00" X 1.97") (Not Including Accessory Mount) Weight 545 grams (1.2 lbs.) Case Sealing IP-42 per IEC 529, Section 3 Warranty Lifetime Calibration Interval 1 year Above specifications are subject to change without notice. Fluke. Keeping your world up and running.