Table 22, Event Messages
ID #
Event Message Text
114
Explanation:
The air detector sensed air in the blue line during the Conc phase. The RBC bag may be empty.
Sequestration has paused.
Corrective Action:
If the RBC bag is not empty:
1. Ensure the blue line clamp is open.
2. Check the tubing for kinks, occlusions, and correct placement.
3. Touch Continue Using RBC Bag to continue.
If the RBC bag is empty:
1. Replace the blood bag.
2. Touch Continue Using Blood Bag to proceed.
To end the cycle:
1. Touch End Cycle to end the current cycle.
2. The device will empty the bowl contents to the RBC bag.
115
Explanation:
The air detector sensed air before the expected volume was pumped from the bowl.
Corrective Action:
1. Check the tubing for correct placement in the air detector and pump.
2. Check the tubing for kinks and occlusions.
3. Touch Continue to continue.
NOTE: When in Quick Transfer mode, the following steps are also displayed:
4. Touch New Transfer to begin a new Fill cycle.
5. Touch End Quick Transfer to end Quick Transfer and return to Cell Salvage automated
mode.
116
Explanation:
An issue has been detected with the air detector.
Corrective Action:
Restart the device.
If the problem persists, service is required.
P/N 130859-US, Manual Revision: AA
Air Detected During Conc
Air Detected Early
Air Detector Failure