Troubleshooting
Table 22, Event Messages
ID #
Event Message Text
305
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• OPEN the white air line clamp.
• CLOSE the blue PRP line clamp.
• CLOSE the yellow PPP line clamp.
2. Touch Continue to empty the bowl.
306
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• CLOSE the white air line clamp.
• CLOSE the blue PRP line clamp.
• OPEN the yellow PPP line clamp.
2. Touch Continue to empty the bowl.
307
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• OPEN the white air line clamp.
• CLOSE the blue PRP line clamp.
• CLOSE the yellow PPP line clamp.
2. Agitate the RBC bag.
3. Touch Continue to concentrate.
®
Haemonetics
Cell Saver
Operator Action Required
Operator Action Required
Operator Action Required
®
®
Elite
+ User Manual
P/N 130859-US, Manual Revision: AA
209
Need help?
Do you have a question about the Cell Saver Elite+ and is the answer not in the manual?