Your DH-3PLUS Series Light Source package should contain the following:
DH-3PLUS light source unit
Power cord
UV safety goggles
One IC-DB15-2 interface cable for shutter operation
Lamp Calibration Report for using the DH-3PLUS light source with a CC-3-UV-S cosine-
corrected irradiance probe
USB containing lamp calibration files
Uncertainty report for unit
Calibration certificate for unit
Allen wrench for adjusting the cosine corrector
Connecting a Spectrometer to the DH-3PLUS
Before using a DH-3PLUS Series light source for the first time, check for transport damage. Be
sure to adhere to all warnings on the unit and in this operational manual. Make sure you have
your DH-3PLUS light source, your spectrometer, a CC-3 cosine-corrected irradiance probe, and
Ocean Optics spectrometer operating software.
Procedure
Use the following procedure to connect your DH-3PLUS light source to a spectrometer:
1. Loosen the set screw on the DH-3PLUS's SMA connector.
2. Screw the CC-3 cosine corrector onto the end of the fiber. The connection should be
tight. Insert the CC-3 all the way into the DH-3PLUS's SMA connector.
3. Tighten the setscrew on the SMA connector of the DH-3PLUS with an Allen wrench.
Connect the other end of the fiber into the SMA connector of the spectrometer.
4. Plug in the power cord.
5. Turn the unit on using the ON/OFF switch on the back of the light source.
6. Turn on the deuterium light and the halogen light on the front of the light source.
7. Allow the lamp to warm up for at least 15 minutes before using it. Place the DH-3PLUS
horizontally.
8. Insert the calibration USB drive that came with your light source into your computer. This
disk contains one ASCII file with the same information as the Lamp Calibration Reports
that came with your DH-3PLUS. The file name contains the lamp's serial number,
followed by CC.LMP. Copy these files to the desired directory on your computer.
9. Start your spectrometer operating software. See your spectrometer manual for
instructions on calibrating the spectrometer with the DH-3PLUS.