5 ADDITIONAL INFORMATION
5.1
Maintenance
camSPECS does not need any special maintenance.
•
Do not touch or pollute the surfaces of the filters.
•
Remove dust on the filters with pressurized-air spray.
•
Remove undesired fingerprints or fat on the filters carefully with a soft and dry tissue.
If the bulb is defective, replace it with the replacement bulb:
•
Open the sliding port on the top side of camSPECS.
•
Remove the top cover of the lamp housing by releasing the four screws with an Allen key.
•
Remove the old bulb.
•
Insert the new bulb. Do not touch the new bulb with bare hands or fingertips, use a tissue
or fabric gloves.
The calibration device requires a recalibration yearly, regardless of the operating hours. If calibration is
necessary, please contact Image Engineering.
•
Do not touch, scratch, or pollute the measuring head.
•
Do not remove the fiber from the calibration device; otherwise, the calibration is invalid.
5.2
Disposal instructions
After the service life of camSPECS and the spectroradiometer, they must be disposed of properly.
Electrical and electromechanical components are included in camSPECS and the spectroradiometer.
Observe all national regulations and ensure that camSPECS cannot be used by third parties after
disposing of it.
Contact Image Engineering if assistance for disposal is required.
Image Engineering
Page 35 of 36