  • Page 1 Handbook Handbook Handbook Handbook Signal converter for electromagnetic flowmeters ER 3.2.xx (SW.REV. 3.2.xx) The documentation is only complete when used in combination with the relevant documentation for the sensor. © KROHNE 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 2 KROHNE Messtechnik GmbH & Co. KG. Subject to change without notice. Copyright 2009 by KROHNE Messtechnik GmbH & Co. KG - Ludwig-Krohne-Straße 5 - 47058 Duisburg www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 3: Table Of Contents

    4.3 Electrical cables for remote device versions, notes............25 4.3.1 Notes on signal cables A and B .................... 25 4.3.2 Notes on field current cable C....................25 4.3.3 Requirements for signal cables provided by the customer ..........26 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 4 4.11.4 Modular inputs/outputs and bus systems ................68 4.11.5 Exi inputs/outputs ....................... 77 ® 4.11.6 HART connection ......................82 5 Start-up 5.1 Switching on the power ....................84 5.2 Starting the signal converter ..................84 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 5 8.3 Dimensions and weights ....................133 8.3.1 Housing ..........................133 8.3.2 Mounting plate, field housing ..................... 134 8.3.3 Mounting plate, wall-mounted housing ................134 8.4 Flow tables ........................135 8.5 Accuracy ........................137 9 Notes 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 6: Safety Instructions

    MA IFC 300 R02 (SW.REV.3.10 (2.21)) 2007-05-25 ER 3.1.3x 1; 3-I MA IFC 300 R02 (SW.REV.3.10 (2.21)) 2008-05-13 ER 3.2.0x 1; 2-X; 3-X; 4 MA IFC 300 R03 (SW.REV.3.20 (3.00)) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 7: Intended Use

    The manufacturer certifies successful testing of the product by applying the CE marking. DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 8: Safety Instructions From The Manufacturer

    (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 9: Information Concerning The Documentation

    This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 10: Warnings And Symbols Used

    In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 11: Device Description

    1 Device in the version as ordered 2 Documentation (calibration report, Quick Start, CD-Rom with product documentation for measuring sensor and signal converter) 3 Signal cable (only for remote version) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 12: Device Description

    • Remote version (electrical connection to the measuring sensor via field current and signal cable) Figure 2-2: Device versions 1 Compact version 2 Measuring sensor with connection box 3 Field housing 4 Wall-mounted housing 5 19" rack-mounted housing www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 13: Field Housing

    Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin- free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 14: Wall-Mounted Housing

    6 Cable entry for field current cable 7 Cable entry for inputs and outputs 8 Cable entry for power supply 1 Turn lock to the right and open the cover. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 15: Nameplates

    5 Approvals-related pressure and temperature thresholds 6 Power supply; protection category; materials of wetted parts 7 GK/GKL values (measuring sensor constants); size (mm /inches); field frequency 8 Product designation, serial number and manufacturing date 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 16: Remote Version (Example)

    3 GK/GKL values (measuring sensor constants); size (mm /inches); field frequency 4 Power supply 5 Materials of wetted parts 6 Field coil resistance 7 Approvals-related data (e.g. accuracy class, measuring range, temperature thresholds, pressure thresholds and vis- cosity thresholds) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 17: Electrical Connection Data Of Inputs/Outputs (Example Of Basic Version)

    • A = active mode; the signal converter supplies the power for connection of the subsequent devices • P = passive mode; external power supply required for operation of the subsequent devices • N/C = connection terminals not connected 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 18: Installation

    3.5 Mounting of the compact version INFORMATION! The signal converter is mounted directly on the measuring sensor. For installation of the flowmeter, please observe the instructions in the supplied product documentation for the measuring sensor. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 19: Mounting The Field Housing, Remote Version

    3.6.1 Pipe mounting Figure 3-1: Pipe mounting of the field housing 1 Fix the signal converter to the pipe. 2 Fasten the signal using standard U-bolts and washers. 3 Tighten the nuts. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 20: Wall Mounting

    3 Fasten the housing securely to the wall. Mounting multiple devices next to each other a ≥ 600 mm / 23.6" b ≥ 250 mm / 9.8" www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 21: Turning The Display Of The Field Housing Version

    Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin- free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 22: Mounting The Wall-Mounted Housing, Remote Version

    Figure 3-4: Pipe mounting of the wall-mounted housing 1 Fasten the mounting plate to the pipe with standard U-bolts, washers and fastening nuts. 2 Screw the signal converter to the mounting plate with the nuts and washers. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 23: Wall Mounting

    2 Fasten the mounting plate securely to the wall. 3 Screw the signal converter to the mounting plate with the nuts and washers. Mounting multiple devices next to each other a ≥ 240 mm / 9.4" 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 24: Electrical Connections

    In the event of separate delivery or the installation of devices that have not been configurated • together, the signal converter must be set to the DN size and GK/GKL of the measuring sensor, refer to Function tables on page 91 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 25: Electrical Cables For Remote Device Versions, Notes

    NOT NOT be connected in the housing of the signal converter. INFORMATION! The field current cable is not part of the scope of supply. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 26: Requirements For Signal Cables Provided By The Customer

    • Insulated conductor / insulated conductor 1000 V • Insulated conductor / outer shield 1000 V Twisting of the insulated conductors • At least 10 twists per meter, important for screening magnetic fields. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 27: Preparing The Signal And Field Current Cables

    Cu / AWG 20 3 Insulated wire (3), 0.5 mm Cu / AWG 20 4 Outer sheath 5 Insulation layers 6 Stranded drain wire (6) for the outer shield (60) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 28: Preparing Signal Cable A, Connection To Signal Converter

    3 Slide an insulating tube over the stranded drain wire (1). 4 Crimp the wire end ferrules onto the conductors (2, 3) and stranded drain wire. 5 Pull the heat-shrinkable tubing over the prepared signal cable. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 29 4 Crimp the push-on connector onto the stranded drain wire (6). 5 Crimp the wire end ferrules onto the conductors (2, 3) and stranded drain wire (1). 6 Pull the heat-shrinkable tubing over the prepared signal cable. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 30: Length Of Signal Cable A

    1 Maximum length of signal cable A between the measuring sensor and signal converter [m] 2 Maximum length of signal cable A between the measuring sensor and signal converter [ft] 3 Electrical conductivity of the medium being measured [μS/cm] www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 31: Signal Cable B (Type Bts 300), Construction

    • Wire end ferrule to DIN 46 228: E 1.5-8 for the stranded drain wire (1) • 4 wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors 2 and 3 and the stranded drain wires (20, 30) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 32 4 Slide the insulation tubing over the stranded drain wires (1, 20, 30). 5 Crimp the wire end ferrules onto the conductors and stranded drain wires. 6 Pull the heat-shrinkable tubing over the prepared signal cable. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 33 4 Crimp the push-on connector onto the stranded drain wire (6). 5 Crimp the wire end ferrules onto the conductors and stranded drain wires (1, 20, 30). 6 Pull the heat-shrinkable tubing over the prepared signal cable. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 34: Length Of Signal Cable B

    1 Maximum length of signal cable B between the measuring sensor and signal converter [m] 2 Maximum length of signal cable B between the measuring sensor and signal converter [ft] 3 Electrical conductivity of the medium being measured [μS/cm] www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 35: Preparing Field Current Cable C, Connection To Signal Converter

    3 x 0.75 Cu 3 x 18 150...300 500...1000 3 x 1.50 Cu 3 x 14 300...600 1000...2000 3 x 2.50 Cu 3 x 12 1 Cu = copper cross-section 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 36 1 Strip the conductor to dimension a. 2 Remove any shield that is present. 3 Pull a shrinkable tube over the prepared cable. 4 Crimp the wire end ferrules onto the conductors 7, 8 and 9. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 37: Prepare Signal Cable A, Connect To Measuring Sensor

    4 Slide an insulating tube over the stranded drain wire (1). 5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wire (1). 6 Pull the heat-shrinkable tubing over the prepared signal cable. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 38: Preparing Signal Cable B, Connection To Measuring Sensor

    4 Slide an insulating tube over the stranded drain wire (1). 5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wire (1). 6 Pull the heat-shrinkable tubing over the prepared signal cable. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 39: Preparing Field Current Cable C, Connection To Measuring Sensor

    1 Strip the conductor to dimension a. 2 Remove any shield that is present. 3 Pull a shrinkable tube over the prepared cable. 4 Crimp the wire end ferrules onto the conductors 7, 8 and 9. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 40: Connecting The Signal And Field Current Cables

    Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 41: Connecting The Signal And Field Current Cables, Field Housing

    Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin- free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 42: Connecting The Signal And Field Current Cables, Wall-Mounted Housing

    Any shield that is present must NOT NOT be connected. 5 Tighten the screw connections of the cable entry and close the housing cover. INFORMATION! Ensure that the housing gasket is properly fitted, clean and undamaged. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 43: Connecting The Signal And Field Current Cables, 19" Rack-Mounted Housing

    3 Field current cable Figure 4-16: Connection signal cable B and field current cable 1 Signal cable B 2 Shield and insulated wires 2 and 3 3 Field current cable 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 44: Connection Diagram For Measuring Sensor, Field Housing

    1 Electrical terminal compartment in housing of the signal converter for signal and field current cable. 2 Signal cable A 3 Signal cable B 4 Field current cable C 5 Connection box of measuring sensor 6 Functional ground FE www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 45: Connection Diagram For Measuring Sensor, Wall-Mounted Housing

    1 Electrical terminal compartment in housing of the signal converter for signal and field current cable. 2 Signal cable A 3 Signal cable B 4 Field current cable C 5 Connection box of measuring sensor 6 Functional ground FE 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 46: Connection Diagram For Measuring Sensor, 19" Rack-Mounted Housing

    1 Electrical terminal compartment in housing of the signal converter for signal and field current cable. 2 Signal cable A 3 Signal cable B 4 Field current cable C 5 Connection box of measuring sensor 6 Functional ground FE www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 47: Grounding The Measuring Sensor

    ≥ 200 µS/cm Electrical conductivity Signal cable use only A (type DS 300) ≤ 50 m / ≤ 150 ft Signal cable length Ex operation is possible, consult with us first 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 48: Connecting Power, All Housing Variants

    Power supply connection (excluding 19" rack-mounted housing) 1 100...230 VAC (-15% / +10%) 2 24 VDC (-55% / +30%) 3 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 49 (PELV) (as per VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or relevant national regulations). INFORMATION! For 24 VDC, 12 VDC-10% is included in the tolerance range. Power supply connection for 19" rack-mounted housing 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 50: Inputs And Outputs, Overview

    • For hazardous areas, all of the input/output variants for the housing designs C and F with terminal compartment in the Ex-d (pressure-resistant casing) or Ex-e (increased safety) versions can be delivered. • Please refer to the separate instructions for connection and operation of the Ex-devices. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 51: Description Of The Cg Number

    Signal converter monitors cable breaks and short circuits as per EN 60947-5-6. Errors indicated on LCD display. Error messages possible via status output. Active current input Passive current input No additional module installed No further module possible 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 52: Fixed, Non-Alterable Input/Output Versions

    PROFIBUS PA (Ex-i) (Option) D 0 0 FISCO Device FISCO Device D 1 0 active NAMUR passive FISCO Device FISCO Device D 2 0 passive NAMUR passive FISCO Device FISCO Device www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 53 V/D+ V/D- V/D+ V/D- passive FISCO Device FISCO Device E 2 0 passive NAMUR V/D+ V/D- V/D+ V/D- passive FISCO Device FISCO Device 1 function changed by reconnecting 2 changeable 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 54: Alterable Input/Output Versions

    H _ _ max. 2 optional modules for term. A + B Common Sign. B Sign. A (D1) (D0) 1 changeable 2 not active bus terminator 3 active bus terminator www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 55: Description Of The Inputs And Outputs

    For further information refer to Connection diagrams of inputs and outputs on page 63 and refer Technical data on page 121 DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 56: Pulse And Frequency Output

    For further information refer to Connection diagrams of inputs and outputs on page 63 and refer Technical data on page 121 DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 57: Status Output And Limit Switch

    For further information refer to Connection diagrams of inputs and outputs on page 63 and refer Technical data on page 121 DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 58: Control Input

    For further information refer to Connection diagrams of inputs and outputs on page 63 and refer Technical data on page 121 DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 59: Current Input

    For further information refer to Connection diagrams of inputs and outputs on page 63 and refer Technical data on page 121 DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 60: Electrical Connection Of The Inputs And Outputs

    Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin- free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 61: Wall-Mounted Housing, Electrical Connection Of The Inputs And Outputs

    4 Route the connector plugs with the clamped conductors into the sockets provided for that pur- pose. 5 Close the housing cover. INFORMATION! Ensure that the housing gasket is properly fitted, clean and undamaged. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 62: 19" Rack-Mounted Housing, Electrical Connection Of The Inputs And Outputs

    2 Tighten the screw connection of the cable entry securely. 3 Never mount the housing with the cable entries facing upwards. 4 Seal cable entries that are not needed with a plug. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 63: Connection Diagrams Of Inputs And Outputs

    Control input active acc. to NAMUR EN 60947-5-6: Signal converter monitors cable breaks and short circuits acc. to EN 60947-5-6. Errors indicated on LCD display. Error messages possible via status output. Current input active or passive 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 64: Description Of The Electrical Symbols

    Electronic or electromagnetic counter At frequencies above 100 Hz, shielded cables must be used to connect the counters. Internal resistance of the counter Button, NO contact or similar Table 4-1: Description of symbols www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 65: Basic Inputs/Outputs

    = 24 VDC nominal int,nom ≤ 32 VDC • U • I ≤ 22 mA ≥ 1.8 • U ≤ (U • R ) / I Figure 4-26: Current output passive I 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 66 ) / I L, min • Can also be set as a status output; for the electrical connection, see status output connection diagram. Figure 4-27: Pulse frequency output passive P www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 67 Contact closed (on): U = 2.8 mA • Can also be set as a status output; for the electrical connection, see status output connection diagram. Figure 4-29: Control input passive C 1 Signal 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 68: Modular Inputs/Outputs And Bus Systems

    ≤ (U • R ) / I • X designates the connection terminals A, B or C, depending on the version of the signal converter. Figure 4-31: Current output passive I www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 69 ) / I L, min • X designates the connection terminals A, B or D, depending on the version of the signal converter. Figure 4-32: Pulse / frequency output active P 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 70 • Can also be set as a status output; see status output connection diagram. • X designates the connection terminals A, B or D, depending on the version of the signal converter. Figure 4-33: Pulse frequency output passive P www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 71 • X designates the connection terminals A, B or D, depending on the version of the signal converter. Figure 4-34: Pulse and frequency output passive P to NAMUR EN 60947-5-6 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 72 • The output is open when the device is de-energized. • X designates the connection terminals A, B or D, depending on the version of the signal converter. Figure 4-36: Status output / limit switch passive S www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 73 • X designates the connection terminals A, B or D, depending on the version of the signal converter. Figure 4-37: Status output / limit switch S to NAMUR EN 60947-5-6 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 74 Contact closed (on): U = 1.9 mA • X designates the connection terminals A or B, depending on the version of the signal converter. Figure 4-39: Control input passive C 1 Signal www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 75 ≤ 1.2 V with I ≥ 6.7 mA • X designates the connection terminals A or B, depending on the version of the signal converter. Figure 4-40: Control input active C to NAMUR EN 60947-5-6 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 76 • X designates the connection terminals A or B, depending on the version of the signal converter. Figure 4-42: Current input passive IIn 1 Signal 2 2-wire transmitter (e.g. temperature) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 77: Exi Inputs/Outputs

    = (U ) / I L, min • X designates the connection terminals A or C, depending on the version of the signal converter. Figure 4-44: Current output passive I 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 78 • X designates the connection terminals B or D, depending on the version of the signal converter. Figure 4-45: Pulse and frequency output passive P as per NAMUR EN 60947-5-6 Exi www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 79 • The output is closed when the device is de-energized. • X designates the connection terminals B or D, depending on the version of the signal converter. Figure 4-46: Status output/limit switch S to NAMUR EN 60947-5-6 Exi 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 80 Contact open (off): U ≥ 5.5 V with I ≥ 4 mA Contact closed (on): U • X designates the connection terminals B, if available. Figure 4-47: Control input passive C 1 Signal www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 81 • X designates the connection terminals A or B, depending on the version of the signal converter. Figure 4-49: Current input passive IIn 1 Signal 2 2-wire transmitter (e.g. temperature) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 82: Hart ® Connection

    1 Basic I/O: terminals A and A+ 2 Modular I/O: terminals C- and C ® 3 HART communicator communicator must be R ≥ 230 Ω. ® The parallel resistance to the HART www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 83 Figure 4-51: HART connection passive (I 1 Basic I/O: terminals A- and A 2 Modular I/O: terminals C- and C ® 3 HART communicator ® 4 Other HART - capable devices 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 84: Start-Up

    It is possible to change between the two measured value windows, the trend display and the list with the status messages by pressing the keys ↑ and ↓. Possible status messages, their meaning Status messages and diagnostic information and cause refer to on page 113. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 85: Operation

    Esc (> + ↑) Return to menu mode Return to sub-menu or without acceptance of function without data acceptance of data Table 6-1: Description of key functionality 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 86: Display In Measuring Mode With 2 Or 3 Measured Values

    (_ _ _ signalise in this line the end of the list) 6 Current menu, sub-menu or function 7 Previous menu, sub-menu or function (_ _ _ signalise in this line the beginning of the list) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 87: Display When Setting Parameters, 4 Lines

    3 Denotes the change of a parameter (simple to check changed data when browsing through lists) 4 Next parameter 5 Currently set data from 6 6 Current parameter (for selection press key >; then see previous chapter) 7 Factory setting of parameter (non-alterable) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 88: Using An Ir Interface (Option)

    Upon activation, the LED 3 lights up and the optical keys no longer function. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 89: Menu Structure

    A5.4 low flow cutoff A6 GDC IR interface A7 process input B test > B1 simulation > B2 actual values B3 information ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ > 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 90 5.3 1. meas. page 5.4 2. meas. page 5.5 graphic page 5.6 special functions 5.7 units 5.8 HART 5.9 quick setup ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ > www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 91: Function Tables

    1) Setting for main current output (applicable to all flow measurements) Setting: xxx.x s (range: 000.1…100 s 2) Use for all outputs? Make setting, see Fct. A4.1 above! 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 92 Set coil resistance: break (return with ^ key) Standard (= setting of Fct. C1.1.7) automatic (= calibration with the current resistance) Reference value for on-site calibration; range: 1.000...9999 μS/cm A7.8 target conduct. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 93 2; 4/3; 2/3; 1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50 A7.11 flow direction Define polarity of flow direction. forwards (according to the arrow on the measuring sensor) or backwards (in the opposite direction to the arrow) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 94: Menu B, Test

    See also Fct. C1.1.10...C1.1.12 B2.8 act. coil resistance Current resistance of the field coils according to the current coil temperature. B2.9 current input A Displays the active current value. B2.10 current input B www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 95: Menu C, Setup

    Standard (= 20°C) automatic (set current temperature); range: -40.0…+200°C Set coil resistance: break (return with ^ key) Standard (= setting of Fct. C1.1.7) automatic (= calibration with the current resistance) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 96 (measuring & setting; for DC systems fixed setting 50 Hz) Select: 50 Hz or 60 Hz (fixed setting) C1.1.17 act. coil resistance Display of the current value for calculation of the temperature. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 97 Sets output value of all outputs to "0": x.xxx ± x.xxx m/s (ft/s); range: 0.0…20 m/s (0.0...32.8 ft/s) (1st value = switching point / 2nd value = hysteresis), condition: 2nd value ≤ 1st value 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 98 Only available when electrode noise "on" activated in Fct. C1.3.13. Activation takes place only after setting mode is exited! C1.3.16 settling of field Automatic test switched off / on, select: off / on www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 99 Select: off (switched off) / current output / status output / limit switch C2.1.4 terminal D Select: off (switched off) / frequency output / pulse output / status output / limit switch 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 100 2nd value ≤ 1st value C2. .13 information Serial number of the I/O circuit board, software version number and production date of the circuit board C2. .14 simulation Sequence see B1. current output X www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 101 Select: off (no phase shift) / 0° phase shift (between outputs A or D and B, inversion possible) / 90° phase shift (between outputs A or D and B, inversion possible) / 180° phase shift (between outputs A or D and B, inversion possible) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 102 Select: off (no phase shift) / 0° phase shift (between outputs A or D and B, inversion possible) / 90° phase shift (between outputs A or D and B, inversion possible) / 180° phase shift (between outputs A or D and B, inversion possible) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 103 Only appears if output A...D is set under "mode (see above)" and this output is switched off. C2. .3 invert signal off (activated output supplies a high current, switch closed) on (activated output supplies a low current, switch open) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 104 C2. .3 information Serial number of the I/O circuit board, software version number and production date of the circuit board C2. .4 simulation Sequence see B 1. control input X www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 105 Trimming of the current at 4 mA Reset to 4 mA restores the factory calibration. C2. .9 20mA trimming Trimming of the current at 20 mA Reset to 20 mA restores the factory calibration. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 106 Select: no (exits the function without starting the counter) / yes (starts the counter and exits the function) C3. .10 information Serial number of the I/O circuit board, software version number and production date of the circuit board www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 107 Serial no. of the circuit board, version no. of the main software, production date of the circuit board C5.1.6 Electronic Revision ER Shows ID no., electronic revision no. and production date; Contains all hardware and software changes. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 108 / conductivity / counter 1 / counter 2 / counter 3 / current input A / current input B C5. .11 format 3.line Specify decimal places. Select: automatic (adaptation carried out automatically) / X (= none) …X.XXXXXXXX (max. 8 digits) www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 109 (the IR interface (adapter) and interrupt the optical keys) If about 60 seconds pass without a connection being established, then the function is exited and the optical keys are active once again. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 110 Specification of the conversion factor, based on kg/m³: Set free units xxx.xxx refer to on page 111 C5.7.19 pressure Pa; kPa; bar; mbar; psi (no free units possible); only if current input available. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 111: Set Free Units

    Conversion factors Conversion factors Desired unit = [unit see above] * conversion factor Conversion factor Max. 9 digits ↑ to the left and ↓ to the right Shift decimal point: 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 112: Description Of Functions

    2.5...0.0 s, after that release the key. > language 2 x ↓ reset > reset errors > reset? ↓ or ↑ reset? reset errors Error has been reset. 3 x ^ Measuring operation www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 113: Status Messages And Diagnostic Information

    After module change, confirm query for changed configuration. If device configuration unchanged: defective, replace electronic unit. F hardware detection Existing hardware cannot be identified. Replace electronic unit. Defective or unknown modules. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 114 If frequency output: extend values in measured value is limited by a filter setting. Fct. C2.x.5 and Fct. C2.x.7. F over range B Or the demanded pulse rate is too high. F over range C www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 115 S electrode symmetry Impedance of the two measuring electrodes Deposits in measuring tube or electrode not equal. Measured values are still short-circuit to ground. Clean and check supplied. measuring tube! 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 116 C test sensor Test function of the measuring sensor electronics is active. C simulation fieldbus Values on the Foundation Fieldbus interface are simulated. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 117 Profibus. I conductivity off Conductivity measurement switched off. Changing of settings in Fct. C1.3.1. I diagnosis channel off Diagnosis value switched off. Changing of settings in Fct. C1.3.17. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 118: Service

    • such dangerous substances, • to enclose a certificate with the device confirming that is safe to handle and stating the product used. www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 119: Form (For Copying) To Accompany A Returned Device

    We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned. Date: Signature: Stamp: 7.4 Disposal CAUTION! Disposal must be carried out in accordance with legislation applicable in your country. 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 120: Technical Data

    (separate from noise) and convert it into signals for totalising, recording and output processing. 1 Voltage (induced voltage proportional to flow velocity) 2 Electrodes 3 Magnetic field 4 Field coils www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 121: Technical Data

    2 (optional 3) internal counters with a max. of 8 counter places (e.g. for counting volume and/or mass units) Verification Integrated verification, diagnostic functions: measuring device, process, measured value, empty pipe detection, stabilization Communication interfaces ® Foundation Fieldbus, Profibus PA and DP, Modbus, HART 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 122 ±0.15% of the measured value ±1 mm/s, depending on the measuring sensor For detailed information and accuracy curves, see chapter "Accuracy". Repeatability ±0.06% to OIML R117; not valid for WATERFLUX 3000 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 123: Operating Conditions

    For detailed information, refer to chapter "Installation conditions". Inlet/outlet runs Refer to technical data for the measuring sensor. Dimensions and weights For detailed information refer to chapter "Dimensions and weights". 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 124 WSC (only for WATERFLUX 3000) WSC (only for WATERFLUX 3000) Max. length: 25 m / 82 ft Cable entries Standard: M20 x 1.5 (8...12 mm) Option: ½" NPT, PF ½ www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 125: Inputs And Outputs

    ≥ 1.8 V ≥ 4 V ≤ (U ≤ (U ) / I ) / I = 30 V = 100 mA = 1 W = 10 nF ~ 0 mH 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 126 = 22.5 V 0, nom at I = 1 mA = 21.5 V 0, nom at I = 10 mA = 19 V 0, nom at I = 20 mA www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 127 The time constant corresponds to the elapsed time until 67% of the end value has been reached according to a step function. Settings Set in increments of 0.1. 0…100 s 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 128 Passive to EN 60947-5-6 open: open: = 0.6 mA = 0.43 mA closed: closed: = 3.8 mA = 4.5 mA = 30 V = 100 mA = 1 W =10 nF =0 mH www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 129 I < 1.9 mA Detection of cable break: ≥ 8.1 V with I ≤ 0.1 mA Detection of cable short circuit: ≤ 1.2 V with I ≥ 6.7 mA 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 130 = 30 V = 5 V = 100 mA 0, max at I ≤ 22 mA = 1 W = 10 nF ® No HART =0 mH ® No HART www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 131 Modbus RTU, Master / Slave, RS485 Address range 1…247 Supported function codes 03, 04, 16 Broadcast Supported with function code 16 Supported Baudrate 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 132 Electromagnetic compatibility 89/336/EEC and 93/68/EEC in conjunction with EN 61326-1 (A1, A2) (EMC) European Pressure Equipment PED 97/23 (only for compact versions) Directive NAMUR NE 21, NE 43, NE 53 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 133: Dimensions And Weights

    Dimensions and weights in inches and lbs Version Dimensions [inches] Weight [lbs] 7.75 4.75 6.10 10.20 5.40 9.30 7.75 4.75 6.10 11.60 10.90 12.60 7.80 5.40 11.80 5.30 5.59 5.08 7.68 2.65 (28 TE) (3 HE) 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 134: Mounting Plate, Field Housing

    IFC 300 8.3.2 Mounting plate, field housing Dimensions in mm and inches [mm] [inches] Ø9 Ø0.4 8.3.3 Mounting plate, wall-mounted housing Dimensions in mm and inches [mm] [inches] Ø9 Ø0.4 3.85 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 135: Flow Tables

    2000 3393.00 11310.00 33930.00 135720.00 2200 4105.50 13685.00 41055.00 164220.00 2400 4885.80 16286.00 48858.00 195432.00 2600 5733.90 19113.00 57339.00 229356.00 2800 6650.10 22167.00 66501.00 266004.00 3000 7634.10 25447.00 76341.00 305364.00 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 136 95606.51 382426.03 12100.27 40333.83 121002.69 484010.75 14938.92 49795.90 149389.29 597557.18 18075.97 60252.63 180759.73 723038.90 21511.53 71704.38 215115.30 860461.20 25245.60 84151.16 252456.02 1009824.08 29279.51 97597.39 292795.09 1171180.37 33611.93 112038.64 336119.31 1344477.23 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...
  • Page 137: Accuracy

    0.3% of mv + 2 mm/s OPTIFLUX 4300 / 5300 / 6300 <10 <3/8 0.3% of mv + 2 mm/s WATERFLUX 3300 50...600 2...24 0.2% of mv + 1 mm/s 06/2009 - 4000069802 - MA IFC 300 R03 en www.krohne.com...
  • Page 138: Notes

    NOTES IFC 300 www.krohne.com 06/2009 - 4000069802 - MA IFC 300 R03 en...