  • Page 1 DenOptix ® Operator and Service Manual Manuale dell’Operatore e Servizio Manuel de l’Opérateur et de Service Gebrauschsanweisung Manual del Usuario y de Servicio Phosphor Plate Technology...
  • Page 2 English Operator and Service Manual Inglese Anglais Englisch Inglés...
  • Page 3: Table Of Contents

    Preparing for the next patient ........37 Turning the DenOptix QST system on and off ......38...
  • Page 4 Appendix D: If you need assistance ........62 11.4 ©2005 Gendex Dental Systems Gendex, DenOptix QST, AcuCam and Orthoralix are registered trademarks of Gendex Dental Systems. Rinn, XCP is a registered trademark of Dentsply International. VixWin and Concept are trademarks of Gendex Dental Systems.
  • Page 5 • Imaging plates are reusable. • Saves critical time vs. using processing systems. DenOptix QST will process 4 bitewings in under 30 seconds and a FMX will process 20 images in less than 4 seconds per image. Panoramic images can be obtained in less than 80 seconds. A cephalometric image can be obtained in less than 80 seconds.
  • Page 6 ® panoramic, we are dedicated to providing products with the features and high quality you desire. As you enjoy using the new DenOptix QST system, you are taking advantage of our 50 years of manufacturing expertise in the dental industry.
  • Page 7 Indications for use The DenOptix QST is intended for use as a digital dental radiography system using X-ray recording media (phosphor imaging plates) for radiographic diagnostic intraoral and extraoral exposures providing interactive CRT retrieval, viewing and processing of stored computed radiographic images.
  • Page 8 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 9 Contraindications There are no contraindications for use. Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 10 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 11: Safety Requirements

    Safety requirements Important information It is important that all personnel who will operate the DenOptix QST System read and under- stand this manual before operating the device. All personnel should follow all warnings and cautions as outlined in Section 3.2, for their safety and the safety of others around them.
  • Page 12 Standard; Electromagnetic Compatibility Requirements and Tests, class A. Use only grounded electrical connections. Connect the DenOptix QST scanner to a grounded electrical outlet between 100-240 volts AC. Do not reuse the barrier envelopes. Once used within the oral cavity, the barrier envelopes are bio-contaminated and should NEVER be reused.
  • Page 13 Caution statements Reduce the exposure time on your intraoral X-ray unit. The DenOptix QST System is designed to produce high quality, diagnostic images at reduced radiation levels. To get the maximum benefit, we recommend that the exposure be reduced as outlined in Section 5.1.
  • Page 14 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 15: Pre-Installation Information And Recommendations

    This manual provides the essential information necessary for the installation, operation and routine care of the DenOptix QST System. The detailed instructions for the imaging software associated with the scanner can be found in the Imaging Software User Manual and Installation Guide.
  • Page 16 Gendex DenOptix® QST ceph Cephalometric Intraoral Combo Combination system capable of scanning both I/O and panoramic imaging plates TIFF Tagged image file format LP/mm Line pairs per millimeter PhotoStimulable Phosphor D.P.I. Dots per inch Central processing unit (your computer) International Electrotechnical Commission...
  • Page 17: Physical Description

    Gendex DenOptix® QST Physical description The DenOptix QST system has several configurations. Even if the basic hardware is the same, DenOptix QST is provided with different licenses for scanning different types of plates, according to the chosen model. See the following table for a complete list of components for each model.
  • Page 18 Gendex DenOptix® QST Table of models and components Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 19 Yellow Blinking – Scanning Red — Error Figure 4-2 Scanner rear view 1) DenOptix QST scanner power entry module with switch and fuses 2) USB 2.0 connector. A USB 2.0 cable (provided) connects the DenOptix QST scanner to the CPU’s USB 2.0 port.
  • Page 20 Rear (Black) 1) IP serial number for tracking purposes 2) Date of manufacture 3) Available from Gendex Dental Systems 4) Solid orientation dot 5) IP size indicator. Shows the equivalent film size 6) Orientation circle. Visibly shows on final radiograph for faster orientation...
  • Page 21 2) Active imaging area, white color 3) Patient left ( ) and right ( ) side indicators 4) General information that includes: • Available from Gendex • Date of manufacture • Imaging plate size • IP serial number Doc # M010-004WWE July 2005...
  • Page 22 Gendex DenOptix® QST Figure 4-5 Carousels Extraoral QEO/Ceph (QCH) Intraoral (QFMX) 1) Top of the carousel. For the EO/CH carousels, the knob will be positioned based upon whether or not a Pan or Ceph is being scanned. If a Pan or Ceph is loaded, turn the knob towards the mark indicating a Pan or Ceph.
  • Page 23: Computer Requirements

    Power 110 watts maximum Power Cord Standard line cord provided (Medical Grade for 115V) Compliance to standards The DenOptix QST System conforms to the following standards: Standard Title UL 60601-1 Medical Electrical Equipment, Part 1: General requirements for safety 21 CFR Chapter I...
  • Page 24: Installation

    We do not recommend that an untrained person try to install and configure the system. Site selection The DenOptix QST scanner can be located almost anywhere in the dental office. The site you pick should have: • Subdued lighting conditions. You should have the ability to turn down/off the lights and block sunlight with blinds.
  • Page 25: Unpacking The Unit

    Gendex DenOptix® QST Unpacking the unit The DenOptix QST System is shipped in one box. Completely unpack the box and save it in a safe, dry location. You may need to repack the unit for shipping if you ever encounter a main- tenance outage, or relocate your office.
  • Page 26: Hardware Setup And Connections

    Step 3: Turn the power on to the monitor and computer. Step 4: Connect the device end of the USB 2.0 cable to the USB 2.0 connector on the back of the DenOptix QST scanner. Connect the scanner to a grounded power outlet. Make sure the other end of the USB 2.0 cable is connected to the...
  • Page 27: Network Installation

    Gendex DenOptix® QST 4.11 Network installation Refer to your network/computer professional for network installation, configuration and maintenance. Refer to your Imaging Software User Manual and Installation Guide for more information on how to configure the Imaging Software in a network.
  • Page 28 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 29: System Operating Instructions

    80% of the dose required for D-speed intraoral film. More details on how this technology works can be found in appendix A. Almost any intrao- ral or panoramic X-ray unit can be used with the DenOptix QST system. There is no need to purchase a new one.
  • Page 30 Panoramic/cephalometric equipment The DenOptix QST system requires no adjustments to the panoramic/cephalometric exposure settings (kV, mA or time). The DenOptix QST system can be used with almost any existing panoramic/cephalometric unit, regardless of age or manufacturer. The use of a cassette without an intensifying screen is highly recommended. If you choose to use a cassette with an intensifying screen, you must at a minimum remove the front screen.
  • Page 31: Erasing The Imaging Plate

    • If you have a soft vinyl cassette, simply remove both intensifying screens from the cassette. • If you have a hard sided cassette, remove intensifying screens (these are usually glued). • Cassettes without intensifying screens are also available from Gendex. Contact your local dental dealer or the panoramic/cephalometric equipment manufacturer for additional assistance.
  • Page 32: Infection Control

    Gendex DenOptix® QST Infection control Prior to use, the DenOptix QST I/O imaging plates must be placed in barrier envelopes to minimize the possibility of cross-contamination. With proper application and use of the barriers, it will not be necessary to routinely cold sterilize the imaging plates. After erasing...
  • Page 33: Taking An X-Ray

    Gendex DenOptix® QST Taking an X-ray Figure 5-3 Ensure the computer, monitor and DenOptix QST scanner are switched on and properly connected. The green LED indicates that the system is ready to scan. Figure 5-4 Launch your imaging software. Refer to your Imaging Software User Manual for details (VixWin shown).
  • Page 34 Repeat until all I/O imaging plates are mounted. Figure 5-8 Open the carousel well cover on the DenOptix QST scanner and insert the carousel. Close the lid. You are now ready to scan your I/O imaging plates, as outlined in Section 5.5.
  • Page 35 Gendex DenOptix® QST Panoramic/cephalometric imaging Figure 5-9 Place the erased panoramic or cephalomet- ric imaging plate into the cassette with the white side pointed toward the tubehead. Insert the cassette into the panoramic or cephalometric unit and expose in the usual manner.
  • Page 36 Gendex DenOptix® QST Figure 5-12 Wrap the imaging plate around the carousel and push the free end under the edge of the other clip. Push until you hear a “click.” Panoramic shown. Figure 5-13 Open the carousel well cover on the DenOptix QST scanner and insert the carousel.
  • Page 37: Scanning The Imaging Plate

    Gendex DenOptix® QST Scanning the imaging plate Figure 5-14 Once the scanner has been loaded and the patient file has been opened, click on the scanner icon on the imaging software toolbar. The software will automatically sense whether an FMX, EO or Ceph carousel is loaded.
  • Page 38 Gendex DenOptix® QST Figure 5-17 The procedure for scanning a panoramic/cephalometric plate is the same as for the intraorals. It is possible to select how many plates to scan, and what size. As with intraorals, you can choose between 150, 300, and 600 DPI.
  • Page 39: Preparing For The Next Patient

    If the outside of the DenOptix QST scanner has been touched by a contaminated surface, it must be cleaned. BEFORE PROCEEDING, TURN THE SCANNER OFF AND DISCONNECT FROM THE POWER OUTLET.
  • Page 40: Turning The Denoptix Qst System On And Off

    Turning the DenOptix QST system on and off The DenOptix QST System is designed to be left on continuously. The laser is only operational while the unit is actually scanning. If you wish to turn the system off, please remove the carousel and use the following procedure.
  • Page 41 The DenOptix QST imaging systems can operate with different imaging software. For example, VixWin imaging software from Gendex optimizes digital X-ray images from the DenOptix QST systems. Please refer to your Imaging Software User Manual and Installation Guide for detailed instructions.
  • Page 42 Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 43: Maintenance Procedures

    Maintenance procedures The DenOptix QST System is designed for many years of trouble-free operation. It is manufac- tured from the highest quality components to ensure excellent performance. Maintenance that can be performed by the operator is minimal. Cleaning the system If the scanner, imaging plates or carousel becomes contaminated, clean in accordance with the directions given in Section 5.6.
  • Page 44 Gendex DenOptix® QST Cleaning the carousel Figure 7-1 Remove the intraoral imaging plate holder from the carousel by removing the three screws along the vertical plastic retention piece. After removing the three screws, the intraoral plate holder will come free from the carousel.
  • Page 45 Gendex DenOptix® QST Cleaning the EO/CH (panoramic and cephalometric) carousel Figure 7-4 To clean the EO/CH carousel, you will follow similar steps to the intraoral carousel. Remove the carousel imaging plate holder by removing the three screws of the plastic retention piece on both sides of the carousel.
  • Page 46: Operator Maintenance

    Gendex DenOptix® QST Operator maintenance Scanner maintenance The only part of the DenOptix QST scanner that may require operator maintenance is the fuse. If you experience any other service-related problem, contact your local Gendex dealer. Follow these directions to change the fuse.
  • Page 47 Section 5.7. Carousel maintenance The carousel from the DenOptix QST scanner can be maintained by the operator. If the intra- oral imaging plate holder on an intraoral, panoramic or cephalometric carousel becomes damaged, it can be replaced by the operator. Order a replacement holder from your local Gendex dealer.
  • Page 48: Troubleshooting

    Gendex DenOptix® QST Troubleshooting Corrective action Probable cause Trouble Check power cord at outlet and Scanner not plugged in No power at the connection on the back of No green light the scanner. Turn on the scan- ner, then turn on the computer and monitor.
  • Page 49 The computer does not Turn off the scanner and recognize that the scan- re-install the DenOptix QST ner is connected or the systems software driver. scanner has not been added to the Device Turn the unit on.
  • Page 50 Transfer imaging plates from their protective barriers to the DenOptix QST scanner as quickly as possible using the Dark Box (within one hour of exposure). Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 51 Imaging plates fall out The imaging plates Practice loading erased imaging into the DenOptix QST have not been loaded plates in full daylight until you scanner carousel well properly. See Section 5.4 become familiar with the load- for more information.
  • Page 52: Disposal Of Waste Materials And Inoperative Parts

    Disposal of waste materials and inoperative parts Disposal regulations vary from country to country. Therefore, it is difficult to give specific instructions on the disposal of DenOptix QST waste materials and inoperative parts. In gen- eral, we believe the following guidelines to be true.
  • Page 53: Storage And Shipment

    Then, follow the guidelines outlined in Section 5. Shipment The DenOptix QST System is designed to be shipped in its original shipping container by normal commercial carriers. Ensure that the carousel has been removed from the scanner and that the carousel well is empty.
  • Page 54 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 55: Warranty Statement

    (2) years from date of purchase. Parts will be repaired or replaced at our option. The DenOptix QST Scanner and Carousels must be installed and operated in accordance with the Gendex written instruc- tions furnished with the unit.
  • Page 56 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 57: Technical Specifications

    15° to 35°C (59° to 95° F); 5%-95% RH NC Storage and Shipping -40° to 70°C (-40° to 158°F) Temperatures Class I Grounded For Indoor Use Only Suitable for continuous operation DenOptix QST intraoral imaging plates Size 0 Size 1 Size 2 Size 3 Size 4 I/O dimensions...
  • Page 58 Gendex DenOptix® QST DenOptix QST panoramic/cephalometric imaging plates Dimensions Pan: 5x12 in; 15x30 cm Ceph: 8x10 in; 18x24 cm Recommended Dose Equal to standard panoramic systems equipped with rare-earth intensifying screens (400 speed) Storage Store in their clear plastic shipping envelope Operating and 18°...
  • Page 59: Appendix

    How storage phosphor imaging plates work Each DenOptix QST Imaging Plate is made up of a very thin layer of tiny storage phosphor crystals that are bonded together and coated on a flexible sheet of plastic. These storage phosphor crystals have the ability to capture the energy of X-rays and store the pattern as a latent image (see diagram).
  • Page 60: Appendix B: Lighting Conditions For Handling Or Erasing Imaging Plates

    The DenOptix QST scanner further excites the phosphor crystals to an unstable state by exposing them to a red laser. The phosphor crystals then release a blue light and return to their stable ground state. The DenOptix QST scanner reads this blue light and, with the imag- ing software, produces an image.
  • Page 61 Additionally, it is very difficult to judge the intensity of sunlight. Loading carousels under sunlight must be avoided. If you cannot darken your mounting area to 20 lux or less, contact your local Gendex representative for suggested mounting times.
  • Page 62 Gendex DenOptix® QST 4. Using the exposure conditions outlined in Section 5.1, expose each imaging plate. If pos- sible, place an object, such as an extracted tooth, in the path of the X-ray beam. 5. Turn on your erasing light source. We recommend a light box or a minimum l00-watt incandescent light bulb (see Section 5.2).
  • Page 63: Appendix C: Optional Printer

    Thermal printers print in shades of grey with at least 64 levels. Thermal printers use a special type of paper which guarantees a print life of about 3 years. Dye sublimation printers Suitable for high printing quality whenever the DenOptix QST System is used in combination with an intraoral camera. Laser printers If a Laser Printer is adopted, contact the Printer manufacturer or a representative to make sure that it can achieve at least 256 gray shades.
  • Page 64: Appendix D: If You Need Assistance

    11.4 Appendix D: If you need assistance The DenOptix QST System is designed to provide years of trouble-free service, and carries a limited warranty. If you need assistance, first contact your local Gendex dealer. They have been trained to handle most technical service issues. If you cannot get the information you need there, simply call the Gendex office nearest to you.
  • Page 65 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 66 Gendex DenOptix® QST Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 67 Doc # M010-004WWE July 2005 User and Service Manual...
  • Page 68 © 2005 Gendex Dental Systems – M010-004WWE...