3. Insert 5.0 mm punch into Fowler retaining lug (Effectively hinge for whole fowler frame
located at foot end of fowler frame either end of 25 mm Dia. tube.) Push back retaining pin
until pin is free of nylon bush and raise that side of fowler slightly to clear the mount.
Fowler frame and sheet should then move sideways and remove from top frame.
Replacement of fowler frame is the reverse of the above procedure.
Use new pivot bushes and other components if they are worn.
Pivot Points:
Replace pivot pins, bolts or removable bushes (where fitted) if there is significant play or the
there is wear greater than 1mm in diameter.

11. Technical specifications:

Width outside: 92cm
Length outside: 215cm
Width inside: 80cm
Length inside: 190cm
Calf rest lift:
Thigh angle:
o
0-38
0-13
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2010 Invacare Asia Pacific
1600 User's Manual
Height adjustability:
20-67cm
Tilt option:
o
N/A
Mattress support
Backrest angle:
dimensions: 190x80cm
0-70
Safe working load:
Max user weight:
165kg
200kg
14
All measurements are
stated in millimetres
(mm). All angles are
stated in degrees.
All measurements
and angles are stated
without tolerances.
Invacare reserves the
right to change the
stated measurements
and angles without
warning.
Maximum weight
carrying capacity is
200kg, tested as per
ASNZ standards
(includes patient
weight of 165kg).