Deleting Tool Error Compensation Values; Reading Tool Error Compensation Values; Multiple Remount Balancing - Hofmann MC10 HS Operator's Manual

Unbalance measuring system
Table of Contents

Advertisement

MC10 HS
3
2
Operation
Remove rotor.
INFORMATION
Remount balancing and the determination of tool er-
ror compensation values is now complete. The tool
error compensation values are included in the dis-
played measurement values.
5.5.2
Deleting Tool Error Compensation
Values
Press "Index. Bal." key (1).
Press "Clear remount-value" key (2).
INFORMATION
The tool error compensation run is deleted. No tool
error compensation values are included in the dis-
played measurement values.
5.5.3
Reading Tool Error Compensation
Values
Press "Index. Bal." key (1).
Press "Recall last value" key (3).
INFORMATION
The tool error compensation run is now read. Tool er-
ror compensation values determined during the last
compensation run are included in the displayed
measurement values.
5.5.4

Multiple Remount Balancing

INFORMATION
The unbalance measuring system provides the pos-
sibility of multiple remount balancing.
In order to obtain most accurate tool error compensa-
tion values, rotors are rotated by 180°.
However, there are rotors and adaptors, where this is
not possible; e.g. a rotor fixed to the adaptor by three
bolts. The second measuring run is then carried out
for the rotor rotated by 120°.
In this case, a second measuring run is now possible
for a remount angle of 240°. The values obtained
from this measurement are averaged with the previ-
ously obtained ones.
For multiple remount balancing, the remount
angle must also be between 90° and 270°. The
0° position cannot be repeated.
- Remount Balancing (Option)
5 - 21