3: Operating Instructions
Shutter Switch
Manually set the operating mode of the DH-3PLUS with the Shutter Switch as follows:
OPEN - Shutter open
CLOSE - Shutter closed
TTL – Remote operation of shutter
Remote Operation of the LAMP and Shutter
For remote operation, plug the D-SUB 15-pin connector into the appropriate socket on your
spectrometer.
USE ONLY THE CABLE PROVIDED OR A CUSTOM CABLE DESIGNED TO INTERFACE
WITH THE PINOUTS AS DESCRIBED IN THE PINOUT INFORMATION SECTION. DO NOT
USE A VGA VIDEO CABLE.
Set the front panel shutter switch to the TTL position
Pin 2 of the connector is the Deuterium LAMP ON/OFF control.
Pin 3 of the connector is the Halogen LAMP ON/OFF control.
Pin 13 of the connector is the shutter control. (HIGH=OPEN, LOW=CLOSE)
Refer to
Pinout Information
14
for information on the operation of lamp and shutter.
355-00000-003-01