Power C onsumption:
Pressure Relief M echanism Operational a t:
Change in maximum recommended flow when back pressure of 7 kPa is
applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient T emperature and Humidity:
Cabinet:
Standards a nd Regulatory Listing:
1193322-A-04
325 W Typical
30 psi t o 45 psi ( 207 kPa to 310 kPa )
0.7 L/min (Back Pressure of 7 kPa applied)
Cabinet, O utlet H EPA and Compressor I nlet
Current o verload or l ine surge shutdown.
High temperature compressor s hutdown.
High Pressure Alarm with compressor s hutdown.
Low Pressure Alarm with compressor s hutdown.
Battery Free Power L oss A larm.
SensO
Oxygen System.
2
Possible Obstruction Alert.
15 in ± 3/8 in (38.1 cm ± 1 cm)
23 in ± 3/8 in (58.4 cm ± 1 cm)
12 in ± 3/8 in (30.5 cm ± 1 cm)
40 lbs ± 2 lbs ( 18.1 kg ± 1 kg)
45 lbs ± 2 lbs ( 20.4 kg ± 1 kg)
50°F to 95°F (10°C to 35°C) a t u p to 60% relative humidity
Impact R esistant f lame-retardant p lastic cabinet t hat c onforms t o UL
94V–0
ETL certified complying with AAMI E S60601-1, IEC 60601-1, IEC
60601-1-2, I EC 60601-1-6, I EC 60601-1-11, C SA C22.2#60601-1, I SO
80601-2-69