Quick Reference Guide This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ®...
Page 2
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Prepare the device: 4. Power on device and allow system self test to finish. 1. Unfold biohazard waste bag, stored in tray on underside of device.
Page 3
Load Disposable / Initiate Procedure (page 2 of 3) Load processing set: 3. Thread pump tubing around pump and place tubing manifold into Open Cell Saver Elite+ processing position. set and check for integrity. 4. Floss tubing into air detector and 2.
Page 4
IV pole. Close device cover. port on bag is securely closed. Close both yellow line tubing 2. Touch Active Settings: Haemonetics clamp(s) and spike saline bag(s). Default or Fat reduction. 3. Touch Start Procedure. NOTE: System will begin processing when minimum fluid level is collected.
Page 5
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Supply Storage location Cell Saver Elite + system ® ® Wash solution 1 liter or 3 liter bag of sterile 0.9% saline for injection, USP ...
Page 6
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® After surgery is complete: Follow the steps below to transition the device to postoperative operation. 1. Turn off suction. 2.
Page 7
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® NOTE: When emptying/changing waste bag, do not lose the displacement air from the system; if this occurs, the bowl may not empty properly. To ensure this does not happen do not empty the waste bag below the indicated line.
Page 8
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® NOTE: Manual Mode must be enabled in a settings group before it can be utilized. Please refer to Operation Manual for instructions to enable access and activate Manual Mode in a new Settings Group. This feature enables the user to control the Cell Saver Elite+ device manually for cell salvage with minimal sensor monitoring.
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Filling the Bowl 1. Touch Fill to initiate the Fill phase. NOTE: The Fill phase does not stop until the user presses another key, such as the Wash key. Washing the Cells As the bowl is filling during either the Fill or Concentrate phase, observe when the RBC/supernatant interface reaches a point just below the neck of the bowl.
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® In emergency situations where users are managing high blood-loss, it may be necessary to accelerate the processing time to quickly return blood to the RBC bag for reinfusion to the patient.
Page 11
Emergency Options (page 2 of 3) Emergency Mode During Emergency Mode the device processes blood continuously at high speeds through the Fill, Wash, and Empty phases until the air detector senses air for the first time in the Fill phase, indicating that the reservoir is empty. The device then reverts to the previous settings group and enters Standby.
Page 12
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Quick Transfer Mode NOTE: Quick Transfer Mode must be enabled in a settings group before it can be utilized. Please refer to Operation Manual for instructions to enable access and activate Quick Transfer Mode in a new Settings Group.
Page 13
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Use this method to process a partial bowl of blood at the end of a Cell Saver Elite+ procedure or whenever it is necessary to process blood before a full bowl has been collected.
Page 14
This guide is not intended as a substitute for the Cell Saver This guide is not intended as a substitute for the Cell Saver Elite Elite + System Operation Manual or authorized training. + System Operation Manual or authorized training. ®...
Page 15
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® When procedure is complete: 1. Touch End Procedure, then confirm by touching again End Procedure. System empties blue line and Procedure Complete screen appears.
Page 16
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Use this method if it is not clear that sufficient volume will be collected during a procedure to process for reinfusion. 1.
3. Use arrows to select a parameter. 3. Enter name for new 4. Use arrows to change parameter. settings group. 5. Touch "done" to save parameters. NOTE: Haemonetics Default and Fat reduction are settings groups that cannot be edited or deleted. TRN-QRG-100125-US(AC) Protocol Settings...
Page 18
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® WARNING: Avoid suction of the following substances into the Cell Saver Elite+ system reservoir in accordance with AABB Guidelines. A more complete list can be found in the operation manual.
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Troubleshooting scenarios Problem Possible cause Action 1. Ensure vacuum line is securely connected to device and reservoir and ...
Page 20
Troubleshooting/Tips (page 2 of 3) Event messages Message Action 1. Empty or replace the waste bag. 2. Touch Continue to continue. Waste Bag Full NOTE: When emptying waste bag, ensure fluid level in bag stays above 1 liter mark. NOTE: When replacing waste bag, ensure bowl is empty.
Page 21
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® Changing target wash volume during a procedure: 1. Touch Cycle Wash Volume. A dialog box appears. 2. Use arrows to change target wash volume for current wash cycle.
Page 22
This guide is not intended as a substitute for the Cell Saver Elite + System Operation Manual or authorized training. ® ® The Records screen displays information about the consumables, procedures and device and allows you to enter additional information and export it to an external storage device. NOTE: The barcode reader is active when the device is in the Records screen.