Invacare® Perfecto ₂™V Oxygen Concentrator
10 Technical d ata

10.1 Specifications

Electrical R equirements:
Rated Current I nput:
Sound Pressure Level:
Sound Power L evel:
Altitude:
Oxygen Output C oncentration Levels:
Maximum Outlet P ressure:
Low Flow Alarm:
Potential O bstruction Alert:
30
120 VAC + 10, – 15% (132 VAC/102 VAC), 6 0 Hz
3 A
< 53 dBA weighted for 3 L/min flow
< 53 dBA weighted for 5 L/min Flow
< 53 dBA weighted for 5 L/min Flow
Up to 8,000 ft ( 2438 m) a bove sea l evel without d egradation of
concentration levels.
While filling a HomeFill S ystem:
sea level w ithout d egradation of c oncentration levels.
Atmospheric Pressure Range: 101.3 kPa – 75.0 kPa
87% to 95.6% at 0 .5 to 5 L/min
Concentration levels a chieved after i nitial w arm-up period
(approximately 30 minutes)
5 psi ± 0.5 psi ( 34.5 kPa ± 3.45 kPa)
0.5 to 5 L/min (maximum). For f lowrates l ess t han 1 L/min, we
recommend the use of t he Invacare Pediatric Flowmeter A ccessory
(IRCPF16).
0 L/min to 0.5 L/min
The concentrator d etects a condition that m ay indicate a potential
obstruction of t he output o xygen. Rapid audible beeping alert ( this
alert i s deactivated when accessories are connected) m ay be associated
with flow setting of 0 .5 L/min or l ess.
Up to 6,000 ft ( 1828 m) a bove
1193322-A-04