Troubleshooting
Table 22, Event Messages
ID #
Event Message Text
253
Explanation:
A disposable must be loaded to begin a procedure.
Corrective Action:
Load a disposable and touch Start Procedure when ready to begin a procedure.
254
Explanation:
A device event has caused the centrifuge to stop and the PRP separation to be lost.
The RBCs in the centrifuge should be emptied to the RBC bag and the cycle started over.
Corrective Action:
1. Touch Empty Bowl to empty RBCs to the RBC bag.
2. Touch Conc to repeat Sequestration with RBCs from the RBC bag.
OR
1. Touch Empty Bowl to empty RBCs back to the RBC bag.
2. Touch Fill to repeat Sequestration with RBCs from the blood bag.
255
Explanation:
The procedure is establishing a platelet layer. The pump will restart after a 30 second delay
when separation is established.
Corrective Action:
Touch to override the delay.
256
Explanation:
The device was transmitting data to a storage device and the transmission was interrupted. This
could occur if the storage device was removed from the USB port before completing data
transfer or if the connection between the storage device and the USB port was loose.
Export of the data will have to be repeated.
Corrective Action:
1. Ensure the storage device is securely installed in the USB port.
2. Touch Export to begin the data transfer.
®
Haemonetics
Cell Saver
PRP Separation Lost
Data Transfer Interrupted
®
®
Elite
+ User Manual
Load Disposable
Recentrifuge Delay
P/N 130859-US, Manual Revision: AA
199