Standard Uncertainties and Uncertainty Budgets in % at Wavelength
For DH-3P (-BAL) CAL Cosine Corrector Calibration
UV Calibration
230
Standard
deviation
k = 1
7.7
k = 2
15.4
For DH-3P (-BAL) CAL Bare Fiber Calibration
UV Calibration
230
Standard
deviation
k = 1
N/A
k = 2
N/A
Place the light source horizontally. A value of 6 % of expanded uncertainty must be taken into
account for vertical placement.
000-10000-191-02-201601
Uncertainty Budgets in [%] at Wavelength
Standard VIS Calibration
250
300
400
500
6.9
6.5
6.0
4.1
13.8
13.0
12.0
8.2
Uncertainty Budgets in [%] at Wavelength
Standard VIS Calibration
250
300
400
500
9.5
8.7
8.1
19.0
17.4
16.1
13.6
Wavelength (nm)
600
1000
1400
3.9
3.8
7.7
7.7
Wavelength (nm)
600
1000
1400
6.8
6.7
6.7
13.4
13.3
A: Calibration Basics
Extended Calibration
1600
2000
4.3
5.6
10.6
8.5
11.2
21.2
Extended Calibration
1600
2000
6.9
7.8
11.9
13.8
15.6
23.8
2400
11.7
23.5
2400
12.9
25.9
19