Page 1
VS 8/VS 8A Vital Signs Monitor Operator’s Manual To Purchase, Visit Avobus.com or call 1-800-674-3655...
Page 2
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) Release time: January 2022 owns the intellectual property rights to this Mindray product and this manual. This Revision: 2.0 manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
Page 3
This manual is based on the maximum configuration and therefore some contents may Please include the instrument model number, the serial number, and a description of the not apply to the product. If you have any question, please contact Mindray. problem with all requests for service.
Page 6
14.2.2 Example Review Dialog ......................14 - 1 16.6.3 Enabling Black Standby Screen ..................16 - 14 14.2.3 Reviewing the Tabular Trends .................... 14 - 2 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
Page 7
16.13.1 Setting Scanner Information ...................16 - 21 18.3 Cleaning and Disinfecting the Equipment ..................18 - 2 16.13.2 Identifying the Barcode Reader (for the non-Mindray Custom 2D Barcode Reader) 18.3.1 Approved Cleaning and Disinfecting Agents ..............18 - 2 16 - 21 18.3.2 Cleaning the Equipment .......................
Page 8
B.2 Radio Regulatory Compliance .........................B - 7 C Default Settings .........................C - 1 C.1 Parameters Default Settings ........................C - 1 C.1.1 SpO2 Default Settings ........................ C - 1 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
Page 9
CAUTION • Use only parts and accessories specified in this manual. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 1 - 1 1 - 2...
Page 10
1.1.3 Notes USB connector Equipotentiality NOTE • The equipment software copyright is solely owned by Mindray. No DEFIBRILLATION-PROOF TYPE BF DEFIBRILLATION-PROOF TYPE organization or individual shall resort to modifying, copying, or exchanging APPLIED PART CF APPLIED PART it or to any other infringement on it in any form or by any means without due permission.
Page 11
European Community Manufacturer MR Unsafe (magnetic resonance) TrueBP is Mindray's new non-invasive blood pressure measurement algorithm using linear inflation technology, which can measure the blood pressure quickly and comfortably during cuff inflation. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
Page 12
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 8/VS 8A Vital Signs Monitors are not intended for helicopter transport, hospital ambulance, or home use.
Page 13
USB connectors Connect approved USB devices, for example the keyboard, mouse, and barcode reader. Network connector VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 2 - 3 2 - 4 To Purchase, Visit Avobus.com...
Page 14
When using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential difference between them. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 2 - 5...
Page 15
Passwords should only be changed by authorized personnel. Contact to the IEC 60601-1. If you have any questions, please contact Mindray. the department manager or biomedical engineering department for the •...
Page 16
To use the AC power source, follow this procedure: Press the power switch to turn on the monitor. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 3...
Page 17
Tapping the screen Do not use the monitor on a patient if you suspect it is not working properly, or if it is mechanically damaged. Contact the service personnel or Mindray. ◆ To select an item from menus or lists, or select a quick key, tap on it with your finger.
Page 18
The barcode reader is connected to the monitor through the USB connector. If you are using the Mindray custom 2D barcode reader (Model HS-1R, HS-1RS or HS-1M), before using it for the first time, clear old data formats and configure the barcode reader.
Page 19
Wired network is Wired network is not connected. connected. The following table shows available quick keys. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 9 3 - 10 To Purchase, Visit Avobus.com...
Page 20
Select the Main Menu quick key → from the Configuration column select ■ Manage → input the required password → select → Workflow Setup. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 11 3 - 12...
Page 21
Monitoring NIBP monitoring on the same limb. The SpO alarm status is locked until the NIBP measurement ends. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 13 3 - 14...
Page 22
Brightness On Battery. You can set the brightness to 1 to 10, in parameter which 10 is the brightest, and 1 is the least bright. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 15...
There will be no physiological alarm related indications. ■ The alarm setup menus will not be displayed. ■ Alarm limits cannot be set. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 17 3 - 18...
You can connect the monitor to the BeneVision CMS through wired LAN or wireless LAN. after the Auto Enter Standby Time. When connected to the CMS, the system provides the following function. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 19...
Stopping a Parameter Measurement To stop monitoring a parameter, follow this procedure: Remove corresponding sensors from the patient. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 3 - 21 3 - 22 To Purchase, Visit Avobus.com...
30 minutes, you will be prompted whether to start monitoring a new patient if any of the above vital signs are detected again. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 4 - 1...
Every time you switch to a workflow, the alarm settings of the workflow are loaded. For details, see 3.9.1.4 Changing Alarm Settings. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
■ Physiological alarms are switched off. The alarm lamp does not flash and alarm sound is not issued. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 5 5 - 6...
Pause Time), pressing the Audio Pause quick key permanently switches off all alarm sound. The audio off state has the following features: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 7...
The nurse call system is enabled. ■ A user-defined alarm occurs. ■ Alarms are not paused or reset. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 5 - 9 5 - 10...
• A functional tester or SpO simulator cannot be used to assess the SpO accuracy. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 1 6 - 2...
Cuff or arterial blood pressure measurement device on the same limb as the Patient physiological characteristics: sensor ◆ Cardiac arrest ◆ Hypotension VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 3 6 - 4 To Purchase, Visit Avobus.com or call...
Select an appropriate extension cable according to the connector type and plug the cable into the SpO connector. Connect the sensor to the extension cable. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 5 6 - 6...
Sensitivity (for Masimo SpO For Masimo SpO , selects the Sensitivity as per signal quality and patient motion. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 7 6 - 8...
The reliability of FastSAT is dependent on the setting for the averaging time and the input signal. FastSAT is disabled by default. To enable FastSAT, follow this procedure: Set Averaging. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 9...
Page 38
This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http:// www.covidien.com/patents. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 6 - 11 6 - 12...
Page 39
Shenzhen Mindray. In no event may you transfer, assign, rent, lease, sell, to this device.
Page 40
Changing the PR Alarm Settings In Continuous Monitoring mode, you can change the PR alarm settings. Follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 7 - 1 7 - 2 To Purchase, Visit Avobus.com...
Page 41
Wash hands after temperature is taken. This will significantly reduce the risk Temp unit of cross contamination and nosocomial contamination. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 1 8 - 2...
Page 42
Axillary. The measurement site cannot be changed. The temperature type automatically changes to Predictive mode when the probe returns to the probe well. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 3...
Page 43
There are no parts that you can service yourself except for the battery, which you should replace when low by following the instructions in this manual. For To disinfect the temperature probe: service, repair, or adjustments, return your thermometer to Mindray. Disconnect the temperature probe from Temp connector. •...
Page 44
Release button remove from head and read measurement. High Ambient HI A >104 °F (40 °C) NOTE Low Ambient LO A <60 °F (16 °C) VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 7 8 - 8...
Page 45
You should clean the product as per your hospital’s regulations after each use. Sterilization is not allowed for the product. ■ Cleaning the case VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 9 8 - 10...
Page 46
You can manually input a Temp value and save the data for future reference. Follow this procedure: Select in the Temp area. For monitors configured with Mindray SmarTemp™ module, to input Temp value in Continuous Monitoring mode, press and hold in the Temp area.
Page 47
• NIBP measurement can be performed during electro-surgery and discharge of a defibrillator. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 1 9 - 2...
Page 48
Manual: measurement on demand. ■ Auto: repeated measurement at set interval or an interval synchronized with the real time clock. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 3 9 - 4...
Page 49
For patients taking orthostatic BP measurement, refer to 9.10 Orthostatic BP For more information, see 9.8.9 Correcting the NIBP Measurements. Measurement. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 5...
Select STAT. Patient position and site of the first measurement Patient position and site of the second measurement VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 7 9 - 8...
Select the Setup tab. Select the NIBP numeric area to enter the NIBP dialog. Set Display Format. Set Start Mode. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 9 9 - 10...
Select on the right of the workflow to be set. Make sure the workflow is under Spot Check. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 13 9 - 14...
If the problem persists, contact your service personnel. NOTE • For the physiological and technical alarm messages, see Appendix D Alarm Messages. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 9 - 17 9 - 18 To Purchase, Visit Avobus.com...
RR, follow this procedure: Select the Main Menu quick key → from the Configuration column select Workflow → input the required password → select VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 10 - 1...
Return to the main screen and switch workflows for the setting to take effect. This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 11 - 3...
MEWS, NEWS and NEWS2 are intended for adult patients only. The patient category applied to the Custom Score is defined by Mindray Clinical Score Configuration Tool. For more information, see Mindray Clinical Scoring Config Tool Instruction for Use (P/N: 046- 12.1.2...
Page 60
Alarm: the monitor automatically starts scoring when an alarm occurs to the parameter for scoring. ◆ If no option is selected, the monitor does not initiate auto scoring. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 3 12 - 4...
Page 61
To enable the auto refreshing score function, follow this procedure: viewed. From the EWS page select Setup. Select Filter and choose a protocol if needed. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 5 12 - 6 To Purchase, Visit Avobus.com...
Page 62
11.4 Changing the Display of Manual Parameters to add GCS to the list. • For details about how to import scoring tools, refer to Mindray Clinical Scoring Config Tool Instruction for Use (P/N: 046-012986-00). 12.2 Glasgow Coma Scale (GCS) The Glasgow Coma Scale (GCS) function is based on Teasdale’s Assessment of Coma and...
Page 63
From the Eye Opening area, Verbal Response area, and Motor Response area, Pain Score is applicable for adult, pediatric, and neonatal patients. respectively select an item that represents the patient’s status. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 9...
Page 64
Goal quick key. ■ Select the desired waveform area, parameter numeric area, or dashboard to enter corresponding parameter setup menu. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 11 12 - 12 To Purchase, Visit Avobus.com...
Page 65
From the SpO statistics area, select the duration to redefine the length of SpO statistics. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 12 - 13 12 - 14...
Page 66
Select Recorder Paper Speed to set the speed for recording waveforms. ■ Press the hardkey on the front of the recorder. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 13 - 1 13 - 2...
Page 67
If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam. If a paper jam is detected, follow this procedure to remove it: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
Page 68
(Temp, Scoring etc.) or type of data (Manually Saved for example). Select a record to be viewed and select Detail to check more information. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
Page 69
Different color blocks are displayed on the left of each event to indicate different event types. ■ Red: high priority alarm event VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 14 - 3 14 - 4...
Select the workflows and User Maintenance Settings to import. Select a department. Select OK. Select Import. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 15 - 1 15 - 2 To Purchase, Visit Avobus.com...
30 minutes, the monitor behaves the same as it is normally turned off. This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 15 - 3...
From the Maintenance area, set User Maintenance to select the password for data is cleared. accessing the monitor’s Maintenance dialog. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 3 16 - 4 To Purchase, Visit Avobus.com...
You can set the monitor to automatically log out of the clinician account after saving data. do so, follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 5...
High Alarm Interval 3 s to 15 s 10 s Med Alarm Interval 3 s to 30 s 20 s VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 7 16 - 8...
Off When Reset: when the alarm system is reset, both the alarm tone and Permanent. The default audio pause time is two minutes. alarm lamp of the current alarms are switched off. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 9...
Pulse: the nurse call signal is a pulse signal and each pulse lasts one second. so, follow this procedure: When multiple alarms simultaneously occur, only one pulse signal is VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 11...
→ input the required password → select To do so, follow this procedure: VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 13...
Select the Main Menu quick key → from the System column select Maintenance Set high limit, low limit and color for each level. → input the required password → select VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 15...
From the Pain Score Method area, switch on the pain scales as needed. → input the required password → select Select Module → Other tab. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 17...
Select the Time tab→ Time Synchronization tab. → input the required password → select Switch on Start NTP Time Sync. Select the Version tab. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 19 16 - 20...
Page 82
• Wireless network designing, deploying, debugging, and maintenance should To select desired patient information that will be read by the barcode reader, follow this be executed by Mindray service personnel or authorized technicians. procedure: • Always set the wireless network according to local wireless regulations.
Page 83
Auto, which means the monitor will automatically use LAN1 IP when detected and the available WLAN when LAN1 IP is not detected. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual...
Page 84
Select the Main Menu quick key → from the System column select Maintenance Set Multicast TTL. → input the required password → select VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 25...
Page 85
Set Automatically Logout after as needed. From the General Connectivity area, set the Master Server Address, Context Root, and Organization ID. Enable Local CSI as needed. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 27...
Page 86
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Other tab. Select Export Patient Data. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 29 16 - 30...
Page 87
Keep batteries out of the reach of children. • The on-screen battery symbols indicate the battery status as follows: Use only Mindray specified batteries. Use of a different battery may present a risk of fire or explosion. ■ indicates that the battery power is sufficient. The green portion represents •...
Page 88
Allow the monitor to run on the battery until the battery is completed and the monitor automatically shuts down. Fully charge the battery again for use or charge it to 40 – 60% for storage. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 17 - 3...
• Storing batteries in a cool place can slow the aging process. Ideally the batteries should be stored at 15 °C. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 17 - 5...
Isopropanol Liquid Isopropanol 70% CAUTION 1-Propanol Liquid 1-Propanol 50% • If you spill liquid on the equipment or accessories, contact Mindray or your Metrex CaviCide1™ Liquid Diisobutylphenoxyethoxyethyl dimethyl service personnel. benzyl ammonium chloride 0.28%, • Isopropanol 17.2% Avoid wetting the pins and metal parts of the equipment, mounting kits or accessories during cleaning and disinfection.
Never clean or disinfect the metalic connectors at either end of the Disposable Wipe n-Alkyl dimethyl benzyl ammonium chlorides 0.14% accessories. • Use only Mindray approved cleaners and disinfectants and methods listed in PDI Sani-Cloth® Plus Wipes n-Alkyl dimethyl ethylbenzyl ammonium chlorides Germicidal 0.125%,...
This equipment contains no user serviceable parts. 19.3 Maintenance and Testing Schedule • Software upgrades to the monitor can only be done by Mindray service Follow the maintenance and testing schedule or local regulations to perform testing and engineers with Mindray upgrading tools. •...
If your monitor needs a safety test and performance test, contact the service personnel. The NIBP accuracy test should be performed once every two years or when you doubt the NIBP measurements. The NIBP accuracy test should be performed by Mindray- 19.5.1 Performing Visual Inspection qualified service personnel only.
Page 96
For disposal of parts and accessories, where not otherwise specified, follow local regulations regarding disposal of hospital waste. This page intentionally left blank. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 19 - 5...
Disposable Infant (3 to therapy is performed. 20 kg) adhesive, thumb or great toe VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 20 - 1 20 - 2 To Purchase, Visit Avobus.com...
Page 98
Pediatric/ Neonate LNCS Inf 0600-00-0158 sensor, 20 pcs/ Disposable Infant (3-20 box, adhesive, thumb or great toe VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual 20 - 3 20 - 4...
Page 101
Current 0.9 to 0.5 A Frequency 50/60 Hz (±3 Hz) Fuse T2AL-250V VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 1 A - 2...
Page 102
Method Thermal dot array remote equipment Paper speed 25 mm/s Number of waveform channels A.5.7 Wireless Network VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 3 A - 4...
Page 103
ESU equipment. The 7.0m Co-channel 8.5m 5210MHz) interfering devices do not include Wi-Fi devices. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 5 A - 6 To Purchase, Visit Avobus.com...
Page 104
The total delay of data transmission from the monitor to the bluetooth node: ≤ 6 • seconds. Do not use Bluetooth nodes to perform realtime vital signs monitoring. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 7 A - 8...
Page 105
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s Low perfusion Pulse amplitude: >0.02% conditions Light penetration: >5% VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 9 A - 10...
Page 106
Mean: 15 to 260 15 to 215 15 to 125 Resolution 1 bpm VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 11 A - 12 To Purchase, Visit Avobus.com or call...
Page 107
* Measurement accuracy verification: In adult and pediatric modes, the blood pressure Oral 0.02°C 0.33°C 0.1°C measurements measured with this device are in compliance with the American National VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 13 A - 14...
Page 108
Alkaline battery This page intentionally left blank. Voltage Run time Approximately one hour continuous use or 15000 readings VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 15 A - 16...
Page 109
Otherwise, degradation of the performance of this device ■ Detect for connection could result. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 1 B - 2 To Purchase, Visit Avobus.com...
Page 110
±1 kV line(s) to ±1 kV line(s) to 61000-4-5 line(s) line(s) ±2 kV line(s) to ±2 kV line(s) to earth earth VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 3 B - 4...
Page 111
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 5...
Page 112
(2) this device must accept any interference received, including interference that may cause undesired operation. Any changes or modifications to this equipment not expressly approved by Mindray may cause harmful radio frequency interference and void your authority to operate this equipment.
Off, Med Item Default Setting Off, Med Alarm Volume Temp Off, Med VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 5 C - 6 To Purchase, Visit Avobus.com or call...
System Time Default Settings This page intentionally left blank. Item Default Setting Date Format yyyy-mm-dd 24-Hour Time Daylight Savings Time VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 7 C - 8...
In the following tables, A, B, and C are used to refer to the responses on alarm reset. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 1...
Check SpO2 No SpO sensor is connected. Sensor Connect an SpO sensor. Connection VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 3 D - 4 To Purchase, Visit Avobus.com or call...
If patient category is service personnel. correct, check that the tubing is not bent and the airway is not occluded. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 5 D - 6...
IP Address wireless network IP address. Check the network settings. Comm Error the alarm persists, contact your service personnel. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 7 D - 8...
Please patient. This page intentionally left blank. delete some discharged patients. VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual D - 9 D - 10 To Purchase, Visit Avobus.com or call...
Table information for the plots below show A values measured with Masimo Set Oximetry Technology in a clinical study. Adtx/Pdtx/4000/4001 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual E - 1 E - 2...
Page 123
DCI/DCIP/4050/4051 LNCS Y 4053 VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual E - 3 E - 4...
~ meter + plus milliampere hour = equal to VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 1 F - 2...
COHb carboxyhemoglobin intensive care unit COPD chronic obstructive pulmonary disease identification VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 3 F - 4 To Purchase, Visit Avobus.com or call...
Page 128
Sync synchronization OxyCRG oxygen cardio-respirogram systolic pressure pArt artery pressure blood temperature pArt-D diastolic artery pressure temperature difference VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 5 F - 6...
This page intentionally left blank. consumption for one breath oxygen consumption index VS 8/VS 8A Vital Signs Monitor Operator’s Manual VS 8/VS 8A Vital Signs Monitor Operator’s Manual F - 7 F - 8...
Page 130
P/N: 046-024296-00(2.0) To Purchase, Visit Avobus.com or call 1-800-674-3655...