Haemonetics Cell Saver Elite+ User Manual page 168

Autotransfusion system
Hide thumbs Also See for Cell Saver Elite+:
Table of Contents

Advertisement

168
Table 22, Event Messages
ID #
Event Message Text
108
Explanation:
The air detector sensed air in the yellow line. The saline bag may be empty.
Corrective Action:
1. Replace the saline bag if it is empty.
2. Ensure the yellow line clamp is open.
3. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
4. Touch Continue to continue.
110
Explanation:
The air detector sensed air in the blue line during the Concentrate phase. The RBC bag may be
empty.
Corrective Action:
1. Ensure the blue line clamp is open.
2. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
3. Wait for additional fluid in the reservoir.
OR
Touch Wash to wash a partial bowl.
111
Explanation:
The optics have detected red cells. The bowl should be full and is ready to transition to the Wash
phase.
Corrective Action:
1. Touch Wash to enter the Wash phase OR
2. Touch  to continue filling the bowl OR
3. Touch Return to return cells to the reservoir.
P/N 130859-US, Manual Revision: AA
Air Detected in Yellow Line
Air Detected During Conc
Bowl is Ready to Wash
Haemonetics
®
®
®
Cell Saver
Elite
+ User Manual