VS 8/VS 8A
Vital Signs Monitor
Operator's Manual
To Purchase, Visit
Avobus.com
or call
1-800-674-3655

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VS 8 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mindray VS 8

  • Page 1 VS 8/VS 8A Vital Signs Monitor Operator’s Manual To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 2 SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) Release time: January 2022 owns the intellectual property rights to this Mindray product and this manual. This Revision: 2.0 manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
  • Page 3 This manual is based on the maximum configuration and therefore some contents may Please include the instrument model number, the serial number, and a description of the not apply to the product. If you have any question, please contact Mindray. problem with all requests for service.
  • Page 4: Table Of Contents

    3.9.5 Setting the Date and Time ...................... 3 - 17 5.10 Nurse Call ..............................5 - 9 3.9.6 Checking Software Licenses ....................3 - 18 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual To Purchase, Visit Avobus.com...
  • Page 5 11 Manual Parameters ........................11 - 1 8.4.6 Troubleshooting ......................... 8 - 10 11.1 Manual Input Parameters ........................11 - 1 8.4.7 Cleaning the TemporalScanner™ Thermometer ............8 - 10 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 6 14.2.2 Example Review Dialog ......................14 - 1 16.6.3 Enabling Black Standby Screen ..................16 - 14 14.2.3 Reviewing the Tabular Trends .................... 14 - 2 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 7 16.13.1 Setting Scanner Information ...................16 - 21 18.3 Cleaning and Disinfecting the Equipment ..................18 - 2 16.13.2 Identifying the Barcode Reader (for the non-Mindray Custom 2D Barcode Reader) 18.3.1 Approved Cleaning and Disinfecting Agents ..............18 - 2 16 - 21 18.3.2 Cleaning the Equipment .......................
  • Page 8 B.2 Radio Regulatory Compliance .........................B - 7 C Default Settings .........................C - 1 C.1 Parameters Default Settings ........................C - 1 C.1.1 SpO2 Default Settings ........................ C - 1 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 9 CAUTION • Use only parts and accessories specified in this manual. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 1 - 1 1 - 2...
  • Page 10 1.1.3 Notes USB connector Equipotentiality NOTE • The equipment software copyright is solely owned by Mindray. No DEFIBRILLATION-PROOF TYPE BF DEFIBRILLATION-PROOF TYPE organization or individual shall resort to modifying, copying, or exchanging APPLIED PART CF APPLIED PART it or to any other infringement on it in any form or by any means without due permission.
  • Page 11 European Community Manufacturer MR Unsafe (magnetic resonance) TrueBP is Mindray's new non-invasive blood pressure measurement algorithm using linear inflation technology, which can measure the blood pressure quickly and comfortably during cuff inflation. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 12 The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 8/VS 8A Vital Signs Monitors are not intended for helicopter transport, hospital ambulance, or home use.
  • Page 13 USB connectors Connect approved USB devices, for example the keyboard, mouse, and barcode reader. Network connector VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 2 - 3 2 - 4 To Purchase, Visit Avobus.com...
  • Page 14 When using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential difference between them. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 2 - 5...
  • Page 15 Passwords should only be changed by authorized personnel. Contact to the IEC 60601-1. If you have any questions, please contact Mindray. the department manager or biomedical engineering department for the •...
  • Page 16 To use the AC power source, follow this procedure: Press the power switch to turn on the monitor. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 3...
  • Page 17 Tapping the screen Do not use the monitor on a patient if you suspect it is not working properly, or if it is mechanically damaged. Contact the service personnel or Mindray. ◆ To select an item from menus or lists, or select a quick key, tap on it with your finger.
  • Page 18 The barcode reader is connected to the monitor through the USB connector. If you are using the Mindray custom 2D barcode reader (Model HS-1R, HS-1RS or HS-1M), before using it for the first time, clear old data formats and configure the barcode reader.
  • Page 19 Wired network is Wired network is not connected. connected. The following table shows available quick keys. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 9 3 - 10 To Purchase, Visit Avobus.com...
  • Page 20 Select the Main Menu quick key → from the Configuration column select ■ Manage → input the required password → select → Workflow Setup. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 11 3 - 12...
  • Page 21 Monitoring NIBP monitoring on the same limb. The SpO alarm status is locked until the NIBP measurement ends. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 13 3 - 14...
  • Page 22 Brightness On Battery. You can set the brightness to 1 to 10, in parameter which 10 is the brightest, and 1 is the least bright. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 15...
  • Page 23: Start Working

    There will be no physiological alarm related indications. ■ The alarm setup menus will not be displayed. ■ Alarm limits cannot be set. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 17 3 - 18...
  • Page 24: Inputting Patient Information

    You can connect the monitor to the BeneVision CMS through wired LAN or wireless LAN. after the Auto Enter Standby Time. When connected to the CMS, the system provides the following function. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 19...
  • Page 25: Connecting The Egateway

    Stopping a Parameter Measurement To stop monitoring a parameter, follow this procedure: Remove corresponding sensors from the patient. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 21 3 - 22 To Purchase, Visit Avobus.com...
  • Page 26: Managing Patients

    30 minutes, you will be prompted whether to start monitoring a new patient if any of the above vital signs are detected again. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 4 - 1...
  • Page 27: Manually Discharging A Patient

    Select the Patient quick key. Select Select Patient. From the patient list select desired patients. This page intentionally left blank. Select Delete. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 4 - 3 4 - 4...
  • Page 28: Alarm Introduction

    + double + triple beeps pattern triple + double beeps VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 1 5 - 2 To Purchase, Visit Avobus.com or call...
  • Page 29: Alarm Status Symbols

    Every time you switch to a workflow, the alarm settings of the workflow are loaded. For details, see 3.9.1.4 Changing Alarm Settings. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 30: Changing The Alarm Volume

    ■ Physiological alarms are switched off. The alarm lamp does not flash and alarm sound is not issued. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 5 5 - 6...
  • Page 31: Pausing Alarm Sound

    Pause Time), pressing the Audio Pause quick key permanently switches off all alarm sound. The audio off state has the following features: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 7...
  • Page 32: Latching Alarms

    The nurse call system is enabled. ■ A user-defined alarm occurs. ■ Alarms are not paused or reset. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 9 5 - 10...
  • Page 33: Monitoring Pulse Oxygen Saturation (Spo )

    • A functional tester or SpO simulator cannot be used to assess the SpO accuracy. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 1 6 - 2...
  • Page 34: Spo Measurement Limitations

    Cuff or arterial blood pressure measurement device on the same limb as the Patient physiological characteristics: sensor ◆ Cardiac arrest ◆ Hypotension VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 3 6 - 4 To Purchase, Visit Avobus.com or call...
  • Page 35: Spo Display

    Select an appropriate extension cable according to the connector type and plug the cable into the SpO connector. Connect the sensor to the extension cable. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 5 6 - 6...
  • Page 36: Nellcor Satseconds Tm Alarm Management

    Sensitivity (for Masimo SpO For Masimo SpO , selects the Sensitivity as per signal quality and patient motion. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 7 6 - 8...
  • Page 37: Enabling Fastsat (For Masimo Spo )

    The reliability of FastSAT is dependent on the setting for the averaging time and the input signal. FastSAT is disabled by default. To enable FastSAT, follow this procedure: Set Averaging. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 9...
  • Page 38 This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http:// www.covidien.com/patents. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 11 6 - 12...
  • Page 39 Shenzhen Mindray. In no event may you transfer, assign, rent, lease, sell, to this device.
  • Page 40 Changing the PR Alarm Settings In Continuous Monitoring mode, you can change the PR alarm settings. Follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 7 - 1 7 - 2 To Purchase, Visit Avobus.com...
  • Page 41 Wash hands after temperature is taken. This will significantly reduce the risk Temp unit of cross contamination and nosocomial contamination. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 1 8 - 2...
  • Page 42 Axillary. The measurement site cannot be changed. The temperature type automatically changes to Predictive mode when the probe returns to the probe well. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 3...
  • Page 43 There are no parts that you can service yourself except for the battery, which you should replace when low by following the instructions in this manual. For To disinfect the temperature probe: service, repair, or adjustments, return your thermometer to Mindray. Disconnect the temperature probe from Temp connector. •...
  • Page 44 Release button remove from head and read measurement. High Ambient HI A >104 °F (40 °C) NOTE Low Ambient LO A <60 °F (16 °C) VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 7 8 - 8...
  • Page 45 You should clean the product as per your hospital’s regulations after each use. Sterilization is not allowed for the product. ■ Cleaning the case VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 9 8 - 10...
  • Page 46 You can manually input a Temp value and save the data for future reference. Follow this procedure: Select in the Temp area. For monitors configured with Mindray SmarTemp™ module, to input Temp value in Continuous Monitoring mode, press and hold in the Temp area.
  • Page 47 • NIBP measurement can be performed during electro-surgery and discharge of a defibrillator. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 1 9 - 2...
  • Page 48 Manual: measurement on demand. ■ Auto: repeated measurement at set interval or an interval synchronized with the real time clock. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 3 9 - 4...
  • Page 49 For patients taking orthostatic BP measurement, refer to 9.10 Orthostatic BP For more information, see 9.8.9 Correcting the NIBP Measurements. Measurement. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 5...
  • Page 50: Performing Auto Measurement

    Select STAT. Patient position and site of the first measurement Patient position and site of the second measurement VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 7 9 - 8...
  • Page 51: Performing Sequence Measurement

    Select the Setup tab. Select the NIBP numeric area to enter the NIBP dialog. Set Display Format. Set Start Mode. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 9 9 - 10...
  • Page 52: Setting The Nibp Alarm Limits Display Switch

    BP values. under Spot Check. Select Parameters Setup → the NIBP tab. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 11 9 - 12 To Purchase, Visit Avobus.com...
  • Page 53: Changing Bp Averaging Settings

    Select on the right of the workflow to be set. Make sure the workflow is under Spot Check. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 13 9 - 14...
  • Page 54: Performing Orthostatic Bp Measurement

    Maintenance → input the required password → select → Module. Select Parameters Setup → the NIBP tab. Under Orthostatic BP Measurement, set Lying Duration. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 15 9 - 16...
  • Page 55: Assisting Venous Puncture

    If the problem persists, contact your service personnel. NOTE • For the physiological and technical alarm messages, see Appendix D Alarm Messages. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 17 9 - 18 To Purchase, Visit Avobus.com...
  • Page 56: Monitoring Respiration

    RR, follow this procedure: Select the Main Menu quick key → from the Configuration column select Workflow → input the required password → select VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 10 - 1...
  • Page 57: Manual Parameters

    Ventilator, Venturi Mask, Oxymizer LOC (ACVPU) Alert, Confusion, Reacting to voice, Reacting to pain, Unresponsive Supp. O Yes, No VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 11 - 1 11 - 2...
  • Page 58: Changing The Display Of Manual Parameters

    Return to the main screen and switch workflows for the setting to take effect. This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 11 - 3...
  • Page 59: Clinical Assistive Applications (Caa)

    MEWS, NEWS and NEWS2 are intended for adult patients only. The patient category applied to the Custom Score is defined by Mindray Clinical Score Configuration Tool. For more information, see Mindray Clinical Scoring Config Tool Instruction for Use (P/N: 046- 12.1.2...
  • Page 60 Alarm: the monitor automatically starts scoring when an alarm occurs to the parameter for scoring. ◆ If no option is selected, the monitor does not initiate auto scoring. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 3 12 - 4...
  • Page 61 To enable the auto refreshing score function, follow this procedure: viewed. From the EWS page select Setup. Select Filter and choose a protocol if needed. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 5 12 - 6 To Purchase, Visit Avobus.com...
  • Page 62 11.4 Changing the Display of Manual Parameters to add GCS to the list. • For details about how to import scoring tools, refer to Mindray Clinical Scoring Config Tool Instruction for Use (P/N: 046-012986-00). 12.2 Glasgow Coma Scale (GCS) The Glasgow Coma Scale (GCS) function is based on Teasdale’s Assessment of Coma and...
  • Page 63 From the Eye Opening area, Verbal Response area, and Motor Response area, Pain Score is applicable for adult, pediatric, and neonatal patients. respectively select an item that represents the patient’s status. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 9...
  • Page 64 Goal quick key. ■ Select the desired waveform area, parameter numeric area, or dashboard to enter corresponding parameter setup menu. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 11 12 - 12 To Purchase, Visit Avobus.com...
  • Page 65 From the SpO statistics area, select the duration to redefine the length of SpO statistics. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 13 12 - 14...
  • Page 66 Select Recorder Paper Speed to set the speed for recording waveforms. ■ Press the hardkey on the front of the recorder. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 13 - 1 13 - 2...
  • Page 67 If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam. If a paper jam is detected, follow this procedure to remove it: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 68 (Temp, Scoring etc.) or type of data (Manually Saved for example). Select a record to be viewed and select Detail to check more information. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 69 Different color blocks are displayed on the left of each event to indicate different event types. ■ Red: high priority alarm event VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 14 - 3 14 - 4...
  • Page 70: Setting Workflows

    Select the workflows and User Maintenance Settings to import. Select a department. Select OK. Select Import. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 15 - 1 15 - 2 To Purchase, Visit Avobus.com...
  • Page 71: Modifying Configuration Password

    30 minutes, the monitor behaves the same as it is normally turned off. This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 15 - 3...
  • Page 72: Password-Protected User Settings

    Select the Patient Management tab → ADT Query tab. network connector. Select the query criteria as desired. To enable this function, follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 1 16 - 2...
  • Page 73: Auto Discharging A Patient After Monitor Power Off

    From the Maintenance area, set User Maintenance to select the password for data is cleared. accessing the monitor’s Maintenance dialog. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 3 16 - 4 To Purchase, Visit Avobus.com...
  • Page 74: Selecting Password For Changing Alarm Settings

    You can set the monitor to automatically log out of the clinician account after saving data. do so, follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 5...
  • Page 75: Setting Items About Sending Data

    High Alarm Interval 3 s to 15 s 10 s Med Alarm Interval 3 s to 30 s 20 s VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 7 16 - 8...
  • Page 76: Setting Alarm Pausing And Resetting Properties

    Off When Reset: when the alarm system is reset, both the alarm tone and Permanent. The default audio pause time is two minutes. alarm lamp of the current alarms are switched off. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 9...
  • Page 77: Latching Physiological Alarms

    Pulse: the nurse call signal is a pulse signal and each pulse lasts one second. so, follow this procedure: When multiple alarms simultaneously occur, only one pulse signal is VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 11...
  • Page 78: Setting The Alarm Delay Time

    → input the required password → select To do so, follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 13...
  • Page 79: Changing Caa Settings

    Select the Main Menu quick key → from the System column select Maintenance Set high limit, low limit and color for each level. → input the required password → select VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 15...
  • Page 80: Setting Temp Measurement Sites

    From the Pain Score Method area, switch on the pain scales as needed. → input the required password → select Select Module → Other tab. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 17...
  • Page 81: Setting Parameter Units

    Select the Time tab→ Time Synchronization tab. → input the required password → select Switch on Start NTP Time Sync. Select the Version tab. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 19 16 - 20...
  • Page 82 • Wireless network designing, deploying, debugging, and maintenance should To select desired patient information that will be read by the barcode reader, follow this be executed by Mindray service personnel or authorized technicians. procedure: • Always set the wireless network according to local wireless regulations.
  • Page 83 Auto, which means the monitor will automatically use LAN1 IP when detected and the available WLAN when LAN1 IP is not detected. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
  • Page 84 Select the Main Menu quick key → from the System column select Maintenance Set Multicast TTL. → input the required password → select VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 25...
  • Page 85 Set Automatically Logout after as needed. From the General Connectivity area, set the Master Server Address, Context Root, and Organization ID. Enable Local CSI as needed. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 27...
  • Page 86 Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Other tab. Select Export Patient Data. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 29 16 - 30...
  • Page 87 Keep batteries out of the reach of children. • The on-screen battery symbols indicate the battery status as follows: Use only Mindray specified batteries. Use of a different battery may present a risk of fire or explosion. ■ indicates that the battery power is sufficient. The green portion represents •...
  • Page 88 Allow the monitor to run on the battery until the battery is completed and the monitor automatically shuts down. Fully charge the battery again for use or charge it to 40 – 60% for storage. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 17 - 3...
  • Page 89: Checking Battery Performance

    • Storing batteries in a cool place can slow the aging process. Ideally the batteries should be stored at 15 °C. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 17 - 5...
  • Page 90: Care And Cleaning Introduction

    Isopropanol Liquid Isopropanol 70% CAUTION 1-Propanol Liquid 1-Propanol 50% • If you spill liquid on the equipment or accessories, contact Mindray or your Metrex CaviCide1™ Liquid Diisobutylphenoxyethoxyethyl dimethyl service personnel. benzyl ammonium chloride 0.28%, • Isopropanol 17.2% Avoid wetting the pins and metal parts of the equipment, mounting kits or accessories during cleaning and disinfection.
  • Page 91: Cleaning The Equipment

    Germicidal Disposable chlorides 0.125%, n-Alky dimethyl benzyl Cloth ammonium chlorides 0.125% VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 18 - 3 18 - 4 To Purchase, Visit Avobus.com...
  • Page 92: Cleaning The Thermal Print Head

    Never clean or disinfect the metalic connectors at either end of the Disposable Wipe n-Alkyl dimethyl benzyl ammonium chlorides 0.14% accessories. • Use only Mindray approved cleaners and disinfectants and methods listed in PDI Sani-Cloth® Plus Wipes n-Alkyl dimethyl ethylbenzyl ammonium chlorides Germicidal 0.125%,...
  • Page 93: Cleaning The Accessories

    Clorox Healthcare® Bleach Wipes Sodium Hypochlorite 0.55% Germicidal Wipes Clorox Healthcare® Wipes Hydrogen Peroxide 1.4% Hydrogen Peroxide Cleaner Disinfectant Wipes VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 18 - 7 18 - 8...
  • Page 94: Maintenance

    This equipment contains no user serviceable parts. 19.3 Maintenance and Testing Schedule • Software upgrades to the monitor can only be done by Mindray service Follow the maintenance and testing schedule or local regulations to perform testing and engineers with Mindray upgrading tools. •...
  • Page 95: Checking Version Information

    If your monitor needs a safety test and performance test, contact the service personnel. The NIBP accuracy test should be performed once every two years or when you doubt the NIBP measurements. The NIBP accuracy test should be performed by Mindray- 19.5.1 Performing Visual Inspection qualified service personnel only.
  • Page 96 For disposal of parts and accessories, where not otherwise specified, follow local regulations regarding disposal of hospital waste. This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 19 - 5...
  • Page 97: Accessories

    Disposable Infant (3 to therapy is performed. 20 kg) adhesive, thumb or great toe VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 20 - 1 20 - 2 To Purchase, Visit Avobus.com...
  • Page 98 Pediatric/ Neonate LNCS Inf 0600-00-0158 sensor, 20 pcs/ Disposable Infant (3-20 box, adhesive, thumb or great toe VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 20 - 3 20 - 4...
  • Page 99 TAT-5000S Temporal Reusable Adult, pediatric, Scanner neonate 115-031807-00 Cuff starter kit, bladderless Reusable Child/Small adult/ Adult/ Adult thigh VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 20 - 5 20 - 6...
  • Page 100 Barcode scanner kit (1D) 115-039575-00 2D Barcode reader, HS-1M, JADAK 115-039635-00 2D Barcode reader, HS-1R, JADAK VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 20 - 7 20 - 8 To Purchase, Visit Avobus.com...
  • Page 101 Current 0.9 to 0.5 A Frequency 50/60 Hz (±3 Hz) Fuse T2AL-250V VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 1 A - 2...
  • Page 102 Method Thermal dot array remote equipment Paper speed 25 mm/s Number of waveform channels A.5.7 Wireless Network VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 3 A - 4...
  • Page 103 ESU equipment. The 7.0m Co-channel 8.5m 5210MHz) interfering devices do not include Wi-Fi devices. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 5 A - 6 To Purchase, Visit Avobus.com...
  • Page 104 The total delay of data transmission from the monitor to the bluetooth node: ≤ 6 • seconds. Do not use Bluetooth nodes to perform realtime vital signs monitoring. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 7 A - 8...
  • Page 105 2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s Low perfusion Pulse amplitude: >0.02% conditions Light penetration: >5% VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 9 A - 10...
  • Page 106 Mean: 15 to 260 15 to 215 15 to 125 Resolution 1 bpm VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 11 A - 12 To Purchase, Visit Avobus.com or call...
  • Page 107 * Measurement accuracy verification: In adult and pediatric modes, the blood pressure Oral 0.02°C 0.33°C 0.1°C measurements measured with this device are in compliance with the American National VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 13 A - 14...
  • Page 108 Alkaline battery This page intentionally left blank. Voltage Run time Approximately one hour continuous use or 15000 readings VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 15 A - 16...
  • Page 109 Otherwise, degradation of the performance of this device ■ Detect for connection could result. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 1 B - 2 To Purchase, Visit Avobus.com...
  • Page 110 ±1 kV line(s) to ±1 kV line(s) to 61000-4-5 line(s) line(s) ±2 kV line(s) to ±2 kV line(s) to earth earth VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 3 B - 4...
  • Page 111 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 5...
  • Page 112 (2) this device must accept any interference received, including interference that may cause undesired operation. Any changes or modifications to this equipment not expressly approved by Mindray may cause harmful radio frequency interference and void your authority to operate this equipment.
  • Page 113: Temp Default Settings (Smartemp)

    Neonate: 90 mmHg Fast SAT (for Masimo SpO Adult: 90 mmHg Pediatric: 70 mmHg Neonate: 40 mmHg Priority VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 1 C - 2...
  • Page 114 Phase D Duration Continuous Neonate: 15 mmHg Interval 1 hr Priority High Phase E Duration Interval 1 hr VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 3 C - 4...
  • Page 115: Ews Default Settings

    Off, Med Item Default Setting Off, Med Alarm Volume Temp Off, Med VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 5 C - 6 To Purchase, Visit Avobus.com or call...
  • Page 116: System Time Default Settings

    System Time Default Settings This page intentionally left blank. Item Default Setting Date Format yyyy-mm-dd 24-Hour Time Daylight Savings Time VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 7 C - 8...
  • Page 117: D Alarm Messages

    In the following tables, A, B, and C are used to refer to the responses on alarm reset. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 1...
  • Page 118: General Technical Alarm Messages

    Check SpO2 No SpO sensor is connected. Sensor Connect an SpO sensor. Connection VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 3 D - 4 To Purchase, Visit Avobus.com or call...
  • Page 119: Temp Technical Alarm Messages

    If patient category is service personnel. correct, check that the tubing is not bent and the airway is not occluded. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 5 D - 6...
  • Page 120: Ews Technical Alarms

    IP Address wireless network IP address. Check the network settings. Comm Error the alarm persists, contact your service personnel. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 7 D - 8...
  • Page 121: Other System Technical Alarm Messages

    Please patient. This page intentionally left blank. delete some discharged patients. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 9 D - 10 To Purchase, Visit Avobus.com or call...
  • Page 122: The Accuracy Of Masimo Spo

    Table information for the plots below show A values measured with Masimo Set Oximetry Technology in a clinical study. Adtx/Pdtx/4000/4001 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual E - 1 E - 2...
  • Page 123 DCI/DCIP/4050/4051 LNCS Y 4053 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual E - 3 E - 4...
  • Page 124: The Accuracy Of Nellcor Spo 2 Sensors

    MAXAI, MAXPI, MAXII 1.62% 1.49% 1.57% 2.50% MAXNI 1.85% 1.71% 1.51% 1.59% SaO2 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual E - 5 E - 6 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 125 Modified Bland-Altman for SpO - D-YS, OXI-A/N, OXI-P/I Sensors: (SpO - SaO vs.SaO This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual E - 7 E - 8...
  • Page 126: Units

    ~ meter + plus milliampere hour = equal to VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 1 F - 2...
  • Page 127: Abbreviations

    COHb carboxyhemoglobin intensive care unit COPD chronic obstructive pulmonary disease identification VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 3 F - 4 To Purchase, Visit Avobus.com or call...
  • Page 128 Sync synchronization OxyCRG oxygen cardio-respirogram systolic pressure pArt artery pressure blood temperature pArt-D diastolic artery pressure temperature difference VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 5 F - 6...
  • Page 129: Temp

    This page intentionally left blank. consumption for one breath oxygen consumption index VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 7 F - 8...
  • Page 130 P/N: 046-024296-00(2.0) To Purchase, Visit Avobus.com or call 1-800-674-3655...