Guidance And Manufacturer's Declaration - Kinetec Breva User Manual

Hide thumbs Also See for Breva:
Table of Contents

Advertisement

14.8.

Guidance and manufacturer's declaration

Guidance and manufacturer's declaration - Electromagnetic emissions
The device KINETEC
Breva is intended for use in the electromagnetic environment specified below. The customer or the user of the device
®
KINETEC
Breva should ensure that it is used in such an environment.
®
Emissions test
Radio frequency emissions CISPR 11
Radio frequency emissions - CISPR 11
Harmonic emissions - IEC 61000-3-2
Voltage fluctuations / Flicker emissions - IEC
61000-3-3
Guidance and manufacturer's declaration - Electromagnetic immunity
The device KINETEC
Breva is intended for use in the electromagnetic environment specified below. The customer or the user of the device
®
KINETEC
Breva should ensure that it is used in such an environment.
®
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient /
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage interruptions
IEC 61000-4-11
Voltage dips
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field - IEC
61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level.
T
Breva
Compliance
Group 1
Class B
Class A
Complies
IEC 60601
Test level
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV
for power supply lines
±0,5 kV, ±1 kV
between lines
±0,5 kV, ±1 kV, ±2 kV
between line and earth
< 5% U
T
(>95% dip in U
)
T
for 5 seconds
< 5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
3A/m
Electromagnetic environment – guidance
The device KINETEC
Breva uses RF energy only for its internal function.
®
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The device KINETEC
Breva is suitable for use in all establishments including
®
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Compliance level
Electromagnetic environment – Guidance
±2 kV, ±4 kV, ±6 kV
contact
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
±2 kV, ±4 kV, ±8 kV
should be at least 30%.
air
±2 kV
Mains power quality should be that of a typical
for power supply lines
commercial or hospital environment.
±0,5 kV, ±1 kV
between lines
Mains power quality should be that of a typical
±0.5 kV, ±1 kV, ±2 kV
commercial or hospital environment.
between line and
earth
< 5% U
T
When the interruption occurs, the device KINETEC
(>95% dip in U
)
T
is reset. After turning on, push START to begin the session.
for 5 seconds
< 5% U
T
(>95% dip in U
)
T
for 0.5 cycle
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
40% U
T
device KINETEC
(60% dip in U
)
T
during power supply interruptions, we recommend
for 5 cycles
powering the device KINETEC
uninterruptible power supply or a battery.
70% U
T
(30% dip in U
)
T
for 25 cycles
Power frequency magnetic fields should be at levels
3A/m
characteristic of a typical location in a typical
commercial or hospital environment.
EN
Breva requires continued operation
®
Breva using an
®
Breva
®
19/20

Advertisement

Table of Contents
loading

Table of Contents