Maintenance; Simple Troubleshooting - S&A CW-5000 User Manual

Industrial chiller
Table of Contents

Advertisement

Condition
Circulating pump works properly
Ultrahigh room temp
Ultrahigh water temp
Ultralow water temp
Faulted room temp sensor
Faulted water temp sensor
Flow rate 1 alarm
Flow rate 2 alarm
Chiller power failure
Note: The alarm output port is connected with a set of normally open and normally closed
contacts of the relay inside the machine. The working current should be smaller than 3A while
the working voltage should be smaller than 300V.
VII.Maintenance
The equipment must be shut down for maintenance, and the power supply must be cut
off. The operation can only be carried out after 3 minutes, otherwise there will be a risk of
electric shock. When the room temperature is lower than 2°C, the internal water must be
drained when the machine is shut down for a long time.
Regularly disassemble the filter gauze for cleaning and use a compressed air gun to blow
away the dust on the condenser (about 1 week is suggested and the air filter must not be
missing for a long time).
VIII. Simple Troubleshooting
Failure
Machine turned on
but unelectrified
Display
Alarm
Buzzer
code
E1
Sounds
E2
Sounds
E3
Sounds
E4
Sounds
E5
Sounds
E6
Sounds
E7
Sounds
Failure Cause
Power cord is not plugged in place
Fuse burnt-out
14
TY-JS-220308-001
OUT H1、H2
OUT H1、H3
DISCONNECTION
BREAKOVER
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
BREAKOVER
DISCONNECTION
Approach
Check and ensure the power
interface and the power plug is
plugged in place and in good
contact.
Pull out the fuse box from the
power supply interface of the
chiller, check the fuse, replace