Serial Commands; Service Code; Aspoll - TSI Instruments DUSTTRAK 8520 Application Note

Aerosol monitor
Hide thumbs Also See for DUSTTRAK 8520:
Table of Contents

Advertisement

Baud Rate:
Data Bits:
Stop Bits:
Hardware Control:

Serial Commands

Command
(Upper
Returned
Case)
Value
Description
Returns a seven character string corresponding to any service conditions which may be
ASRVCK
sssssss
present. See below for detailed description.
Returns the seven character string displayed on the D

ASPOLL

±vvv.vvv
digits and a decimal point followed by three more digits)
Returns a reading every xx seconds averaged over that time interval, where xx can range
from 01 seconds to 60 seconds. The seven character string is in mg/m
ASDATAxx
±vvv.vvv
as "ASPOLL").
NOTE: THE ASDATA FUNCTION DOES NOT WORK PROPERLY FOR D
MONITORS WITH FIRMWARE REVISIONS BELOW v1.9.
AQDATA
none
Terminates the "ASDATAxx" output stream.
ASRVCK
The D
T
monitor can be polled for any service conditions which may be present using the
UST
RAK
command "ASRVCK". This command can be issued to the D
condition which may be present during startup, as well as any time during normal operation. The
D
T
monitor will return a seven character string which corresponds to the seven service
UST
RAK
conditions which may arise. For example a returned string of "0000000" corresponds to no service
conditions (normal operation), a string of "0000001" corresponds to a service condition 1, while a string of
"7000300" corresponds to a service condition 7 and a service condition 3. The returned string may
contain any combination of the seven service codes when they are present.
Service
Code
Service Code Description
1
Memory has been cleared due to temporary loss of power from internal backup battery.
2
Calibration memory error. Internal calibration data corrupted.
3
Backup battery low.
4
The sample inlet nozzle is due for cleaning.
5
Internal filters are due for replacement.
6
The internal pump is failing or has failed.
7
Laser failure.
ASPOLL
The D
T
monitor can be polled for the current aerosol concentration, averaged over the time
UST
RAK
constant being used by the D
character string (three digits, a decimal point and three more digits) represents the average mass sum in
3
mg/m
. If the average mass is negative, then the seven character string will be preceded by a minus sign.
1200
8
1
None
T
monitor, with the command "ASPOLL". The returned seven
UST
RAK
-2-
T
LCD in mg/m
UST
RAK
T
monitor to clear any service
UST
RAK
3
. (three
3
(format is the same
T
UST
RAK

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DUSTTRAK 8520 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents