S C H M I D T c o n t r o l i n s t r u m e n t s Edition FSR 01.E FS Series Model FSR Instruction Manual Valid as of: 01.09.2017 • Please keep the manual for future reference!
1.9 Declaration of conformity, RoHs II and WEEE registration ........4 2 Available models ...................... 5 2.1 General information ....................5 2.2 Specifications of the FS series ................5 2.3 Assignment of the connector .................. 6 2.3.1 Connecting the tension meter................6 2.4 Optional accessories ....................
Warranty: - SCHMIDT tension meters are warranted for 12 months. Parts subject to wear, electronic components and measuring springs are not covered by the warranty. No warranty or liability will be accepted for bodily injury or property damage resulting from one or several of the following causes: - Misuse or abuse of the device.
The device is intended exclusively to be used as a tension meter. Any other use or any use exceeding this intention will be regarded as misuse. Under no circumstances shall Hans Schmidt & Co GmbH be held liable for damage result- ing from misuse.
Available models 2.1 General information These instruction manual refer to the FSR Model of the FS Series: Model Tension Ranges N FSR-10N 0 - 10 FSR-20 N 0 - 20 FSR-50 N 0 - 50 FSR-100 N 0 - 100...
Hans Schmidt & Co GmbH. Certification to the CE specification does not extend to, and shall be invalid for any other combina- tion. Under no circumstances shall Hans Schmidt & Co GmbH be held liable for any damage resulting from the use of non-SCHMIDT sensors or cables.
TSR sensor, SCD-1 display unit and an 5 m long connecting cable (special cable). In no event shall Hans Schmidt & Co GmbH take over any warranty coverage for, or shall be held liable for, any damage resulting from self-made cables.
- If the material path is other than vertical or if the process material deviates significantly from the SCHMIDT calibration material, you need to carry out zero adjustment and gain adjustment as described in chapters 3.4.1 and 3.4.2 before starting measurement.
General information: All tension meters are calibrated with standard materials - such as polyamide monofila- ment (PA) - according to the SCHMIDT factory procedure; therfore the wraping angle is 180°. In any case you need to carry out static zero and gain adjustment as described in chapters 3.4.1 and 3.4.2.
3.4.2 Gain adjustment Requirement: Zero adjustment carried out. Thread the process material in the measuring roller Force direction Measuring roller Gain Zero Weight fig. 3.4.2a fig. 3.4.2b - Hang a weight that corresponds 100% of the tension range from the process material. - Adjust the potentiometer, which you can reach through the gain hole in the housing, with a screwdriver (with a point width of max.
The question of finding the right frequency of calibration accuracy verification depends on several different factors: Operating time and load of the SCHMIDT tension meter Tolerance band defined by the customer Changes to the tolerance band compared to previous calibrations Therefore, the interval between verifications of calibration must be determined by the user‘s Quality Assurance Department, based on the user‘s experience.
Please indicate in your order whether you require an Inspection Certificate 3.1 according to DIN EN 10204. Service address: Hans Schmidt & Co GmbH Schichtstr. 16 84478 Waldkraiburg Germany...
Need help?
Do you have a question about the FS Series and is the answer not in the manual?
Questions and answers