EASTMAN Kodak X-Omat 5000 RA Operator's Manual page 84

Table of Contents

Advertisement

OPERATORS MANUAL
Transport Failure
Surface Artifacts
Abnormal Film Densities
Wet Films
Low Solution Levels
84
Overlapping of Films
If incoming wash water is dirty, clean the Wash Rack, Wash Tank, and Water
Reservoir thoroughly. Change the incoming water Filter. Be sure to use the
correct water Filter.
Check incoming water temperature. Temperature must be between 4°C
(40°F) and 29°C (85°F).
Check that the correct Bulb and Safelight Filter are in the Safelight and at the
correct distance from the Feed Tray and work surface.
Check that the Top Cover and the Panels are secure on the Processor.
If the Processor is installed through the wall, check that there are no leaks in
the Lighttight Gasket.
If 10 x 10 cm films fail to transport reliably, feed the sheets of films diagonally.
Check the time delay. For all transport speeds, the film feed signal should
sound once the trailing edge of the film has advanced 76 mm (3 in.) into the
Processor.
Ambient conditions are out of specification.
If the solution temperature is low, the Processor will not accept film. Check
that the temperature lockout is selected.
Check that solution Level Probes are clean and free from buildup.
Check that all external replenishment lines are without kinks or air bubbles.
May 1997 – 5B6729