Measuring Dc Voltage; Connecting The Dc Voltage Signal; Input Ranges; Optimizing Measurements - National Instruments NI 435 Series User Manual

High-precision temperature and voltage meters
Table of Contents

Advertisement

Measuring DC Voltage

Connecting the DC Voltage Signal

To prevent possible safety hazards, the maximum voltage between any of the
Caution
analog inputs and the computer ground should never exceed ±42 VDC when the NI 435x
is powered on and ±17 VDC when the NI 435x is powered off.

Input Ranges

If scanning voltages in different ranges, the NI 435x uses the widest range for all
Note
channels.

Optimizing Measurements

© National Instruments Corporation
minimize temperature gradients in the space enclosing the thermocouples,
the NI 435x, and its accessories.
The NI 435x accessories—the TBX-68T, CB-68T, and TBX-68 for the
NI 435x for USB, PXI, and PCI—are designed to be used with any
DC voltage signal. Consult the accessory installation guide for instructions
on how to connect the voltage signals.
The NI 435x analog inputs are protected against damage from voltages
within ±42 VDC in all ranges when powered on and ±17 VDC when the
NI 435x is powered off. Never apply voltages above these levels to the
inputs.
The NI 435x has six bipolar input ranges available for measuring DC
voltage. These ranges are ±625 mV, ±1.25 V, ±2.5 V, ±3.75 V, ±7.5 V,
and ±15 V. The NI 435x can measure DC voltage to the specified accuracy
as long as the voltage is within the selected input range. To get the best
resolution, noise rejection, and accuracy, choose the smallest possible
range. Make sure that each signal input to CH+ and CH– is within the input
common-mode limits of this input range. The input common-mode limits
are ±2.5 V and ±15 V for the lower three and higher three input ranges,
respectively.
To make accurate voltage measurements, program the onboard
ground-referencing and open-thermocouple detection appropriately. Also
consider problems associated with AC noise effects, thermal EMFs, and
other errors as discussed in the following sections.